3/15/2017 17:53:27Scott Kurdilla3/15/20175:25:00 PM170315161217CaganSubject
Image Recon error in between Multi Band and T1 MPRAGE
MPRAGENoYesReboot ImagerRe-occuring ErrorYesNo
Second reboot of the imager
Siemens up time notified
3
3/22/2017 11:09:50Tim Keller3/22/201710:00:00 AMNonescannerPhantom
Standard scanner boot up error: unknown software exception (0xc0000005) occurred in the application at location 0x7d620dee.
system bootNoYesClick OKRe-occuring ErrorNoNoClick OK
Recurring error that I just want documented. Seems to have no effect on anything, but I've never know what causes it or if it was the same as the one we got before the software reinstall.
4
4/10/2017 11:49:33Scott Kurdilla4/10/201711:17:00 AM170410102824EackSubjectMPCU Board temp overheatEPIYesNo
Full Scanner Reboot, Call Siemens Uptime
New ErrorYeswaiting on Siemensbreaker reset
5
4/10/2017 15:49:18Scott Kurdilla4/10/20172:24:00 PMqaqaPhantomMPCU temp overheatMBNoNo
Full Scanner Reboot, Call Siemens Uptime
Re-occuring ErrorNoMaybe
Cabinet overheating caused by acs fan breaker being tripped. Reset of breaker cooled cabinet to proper temperatures. Upon
6
4/11/2017 10:17:16Scott Kurdilla3/9/20178:00:00 AMunknownunknownSubjectImage reconstruction errorsMBYesNo
Reboot Imager, Full Scanner Reboot
Re-occuring ErrorYesNo
Collected SaveLogs. Uploaded to servers for analysis.
7
4/11/2017 10:23:52Scott Kurdilla3/6/20178:00:00 AMunknownunknownPhantomAdvance shim failureMBNoYes3D shimNew ErrorYesNo
Installed supply imager for testing purposes. Collected SaveLog for uploading to our servers. Continuing to monitor
8
4/11/2017 10:30:08Scott Kurdilla3/3/20178:00:00 AMunknownsiemensPhantomimage reconMBNoYesCall Siemens UptimeRe-occuring ErrorNoMaybe
Replaced questionable network switch
9
4/11/2017 10:32:08Scott Kurdilla3/1/20178:00:00 AMunknownunknownPhantomimage reconMBYesNosoftware installRe-occuring ErrorNoMaybe
Reloaded software on mrc from image backup. Performed all updates installed since image was created
10
4/13/2017 17:07:35Scott Kurdilla4/13/20174:38:00 PMunknownTarrSubject
Recon Imager crashed / communication error
fieldmapYesNoFull Scanner RebootRe-occuring ErrorYesNo
Siemens replacing imager next day
11
4/19/2017 18:07:39Scott Kurdilla4/14/20176:20:00 PM?tarrSubjectimager crash/scanner power lossfieldmapYesNo
Reboot Imager, Full Scanner Reboot, Call Siemens Uptime
Re-occuring ErrorYesYes
Receiver board #1 failed diags on modualtor and imager. Board replaced
did not fix advance shim
12
4/28/2017 12:11:33Scott Kurdilla4/28/20179:00:00 AM?stabilityPhantom
Shim Error. No communication with unit shim
allNoYesFull Scanner RebootNew ErrorYesSam coming inunknown
13
5/4/2017 18:02:11Scott Kurdilla5/3/201710:10:00 AM170503092242AndersonSubjectImage ReconMBYesYesRe-start ScanRe-occuring ErrorNoMaybeReboot
14
5/11/2017 15:45:17Scott Kurdilla5/11/20173:30:00 PM170511145424VerstynenSubjectRecon errorMBYesYes
Re-start Scan, Reboot Imager
Re-occuring ErrorNoNo
15
6/21/2017 15:17:52Scott Kurdilla6/21/20172:40:00 PM?BehrmannSubjectImager not respondingLocNoYesFull Scanner RebootNew ErrorNoYes
16
7/12/2017 15:54:41Scott Kurdilla7/11/20174:45:00 PM170710160509HafemanSubjectMB Recon ErrorMultiBand 3YesYesRe-start ScanRe-occuring ErrorNoMaybeStarted scan again
17
