A | B | C | D | E | F | G | H | I | J | K | L | M | |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | |||||||||||||
2 | |||||||||||||
3 | 405 | 488 | 561 | 639 | |||||||||
4 | installation value | 3.2 | 11 | 4.7 | 2.4 | measured by Michael Pulla through 40x/0.95 | |||||||
5 | date (type yyyy-mm-dd) | time | temperature (at laser) | running hours | mW | mW | mW | mW | comments | ||||
6 | 31/05/2012 | 1:30 pm | 0.5 | 1.02 | 7.23 | 2.25 | 1.73 | borrowed "new" lpm from MPI LMF | still measured through 10x/0.45 | ||||
7 | 13/06/2012 | exchange AOTF controller (due to some troubles and as it was not sure if we have the most up to date version since there was no sticker) | |||||||||||
8 | 13/06/2012 | 0.72 | 7.2 | 2 | 1.3 | measured with Raik Staeber's lpm (PT9610) - low, another laser tuning needed | |||||||
9 | 21/06/2012 | 2.5 | 8.2 | 3.2 | 1.9 | measured with Raik Staeber's lpm (PT9610), after fine laser coupling | |||||||
10 | 25/07/2012 | permanent exchange HRm with new MRm (by Raik Staeber), due to ZEN-HRm-read out problem. results in 3000,- "Gutschrift" | |||||||||||
11 | 30/07/2012 | exchange internal (Observer) RL shutter; exchange power supply PC by Raik Staeber | |||||||||||
12 | 15/10/2012 | 2:30 am | 1.7 | 8.07 | 3.4 | 1.4 | Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||||
13 | 19/11/2012 | 10:00 am | 0.5 | 1.5 | 7.6 | 3.6 | 0.9 | Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | |||||
14 | 03/12/2012 | 10:40 am | 1 | 1.4 | 8.5 | 4.25 | 0.96 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | |||||
15 | 10/12/2012 | 10:00 am | 0.75 | 1.1 | 8.35 | 4.2 | 0.75 | ||||||
16 | 17/12/2012 | Raik Stäber - laser recoupling | |||||||||||
17 | 17/12/2012 | 1:00 pm | 2h | 2,2 | 9,0 | 4,2 | 2,0 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | |||||
18 | 08/01/2013 | 11:50 am | 0,5 | 2,25 | 9,22 | 4,2 | 2,1 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | |||||
19 | 11/02/2013 | 10:40 am | 22 | 1.5 | 1.85 | 9.02 | 4.5 | 2.16 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
20 | 25/02/2013 | 10:55 am | 22 | 2 | 1.61 | 8.8 | 4.5 | 2.19 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
21 | exchange PC by Raik Staeber | ||||||||||||
22 | 11/03/2013 | 10:20 am | 22 | 1 | 1.71 | 8.64 | 4.39 | 2.13 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
23 | 22/04/2013 | 10:45 am | 22 | 1.5 | 1.46 | 8.7 | 4.5 | 2.01 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
24 | 06/05/2013 | 10:30 am | 22 | 0.75 | 1.57 | 8.62 | 4.42 | 1.89 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
25 | 29/05/0213 | 10:05 am | 22 | 1 | 1.7 | 8.9 | 4.66 | 1.89 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
26 | 12/06/2013 | 10:50 am | 22 | 1 | 1.54 | 8.76 | 4.42 | 1.6 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
27 | 24/06/2013 | 11:20 am | 22 | 2 | 1.33 | 8.06 | 4.39 | 1.22 | 20/0,8; Silke+Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
28 | 01/07/2013 | 11:10 am | 22 | 1 | 1.621 | 7.8 | 4.27 | 1.55 | 20/0,8; Silke+Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
29 | 15/07/2013 | 10:40 am | 22 | 2 | 1.5 | 7.8 | 4.38 | 1.39 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610), air condition not working properly --> hot air is blown into the room (Zuluftschacht), temp. control set on 22 | ||||
30 | 29/07/2013 | 11:00 am | 22 | 2 | 1.2 | 8.4 | 4.7 | 1.12 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
31 | 13/08/2013 | 10:15 am | 22 | 1 | 1.2 | 8.3 | 4.59 | 1.24 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
32 | 02/09/2013 | 10:30 am | 22 | 0.5 | 1.5 | 7.88 | 4.35 | 1.86 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
33 | 16/09/2013 | 11:11 am | 22 | 0.5 | 1.19 | 7.9 | 4.28 | 1.67 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
34 | 01/10/2013 | 10:20:00 | 22 | 0.5 | 1.429 | 8.39 | 3.95 | 2.05 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
