HL7
Error Code
HL7 Error DescriptionResolutionExamplePumpGatewayDrug LibraryF2F Proposed CategoryWG discussion KurtAl
2
8801CCA limit exceededEnsure that the CCA is correct and that the rate and patient weight are appropriateXXX
3
8802Infuser limit exceededEnsure rate is < 999 mL/hrXSame as 9014?
4
8803A different medication is currently programmedStop the pump and resubmit smart pump programX Related to 9002, in this case it may not be infusing, just programmed
5
8804Infuser is partially programmedClear the current program then resubmit smart pump programX
6
8805Line is in standbyCancel standby and resubmit smart pump programXXSame as 9032? delayed start = standby?
7
8806Patient BSA outside of allowable rangeEnsure ordered patient BSA is appropriate for the current CCAXXB Braun reports as "9005^Parameter (patient BSA) outside of allowable range"
8
8807Patient height outside of allowable rangeEnsure ordered patient height is appropriate for the current CCAXXB Braun reports as "9005^Parameter (patient height) outside of allowable range"
9
8808Patient weight outside of allowable rangeEnsure ordered patient weight is appropriate for the current CCAXXB Braun reports as "9005^Parameter (patient weight) outside of allowable range"
10
8809Dose is invalidThe ordered dose cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XX
11
8810Program type is not supported (possible cause: infusion cannot be interrupted by piggyback)Loading Dose, Multistep and Bolus require manual programmingXXB Braun reports primary doesn't allow secondary as "9030^Current medication cannot be interrupted"
12
8811Rate on Line A and Line B must each be at least 0.5 mL/hr and the total of rates on both lines cannot exceed 500 mL/hCombined rates of Lines A and B must be between
0.5 and 500 mL/hr
X
13
8812Unable to program piggyback on primary lineSelect Line B and resubmit smart pump program.XXB Braun reports as "10012^Infusion cannot be programmed as primary"
14
8813Ordered patient BSA does not match BSA currently programmed on pump (clear program on pump then try again)Pump is currently using a different BSA value. Clear the current program then resubmit smart pump program.XB Braun reports as "10002^Patient BSA mismatch"same as 10002
15
8814Ordered patient height does not match height currently programmed on pump (clear program on pump then try again)Pump is currently using a different patient height value. Clear the current program then resubmit smart pump program.XB Braun reports as "10001^Patient height mismatch"same as 10001
16
8815Default drug library installed on infuserReturn pump to Biomed for drug library downloadX
17
8816Unable to change rate or titrate a completed infusionProgram a new bag or increase VTBI manually.X
18
8817Rate change program is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XB Braun's "10017^Program is identical" may be one reason for an invalid update orderMore generic condition than 8818
19
8818Unable to change rate or titrate an unknown medicationNot allowed when medication is not in the pump drug library. Program pump manually.XPump is running a manually programmed infusion outside of the library Similar to 8817 but more specific.
20
8819Rate change or titration cannot specify a durationProgram pump manually. Review medication order and pump drug library and correct as needed.XB Braun reports as "10003^Dose over time infusion can only be modified by manual programming"Applies to dose and dose over time infusions
21
8820Keypad is currently lockedUnlock keypad and resubmit smart pump program.XSimilar to 10018
22
8821Rate unit is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XXXB Braun reports as "9004^Invalid program parameter (Give Rate Units)" [but may be changing to use 9008]
23
8822Programming not completePrevious program was not completely cleared. Resubmit smart pump program.XSame as 8804
24
8823Secondary program rejected because primary infusion is completePrimary line must be active (infusion not complete) to program secondary.X
25
8824Medication amount unit is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XX
26
8825Unable to program as secondaryProgram pump manually.XXB Braun reports as "9013^Unable to program medication as piggyback"same as 9013? Deliver alone mode may not be a piggyback
27
8826CCA change is in progressTo use new CCA, stop the pump and clear the current program(s). Resubmit smart pump program.X
28
8827Smart pump programming not enabledPump gateway is not enabled for smart pump programming.XConnectivity = Unable to send order to pump
29
8828Infuser does not support smart pump programmingReturn pump to Biomed for configuration.XConnectivity = Unable to send order to pump

B Braun reports as "10064^Auto-programming is not enabled on the pump"
30
8829The patient ID is blankCheck sender's interface settingsXB Braun reports as ERR-3=“101^Required field missing^HL70357”, with ERR-5 = “^ID Number (PID-3.1) in Patient Identifier List entry is missing.” but may be changing to "9031^Patient ID is missing or invalid"same as 9031?
