ABCDEFGHIJKLM
1
2
3
405488561639
4
installation value3.2114.72.4measured by Michael Pulla through 40x/0.95
5
date (type yyyy-mm-dd)timetemperature (at laser)running hoursmWmWmWmWcomments
6
31/05/20121:30 pm0.51.027.232.251.73borrowed "new" lpm from MPI LMFstill measured through 10x/0.45
7
13/06/2012exchange AOTF controller (due to some troubles and as it was not sure if we have the most up to date version since there was no sticker)
8
13/06/20120.727.221.3measured with Raik Staeber's lpm (PT9610) - low, another laser tuning needed
9
21/06/20122.58.23.21.9measured with Raik Staeber's lpm (PT9610), after fine laser coupling
10
25/07/2012permanent exchange HRm with new MRm (by Raik Staeber), due to ZEN-HRm-read out problem. results in 3000,- "Gutschrift"
11
30/07/2012exchange internal (Observer) RL shutter; exchange power supply PC by Raik Staeber
12
15/10/20122:30 am1.78.073.41.4Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
13
19/11/201210:00 am0.51.57.63.60.9Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
14
03/12/201210:40 am11.48.54.250.9620/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
15
10/12/201210:00 am0.751.18.354.20.75
16
17/12/2012Raik Stäber - laser recoupling
17
17/12/20121:00 pm2h2,29,04,22,020/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
18
08/01/201311:50 am0,52,259,224,22,120/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
19
11/02/201310:40 am221.51.859.024.52.1620/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
20
25/02/201310:55 am2221.618.84.52.1920/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
21
exchange PC by Raik Staeber
22
11/03/201310:20 am2211.718.644.392.1320/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
23
22/04/201310:45 am221.51.468.74.52.0120/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
24
06/05/201310:30 am220.751.578.624.421.8920/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
25
29/05/021310:05 am2211.78.94.661.8920/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
26
12/06/201310:50 am2211.548.764.421.620/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
27
24/06/201311:20 am2221.338.064.391.2220/0,8; Silke+Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
28
01/07/201311:10 am2211.6217.84.271.5520/0,8; Silke+Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
29
15/07/201310:40 am2221.57.84.381.3920/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610), air condition not working properly --> hot air is blown into the room (Zuluftschacht), temp. control set on 22
30
29/07/201311:00 am2221.28.44.71.1220/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
31
13/08/201310:15 am2211.28.34.591.2420/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
32
02/09/201310:30 am220.51.57.884.351.8620/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
33
16/09/201311:11 am220.51.197.94.281.6720/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
34
01/10/201310:20:00220.51.4298.393.952.0520/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
35
21/10/201317:00:00220.751.5297.864.251.6420/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
36
06/11/201311:30:00220.751.68.454.222.2820/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
37
25/11/201310:30:002221.628.44.52.320/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
38
09/12/201312:00:002221.417.974.42.2220/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
39
20/01/201410:50:00221.51.488.374.542.3620/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
40
03/02/201410:40:00220.751.388.234.452.320/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
41
18/02/201410:50:00220.51.417.894.232.2620/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
42
10/03/201410:45:00220.51.337.274.072.0920/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
43
24/03/201410:45:002221.347.584.192.8420/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
44
07/04/201411:00:002211.4628.224.422.1820/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
45
28/04/201410:50:002211.4377.9354.121.8120/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
46
19/05/201411:15:00220.751.428.034.1862.1220/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
47
02/06/201411:10:002211.327.654.21.9520/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
48
16/06/201411:00:002211.297.494.161.7920/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
49
07/07/201411:00:002211.138.334.171.7520/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
50
