Contact Aligner (MA-6)
 Share
The version of the browser you are using is no longer supported. Please upgrade to a supported browser.Dismiss

View only
 
 
ABCDEFGHIJKLMNOPQRSTUVWXYZAA
1
DateName
Lamp Checked/Calibrated
% UniformityLamp power @10mW in CPWEC head levelComments
2
12/3/2018Carmen9.6
3
1/9/2019Carmen/Greg9.8/10<5%315 W
4
1/28/2019Greg9.0/10<5%300 W
Defocused lamp after shutdown
5
2/4/2019Greg9.7/10<2%308 W*Before Jude's visit
6
2/27/2019Carmen9.0/10<5%300 W
7
3/29/2019Carmen10.0/10<2%300 W
8
4/5/2019Carmen10.1/10<5%300 W
9
5/3/2019Carmen10.0/10<5%300 W
10
5/22/2019Greg/Carmen9.4/10.0<5%300 W
11
5/28/2019Carmen10.0/10<5%300 W
12
6/6/2019Carmen10.0/10<5%300 W
13
6/11/2019Carmen10.0/10<5%300 W
14
6/17/2019Carmen10.0/10<5%300 W
15
7/1/2019Carmen10.1/10<5%300 W
16
7/10/2019Carmen10.0/10<5%300 W
17
7/17/2019Carmen10.1/10<5%300 W
18
8/1/2019Carmen9.9/10.0<5%300 W
19
8/7/2019Carmen9.8/10.0<5%300 W
20
9/4/2019Greg9.6/10<5%300 W
21
9/9/2019Carmen10.0/10<5%300 W
22
9/18/2019Carmen10.1/10.0<5%300 W
23
9/23/2019Carmen10.1/10<5%300 W
24
10/14/2019Carmen10.1/10<5%300 W
25
10/28/2019Carmen10.2/10<5%300 W*Lamp changed 10/23/19
26
11/5/2019Carmen10.2/10<5%300 W
27
11/11/2019Carmen9.8/10<5%300 W
28
12/4/2019Carmen9.6/105%300 W
*Lamp will be turned off on 12/6 so it will be recalibrated before 12/9
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
Loading...