Handheld Digitial Multi-Meter Transient Test Results
2
3
Round I Tests using first generation transient generator
4
ManufactuerModelCostRatingResRMSTmpnSmVuANCVAC+DCMax DC (1KV / 600V)Rectified 220V Line Piezo Grill IgnitorESD Gun, 4KV2.0KV, 49 ohm 2.0KV, 14 ohm2.0KV, 2 ohm2.8KV, 2 ohm, 303.2KV, 2 ohm3.7KV, 2 ohm5.0KV, 2 ohm, 405.8KV, 2 ohm, 6.5KV, 2 ohm, 50us FWHH 12KV, 2 ohm, 50us FWHH13KV, 2ohm, 100us FWHH
5
Amprobe AM510PassPassPassPassPassPassPassPassFail
6
BK 2703CPassPassPassFail
7
Cen-Tech/Mastech
P98674/MS8229
NTNTFail
8
ExtechMN16APassPassPassFail
9
Fluke101PassPassPassPassPassPassPassPassPassPassPassPass
10
Klien ToolsMM500PassPassPassPassPassPassFail
11
MastechMS8264Fail
12
UNI-T UT90APassPassPassPassPassPassFail
13
UNI-T 139CNTNTNTNTNTFail
14
Gardner Bender
GDT-311Fail
15
SearsDM-301PassFail
16
Cen-Tech98025Fail
17
Fluke87VNTNTNTNTNTNTNTNTNTNTFail
18
Cen-Tech P37772PassPass
19
20
Round II Tests using second generation higher energy transient generator
Original Generator Final Design
21
ManufactuerModelCostRatingResRMSTmpnSmVuANCVAC+DCMax DC (1KV / 600V)Rectified 220V Line Piezo Grill IgnitorESD Gun, 4KV1.0KV, 2 ohm, 100us FWHH1.5KV, 2 ohm2.0KV, 2 ohm2.5KV, 2 ohm3.0KV, 2 ohm4.0KV, 2 ohm5.0KV, 2 ohm5.9KV, 2 ohm6KV, 2 ohm, 50us FWHH 8KV, 2 ohm, 50us FWHH 10KV, 2 ohm, 50us FWHH 12KV, 2 ohm, 50us FWHH 14KV, 2 ohm, 50us FWHH 15KV, 2 ohm, 50us FWHH 13KV, 2ohm, 100us FWHH14KV, 2ohm, 100us FWHH15KV, 2ohm, 100us FWHH
22
UNI-T UT90A(repaired)$30 C3/6002KPassPassPassPassFail
23
Klein MM2000PassPassPassPassPassFail
24
UNI-T UT61DFail
25
AmprobeAM530$80 C3/6004KX1XXPassPassPassPassFail
26
RadioShack2200087PassPassPassPassPassPassPassPassPassPassPassNTNTNTNTFail
27
UNI-TUT61EFail
28
Fluke107$110 C3/6006K ACPassPassPassPassPassPassPassPassPassPassPassPassNTNTNTNTPassPassFail
29
TekPowerTP40PassPassPassPassFail
30
TekPower TP2844RPassFailPass
31
BK2705BPassPassPassFail
32
CircuitTestDMR-6550PassPassPassFail
33
UTLUTLDM 2PassPassPassFail
34
SouthWire12070TPassPassPassFail
35
HoldPeakHP-760HPassPassPassPassPassPassPassPassPassFail
36
Innova3320PassPassPassFail
37
ExtechEX430Fail
38
Brymen BM869S$230 C4/1K50KX2XXXXPassPassPassPassPassPassPassPassPassPassPassPassNTNTNTNTNTFail
39
Brymen (2nd Unit)
BM869SNTNTNTNTNTNTNTNTNTNTNTPassPassPassFail
40
Fluke (105xxxxx)
87V6KX1XXXNTNTNTNTPassFail
41
Fluke (105xxxxx)
87V (repaired)6KX1XXXPassPassNTNTPassPassPassPassPassPassPassPassReplace all diodes used for low voltage clamp, BAV199, SOD123 5237
42
Fluke (409xxxxx)
87V6KX1XXXPassPassPassPassPassPassPassPassPassPassPassPassPassPassPassFail
43
UNI-TUT139CPassPassPassPassPassPassPassPassPassFail
44
AmprobeAM510$50 4KXPassNTNTNTNTNTNTPass
45
Fluke115$132 C3/6006KXXPassPassPassPassPassPassPassPassPassPassPassPassPassPassPassPass
46
GreenleeDM-20$30 2KPassPassPassPassPassPassFail
47
UNI-TUT15CPassPassPassPassPassPassPassPassPassPassPassFail
48
KeysightU1231APassPassPassPassPassPassPassPassPassFail
49
ViciVC99PassFail
50
HiokiDT4252$150 C4/6006KXXPassPassPassPassPassPassPassPassPassPassPassPassPassPassPass*Pass*Pass** Added plastic insulator around insert near fuse to prevent breakdown
51
UNI-TUT181A$300 C4/60060KX2XXXXPassPassFail
52
UNI-T Modified*UT181A PassPassPassPassPassNTPassNTNTPassNTPassNTPassNTNTNTPass
* Meter was modified to handle higher transients (including ESD) without damage
53
UNI-T UT10A C2/300PassPassPassPassPassPassPassPassFail
54
UNI-T UT20BC2/300PassFailPass
55
VICTOR VC921C2/300PassPassPassPassFail
56
TPI/SUMMITTIP 194 II$250 C4/60050KX1XXXXPassFail
57
UNI-T Modified*UT61E C4/60020KXXXPassPassPassPassPassPassPassNTPassPassPassPass
* Meter was modified to handle higher transients (including ESD) without damage
58
Brymen
BM235 (EEVBLOG)
$130
C4/300, C3/600
6KX1XXXPassPassPassPassPassPassPassPassPassPassPassPassPassPassPass
59
Mastech MS8211D$20 C3/6002KPassPassPassPassFail
60
MastechMS8211$20C3/6002KPassFailPass
61
CEMDT-9939$120C4/60040KX1XXXPassPassPassPassPassPassPassFail
62
WoodsDMMW3$30C3/3004KPassPassPassFail
63
KASUNTEST
ZT102 / AN8002
$20C3/6006KX1XPassPassPassPassPassPassPassPassFail
64
KASUNTESTKT6000$37C4/6006KX1XXPassPassPassPassPassPassPassPassPassFail
65
Gossen Metrawatt
M248B$850C4/300
300K
X1XXXPassPassPassPassPassPassPassPassPassPassPassPassPassPassPassPass
66
ANENGAN8008$23C3/60010KXXXPassPassPassPassPassPassFail
67
BrymenBM319s$130
C3/600, C4/300
6K1XXPassPassNTPassNTNTPassNTNTPassNTPass
68
BrymenBM839
C3/1000, C4/600
6/20K
X1XXXPassPassNTPassPassPassPassPassPassPassPassPassPassPassPassFail
69
METERKMK01A$9C4/6006KX1XXXPassPassFail
70
Brymen/AMPROBE
BM27s/PM55A$55
C2/600, C3/300
6KXXPassPassPassPassPassPassPassPassPassPassPassPassPassFailMeter was not electrically damaged. More testing is needed.
71
OWONB41T$70C3/100020KX1XXPassFailPassPassMeter was repaired in order to run EDS testing.
72
YokagowaTY720$500
C3/1000, C4/600
50KX1XXXPassPassPassPassPassPassPassPassPassPassPassPassFail
Meter was repaired in order to run EDS testing.