35 | 21/10/2013 | 17:00:00 | 22 | 0.75 | 1.529 | 7.86 | 4.25 | 1.64 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
36 | 06/11/2013 | 11:30:00 | 22 | 0.75 | 1.6 | 8.45 | 4.22 | 2.28 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
37 | 25/11/2013 | 10:30:00 | 22 | 2 | 1.62 | 8.4 | 4.5 | 2.3 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
38 | 09/12/2013 | 12:00:00 | 22 | 2 | 1.41 | 7.97 | 4.4 | 2.22 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
39 | 20/01/2014 | 10:50:00 | 22 | 1.5 | 1.48 | 8.37 | 4.54 | 2.36 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
40 | 03/02/2014 | 10:40:00 | 22 | 0.75 | 1.38 | 8.23 | 4.45 | 2.3 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
41 | 18/02/2014 | 10:50:00 | 22 | 0.5 | 1.41 | 7.89 | 4.23 | 2.26 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
42 | 10/03/2014 | 10:45:00 | 22 | 0.5 | 1.33 | 7.27 | 4.07 | 2.09 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
43 | 24/03/2014 | 10:45:00 | 22 | 2 | 1.34 | 7.58 | 4.19 | 2.84 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
44 | 07/04/2014 | 11:00:00 | 22 | 1 | 1.462 | 8.22 | 4.42 | 2.18 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
45 | 28/04/2014 | 10:50:00 | 22 | 1 | 1.437 | 7.935 | 4.12 | 1.81 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
46 | 19/05/2014 | 11:15:00 | 22 | 0.75 | 1.42 | 8.03 | 4.186 | 2.12 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
47 | 02/06/2014 | 11:10:00 | 22 | 1 | 1.32 | 7.65 | 4.2 | 1.95 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
48 | 16/06/2014 | 11:00:00 | 22 | 1 | 1.29 | 7.49 | 4.16 | 1.79 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
49 | 07/07/2014 | 11:00:00 | 22 | 1 | 1.13 | 8.33 | 4.17 | 1.75 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
50 | 28/07/2014 | 11:30:00 | 22 | 0.75 | 0.94 | 7.22 | 4.5 | 1.02 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
51 | 01/09/2014 | 11:30:00 | 22 | 1 | 1.044 | 7.067 | 4.305 | 1.2 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
52 | 22/09/2014 | 11:00:00 | 22 | 2 | 0.52 | 7.02 | 4.32 | 1.2 | 20/0,8; Isa, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
53 | 02/10/2014 | 11:10:00 | 22 | 1 | 1.14 | 7.27 | 4.5 | 1.7 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
54 | Jahreswartung | ||||||||||||
55 | 26/11/2014 | 1.82 | 9.7 | 4.52 | 2.24 | Raik, Jahreswartung, open laser module, measured with raik powermeter | |||||||
56 | 1.3 | 8.45 | 4.43 | 2.36 | Raik, Jahreswartung, closed laser module, measured with raik powermeter | ||||||||
57 | 08/12/2014 | 11:00:00 | 22 | 0.5 | 1.1 | 7.7 | 2.9 | 2.1 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
58 | 22/01/2015 | 11:00:00 | 22 | 1 | 1 | 8.24 | 4.36 | 2.39 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
59 | 30/03/2015 | 11:00 | 22 | 2 | 0.65 | 7.8 | 3.65 | 2.02 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
60 | 29/04/2015 | 10:30 | 22 | 0.5 | 1.12 | 8.2 | 3.6 | 2.35 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
61 | 01/06/2015 | 10:00 | 22 | 1 | 0.87 | 8.18 | 4.3 | 2.26 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
62 | 23/06/2015 | 10:30 | 22 | 1 | 0.9 | 7.6 | 4.2 | 2.18 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
63 | Jahreswartung Raik Stäber | ||||||||||||
64 | 12/10/2015 | 14:00 | 22 | 3 | 0.7 | 7.6 | 4 | 1.8 | 20/0,8; Raik Stäber, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
65 | 19/04/2016 | 11:00 | 22 | 2 | 0.49 | 6.48 | 3.52 | 1.82 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
66 | 07/07/2016 | 10:30 | 23 | 0.5 | 0.72 | 6.2 | 3.75 | 1.96 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
67 | 06/09/2016 | 14:30 | 23 | 0.5 | 0.65 | 5.74 | 3.54 | 1.78 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | ||||