11/1/2017 13:26:03Scott Kurdilla10/31/201712:00:00 PM171031115526ManelisSubjectPrep of image Recon failedMB4NoYesFull Scanner RebootRe-occuring ErrorNoYesReboot
18
11/21/2017 12:27:42Scott Kurdilla11/21/201711:20:00 AM171121101632VerstynenSubjectscanner stuck in preparing modefieldmapNoNo
Re-start Scan, Reboot Imager
New ErrorYesYesFull scanner reboot
19
11/28/2017 11:57:24Scott Kurdilla11/28/201710:30:00 AM171128102033CreswellSubject
DICO adjustment did not converge
LocNoYesFull Scanner RebootRe-occuring ErrorNoYesReboot
20
12/8/2017 15:58:44Scott Kurdilla12/8/201710:15:00 AM171208092907tarrSubjectreshimMultibandNoYesreshimmedNew ErrorYesNo
Don't know why scanner reshimmed locations were the same
21
12/13/2017 10:01:42Scott Kurdilla12/13/20179:45:00 AM171213094711TarrSubjectMSUP warning (sleep) modeMBYesYesRe-start ScanRe-occuring ErrorNoYes
Did not happen again during scan session
22
3/19/2018 14:18:50Scott Kurdilla3/19/20182:00:00 PMstabbehrmannPhantom
Image recon failed, coil sensitivity did not converge, RF infra structure error, watchdog timeout error
All scans using 32 coilYesNo
Re-start Scan, Reboot Imager, Full Scanner Reboot, Call Siemens Uptime
Re-occuring ErrorNoNoSiemens on way
23
3/29/2018 11:43:00Scott Kurdilla3/29/20189:45:00 AM180329095916EricksonSubject
RF-Infrastructure error sleep mode coil error
EPIYesYes
Call Siemens Uptime, Switched to new 32 channel head coil
Re-occuring ErrorYesYescoil change
Siemens will follow up on April 3rd
24
4/23/2018 14:11:34Scott Kurdilla4/23/20189:30:00 AMstabilitystability qaPhantom
RF Safety Watchdog Error. Hardware: Power Absorption Limit self test failed
boot upNoNoFull Scanner RebootNew ErrorNoYes
Booting up system with no coils plugged in. Tales replaced on Friday April 21st and on initial 2 boot ups it was stuck in high power freq mode due to tales and got confused with head coil. Once all coils were unplugged and system restarted it came up fine.
25
5/1/2018 15:12:09Scott Kurdilla5/1/20182:11:00 PM180501131908ManelisSubject
RF Safety Watchdog error: timeout for energy events. Scanner comm error
DTIYesYesRestart SyngoRe-occuring ErrorYesMaybeEngineer to follow up
Siemens service call #400108800493
26
8/10/2018 15:00:08Tim Keller8/7/20182:00:00 PM180730_bvdwiQAstabilityPhantom
RF amp fault on high b-value DWI scans
cmrr_epse_V80_9b0_b2000_2x2x2 PA and AP
YesNo
Re-start Scan, Full Scanner Reboot, Reshuffling 80 vector table. Full scanner shut down, 15-min wait before turning system on.
Re-occuring ErrorNoMaybe
After a full shut down and boot, couldn't duplicate
At first it was consistently happening on vector 46, so I thought maybe I just needed to reorder the vector table. After reordering table it started to happen on random vectors before and after 46. Did a reboot of the scanner without bringing system down. Still got error. After full system off / 15-min wait / system on have had no problem with the scans since despite multiple attempts to trigger it.
27
2/15/2019 17:51:22Scott Kurdilla2/15/20194:58:00 PM190215151530Behrmann/TarrPhantomTalesIn between subjectsNoYesRe-start ScanNew ErrorNoYes
28
further fa n inspections, found 2 fans out in acs pull out cage. These will be ordered and replaced
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100