31
8830The patient ID is not supportedCheck sender's interface settingsXB Braun reports as "^Unable to determine valid Patient ID." [if PID-3.4 and/or PID-3.5 not recognized]same as 9031?
32
8831Invalid message versionCheck sender's interface settingsXB Braun reports as ERR-3 = 200^Unsupported message type^HL70357 - assumes an HL7 level error eg not a PCD03. 8831 might be a TF version related error after the HL7 is processed. (valid PCD03)
33
8832Malformed smart pump programCheck sender's interface settingsX
34
8833Patient dose lockout period below 5 minute minimumThe order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XX
35
8834Patient dose lockout period exceeds 120 minute maximumThe order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XX
36
8835Invalid maximum dose limit lockout period. Valid hours: 1/4/6/12The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XX
37
8836Missing maximum dose limit or maximum dose limit lockout period (valid hours: 1/4/6/12)The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XX
38
8837Missing patient dose or patient dose lockout period (valid minutes: 5-120)The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XX
39
9001Unknown infuser or channelCheck pump barcode, resubmit smart pump program. If unsuccessful then manually program or return pump to Biomed.e.g., incorrect infuser or channel IDXConnectivityConnectivity = Unable to send order to pump
40
9002Infuser/channel is currently infusingStop the line and resubmit the smart pump program. (possible cause: a different medication is currently programmed)XXUnavailable State
41
9003Missing required program parameter(s) (ParameterName1, ParameterName2, ...)Program pump manually. Review medication order and pump drug library and correct as needed.e.g., Give Amount Minimum (RXG-5) - volume to be infused - is missingXXXNoncompliant
42
9004Invalid program parameter(s) (ParameterName1, ParameterName2, ...)Program pump manually. Review medication order and pump drug library and correct as needed.e.g., volume units are not mLXXXNoncompliant
43
9005Parameter (ParameterName) outside of allowable range (MinValue to MaxValue)Ensure rate and patient weight are appropriate for the current CCA. Rate cannot exceed 999 mL/hr.e.g., ordered rate greater than pump maximumXXXOutside Limits
44
9006Infuser is powered offTurn on pump, manually select CCA and resubmit smart pump programXConnectivityConnectivity
45
9007Infuser is offline or unable to connect to infuserCheck communication icons on pump before resubmitting program, or program pump manually.e.g., infuser not on network or weak wireless signalXConnectivityConnectivity
46
9008Invalid units for parameter (ParameterName)Program pump manually. Review medication order and pump drug library and correct as needed.e.g., Give Strength Volume Units (RXG-24) contains a medication unit value instead of volume unitsXXXNoncompliant
47
9009Dose/Rate Units do not match drug librarye.g., ordered units = mL/hr; drug library units
=mcg/kg/min
XXNonmatch
48
9010Unable to match medication to drug library Check CCA. Review medication order and pump drug library and correct if needed. (possible cause: medication not in CCA)e.g., Medication does not exist in drug libraryXXNonmatch
49
9011Patient weight mismatchPump is currently using a different weight value. Stop the pump, clear the current program then resubmit smart pump program. Ordered patient weight does not match weight currently programmed on pump (clear program on pump then try again)e.g., patient weight known by pump differs from weight sent in [PCD-03]XNonmatch
50
9012Patient ID mismatche.g., patient ID known by pump differs from ID sent in [PCD-03]XXNonmatch
51
9013Unable to program medication as piggybackInfuse in concurrent mode or use a separate pump.e.g., medication not configured for piggyback administration in drug libraryXXUnsupported
52
9014Dose rate or VTBI exceeds maximume.g., greater than pump maximumXXOutside LimitsHow does this differ from 9005?
53
9015Request timed outResubmit smart pump program or program manually.XXConnectivity
54
9016Other errorResubmit smart pump program or program manually.Used when other errors are not applicableXXXMisc.
55
9017Infuser cannot accept program Ensure CCA is selected and resubmit smart pump program. (possible causes: CCA or New Patient not selected on infuser)Infuser is in a state where it cannot accept program; e.g., alarming or in standbyXUnavailable StateAssumes that pump is configured to support auto programming but not in a state to accept it at this time. Temporary condition.
56
9018Parameter does not match drug libraryProgram pump manually. Review medication order and pump drug library and correct as needed.e.g., Give Strength Units (RXG-18) = mg, drug library = mEqXXNonmatch
57
9019Patient weight missingProgram pump manually. Review medication order and interface message and correct as needed.Drug is weight-based or BSA-based, but patient weight OBX is missingXXNoncompliant
58
9020Patient height missingDrug is BSA-based, but patient height OBX is missingXXNoncompliant
59
9021Care area or profile mismatchCare area or profile not cleared on pump; or pump is set to a different care areaXNonmatch
60
9022Requested infusion program is staleThe value of ORC.9 is older than what the IOC will allow to program the pumpXXNoncompliant
61
9023Program rejected by userResubmit smart pump program or program manually.Program rejected or canceled by user prior to starting infusionXMisc.