28/07/201411:30:00220.750.947.224.51.0220/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
51
01/09/201411:30:002211.0447.0674.3051.220/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
52
22/09/201411:00:002220.527.024.321.220/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610)
53
02/10/201411:10:002211.147.274.51.720/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
54
Jahreswartung
55
26/11/20141.829.74.522.24Raik, Jahreswartung, open laser module, measured with raik powermeter
56
1.38.454.432.36Raik, Jahreswartung, closed laser module, measured with raik powermeter
57
08/12/201411:00:00220.51.17.72.92.120/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
58
22/01/201511:00:0022118.244.362.3920/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
59
30/03/201511:002220.657.83.652.0220/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
60
29/04/201510:30220.51.128.23.62.3520/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
61
01/06/201510:002210.878.184.32.2620/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
62
23/06/201510:302210.97.64.22.1820/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
63
Jahreswartung Raik Stäber
64
12/10/201514:002230.77.641.820/0,8; Raik Stäber, measured with new IP-measuremeter (Gigahertz PT 9610)
65
19/04/201611:002220.496.483.521.8220/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
66
07/07/201610:30230.50.726.23.751.9620/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
67
06/09/201614:30230.50.655.743.541.7820/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)
68
25/04/201711:30232.50.286.093.471.5720/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610)service ordered for 405nm
69
02/05/2017Raik Stäber - laser recoupling (405nm laser coupling was very far off)
70
02/05/20171.25.333.21.810x/0.45; Raik measured with DZNE-IP measure meter
71
08/05/201709:30230.50.61---20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610), only 405nm switched on
72
08/05/201709:30230.50.4885.513.11.920/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610), all lasers switched on
73
13/06/2017Raik Stäber - exchange AOTF
74
13/06/201726.663.31.6510x/0.45; Raik measured with DZNE-IP measure meter
75
11/10/2018 - 19/10/2018Move into new building: Spedition Fiedler. responsible technicians:
Daniel Sommer: M1-LSM780NLO + general responsibility for move
Raik Staeber: SD1-Spinning Disk + SD2-Spinning Disk incubation
Robin Dutscheck: C1-LSM700, W1-ApoTome.2 first floor, W2-ApoTome.2
76
77
78
79
air conditioning not working yet in new rooms, very hard to couple in lasers properly, impossible for 405nm. Raik is going to come back for that
80
21/02/201914:0022.30.51.85.82.611.5210x/0.45; measured with DZNE-IP measure meter
81
2019-06-2410:0010.536.084.333.2310x/0.45; measured with DZNE-IP power meter
82
2019-07-030.810x/0.45; measured with DZNE-IP power meter
83
2019-07-04In front of laser4229.54397Service visit Robin Dutscheck; exchange PC - not working, Robin left loan PC here; recoupling 405nm; laser power measured through 10x/0.45 with Robins power meter
84
Out of fibre17.4102254
85
Out of objective1.655.62.751.8
86
1.736.12.91.610x/0.45; measured with DZNE-IP power meter
87
22/07/201911:0022.410.225.72.51.810x/0.45; measured with DZNE-IP power meter
88
23/07/201910:0021.710.510x/0.45; measured with DZNE-IP power meter
89
24/07/201915:0021.20.50.485.72.541.3410x/0.45; measured with DZNE-IP power meter
90
25/07/201910:1521.90.750.485.352.851.110x/0.45; measured with DZNE-IP power meter
91
26/07/201910:1522.210.495.332.851.0610x/0.45; measured with DZNE-IP power meter
92
29/07/201910:0022.410.244.92.550.6110x/0.45; measured with DZNE-IP power meter
93
30/07/201914:3021.94.50.214.472.320.510x/0.45; measured with DZNE-IP power meter
94
31/07/201910:302210.134.972.720.7310x/0.45; measured with DZNE-IP power meter
95
01/08/201914:0022.80.51.534.862.61.41Service visit Robin Dutscheck; exchange PC again - new PC Z6 now working fine; exchange laser fibre and recoupling lasers; laser power measured through 10x/0.45 with Robins power meter
96
02/08/201914:3021.53.51.675.512.411.5710x/0.45; measured with DZNE-IP power meter
97
05/08/201916:3021.55.51.665.272.361.5510x/0.45; measured with DZNE-IP power meter
98
08/08/201914:0021.44.51.675.032.571.5410x/0.45; measured with DZNE-IP power meter
99
2019-08-1410:0020.711.765.152.931.410x/0.45; measured with DZNE-IP power meter
100
2019-10-2110:4520.70.51.635.072.491.2510x/0.45; measured with DZNE-IP power meter