68 | 25/04/2017 | 11:30 | 23 | 2.5 | 0.28 | 6.09 | 3.47 | 1.57 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610) | service ordered for 405nm | |||
69 | 02/05/2017 | Raik Stäber - laser recoupling (405nm laser coupling was very far off) | |||||||||||
70 | 02/05/2017 | 1.2 | 5.33 | 3.2 | 1.8 | 10x/0.45; Raik measured with DZNE-IP measure meter | |||||||
71 | 08/05/2017 | 09:30 | 23 | 0.5 | 0.61 | - | - | - | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610), only 405nm switched on | ||||
72 | 08/05/2017 | 09:30 | 23 | 0.5 | 0.488 | 5.51 | 3.1 | 1.9 | 20/0,8; Silke, measured with new IP-measuremeter (Gigahertz PT 9610), all lasers switched on | ||||
73 | 13/06/2017 | Raik Stäber - exchange AOTF | |||||||||||
74 | 13/06/2017 | 2 | 6.66 | 3.3 | 1.65 | 10x/0.45; Raik measured with DZNE-IP measure meter | |||||||
75 | 11/10/2018 - 19/10/2018 | Move into new building: Spedition Fiedler. responsible technicians: Daniel Sommer: M1-LSM780NLO + general responsibility for move Raik Staeber: SD1-Spinning Disk + SD2-Spinning Disk incubation Robin Dutscheck: C1-LSM700, W1-ApoTome.2 first floor, W2-ApoTome.2 | |||||||||||
76 | |||||||||||||
77 | |||||||||||||
78 | |||||||||||||
79 | air conditioning not working yet in new rooms, very hard to couple in lasers properly, impossible for 405nm. Raik is going to come back for that | ||||||||||||
80 | 21/02/2019 | 14:00 | 22.3 | 0.5 | 1.8 | 5.8 | 2.61 | 1.52 | 10x/0.45; measured with DZNE-IP measure meter | ||||
81 | 2019-06-24 | 10:00 | 1 | 0.53 | 6.08 | 4.33 | 3.23 | 10x/0.45; measured with DZNE-IP power meter | |||||
82 | 2019-07-03 | 0.8 | 10x/0.45; measured with DZNE-IP power meter | ||||||||||
83 | 2019-07-04 | In front of laser | 42 | 29.5 | 43 | 97 | Service visit Robin Dutscheck; exchange PC - not working, Robin left loan PC here; recoupling 405nm; laser power measured through 10x/0.45 with Robins power meter | ||||||
84 | Out of fibre | 17.4 | 10 | 22 | 54 | ||||||||
85 | Out of objective | 1.65 | 5.6 | 2.75 | 1.8 | ||||||||
86 | 1.73 | 6.1 | 2.9 | 1.6 | 10x/0.45; measured with DZNE-IP power meter | ||||||||
87 | 22/07/2019 | 11:00 | 22.4 | 1 | 0.22 | 5.7 | 2.5 | 1.8 | 10x/0.45; measured with DZNE-IP power meter | ||||
88 | 23/07/2019 | 10:00 | 21.7 | 1 | 0.5 | 10x/0.45; measured with DZNE-IP power meter | |||||||
89 | 24/07/2019 | 15:00 | 21.2 | 0.5 | 0.48 | 5.7 | 2.54 | 1.34 | 10x/0.45; measured with DZNE-IP power meter | ||||
90 | 25/07/2019 | 10:15 | 21.9 | 0.75 | 0.48 | 5.35 | 2.85 | 1.1 | 10x/0.45; measured with DZNE-IP power meter | ||||
91 | 26/07/2019 | 10:15 | 22.2 | 1 | 0.49 | 5.33 | 2.85 | 1.06 | 10x/0.45; measured with DZNE-IP power meter | ||||
92 | 29/07/2019 | 10:00 | 22.4 | 1 | 0.24 | 4.9 | 2.55 | 0.61 | 10x/0.45; measured with DZNE-IP power meter | ||||
93 | 30/07/2019 | 14:30 | 21.9 | 4.5 | 0.21 | 4.47 | 2.32 | 0.5 | 10x/0.45; measured with DZNE-IP power meter | ||||
94 | 31/07/2019 | 10:30 | 22 | 1 | 0.13 | 4.97 | 2.72 | 0.73 | 10x/0.45; measured with DZNE-IP power meter | ||||
95 | 01/08/2019 | 14:00 | 22.8 | 0.5 | 1.53 | 4.86 | 2.6 | 1.41 | Service visit Robin Dutscheck; exchange PC again - new PC Z6 now working fine; exchange laser fibre and recoupling lasers; laser power measured through 10x/0.45 with Robins power meter | ||||
96 | 02/08/2019 | 14:30 | 21.5 | 3.5 | 1.67 | 5.51 | 2.41 | 1.57 | 10x/0.45; measured with DZNE-IP power meter | ||||
97 | 05/08/2019 | 16:30 | 21.5 | 5.5 | 1.66 | 5.27 | 2.36 | 1.55 | 10x/0.45; measured with DZNE-IP power meter | ||||
98 | 08/08/2019 | 14:00 | 21.4 | 4.5 | 1.67 | 5.03 | 2.57 | 1.54 | 10x/0.45; measured with DZNE-IP power meter | ||||
99 | 2019-08-14 | 10:00 | 20.7 | 1 | 1.76 | 5.15 | 2.93 | 1.4 | 10x/0.45; measured with DZNE-IP power meter | ||||
100 | 2019-10-21 | 10:45 | 20.7 | 0.5 | 1.63 | 5.07 | 2.49 | 1.25 | 10x/0.45; measured with DZNE-IP power meter |