62
9024Drug library hard limit exceededDose or rate is outside the drug library limit for the medication. Review medication order and pump drug library and correct as needed.e.g., dose exceeds maximum allowable for medicationXXOutside LimitsDIfferent from 9005 in that 9024 is related to library configuration vs pump Max. 9024 could also be that the value is below the minimum limitsHow does this differ from 9005? Because it is a drug library hard limit instead of pump limit?
63
9025Lockout interval missingLockout Interval is required when Patient Dose is presentXXNoncompliant
64
9026Dose limit missingPCA dose limit is required but missingXXNoncompliant
65
9027Patient BSA missingProgram pump manually. Review medication order and interface message and correct as needed.Drug is BSA-based but BSA OBX is missingXXNoncompliant
66
9028Tubing cassette or syringe not installed on pumpInstall and resubmit smart pump program.Used when required by pumpXUnmet Prereqs
67
9029Care area or profile not selectedSelect CCA and resubmit smart pump program.Used when required by pumpXUnavailable State
68
9030Current medication cannot be interruptedInfuse in concurrent mode or use a separate pump.Medication on primary cannot be interrupted by a piggybackXUnavailable State
69
9031Patient ID missing or invalidXXNoncompliant
70
9032Pump is in delayed startCancel delayed start and resubmit smart pump program.XUnavailable State
71
9033Pump is alarmingResolve alarm then resubmit smart pump program.XXUnavailable State
72
9034Not enough fluid or medication in container to deliver bolusApplies to bolus from existing infusion onlyXOutside Limits
73
9035Parent order ID does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatch
74
9036Medication does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatch
75
9037Medication concentration does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatch
76
9038Bolus order type is not supportedORC-1 = “CH” not supported by pumpXXXUnsupported
77
9039Medication does not support bolusMedication in drug library not configured for bolusXXUnsupported
78
9040Bolus dose units do not match the currently programmed ordere.g., “mg” instead of “mL”XXNonmatch
79
9041Unable to parse multistep programApplies to multistep infusion onlyXXNoncompliant
80
9042Medication is not configured for loading doseApplies to multistep infusion onlyXXUnsupported
81
9043Medication does not support multistepApplies to multistep infusion onlyXXUnsupported
82
9044Multistep order type is not supportedApplies to multistep infusion onlyXXUnsupported
83
10000Pump stopped communicating during auto-programmingUnable to determine outcome of programming attempt because the pump stopped communicating
84
10001Patient height mismatchIf an update order and the new height doesn't match the originalXsame as 8814 related to 9011(weight)Add 'X' in Pump column
85
10002Patient BSA mismatchIf an update order and the new BSA doesn't match the originalXsame as 8813 related to 9011(weight)Add 'X' in Pump column
86
10003Dose over time infusion can only be modified by manual programmingXSee 8819, and add 'X' in Pump column
87
10004Syringe pump does not allow a secondaryXSimilar to 8825Add 'X' in Pump column
88
10005Timeout waiting for order reviewUser did not review and accept the order before timeoutXAdd 'X' in Gateway column
89
10006Drug not found in selected care unitXXAdd 'X' in Pump and Drug Library columns
90
10007Secondary orders are only allowed if the primary infusion is programmed as continuousXAdd 'X' in Pump column
91
10008Concentration mismatchConcentration in order does not match concentration(s) in drug library for selected care unitXXAdd 'X' in Pump and Drug Library columns
92
10009Therapy not supportedIntermittent order received for drug name that is defined in drug library as continuous, or vice versaXXAdd 'X' in Pump and Drug Library columns
93
10010Incorrect pump typeOrder received for drug that is in drug library but not enabled for the target pump typeXXAdd 'X' in Pump and Drug Library columns
94
10011Concentration missingOrder received for drug that is defined with a concentration, but the order had rate units of ml/hr and no concentrationXXAdd 'X' in Pump and Drug Library columns
95
10012Infusion cannot be programmed as primaryDrug in order is defined as secondary only, but assignment as primary was attemptedXXRelated to 8812See 8812 (not 8825), and add 'X' in Pump and Drug Library columns
96
10013Patient profile not found in selected care unitOrder received for drug that is in drug library, but is not configured for the patient profile selected by the userXXAdd 'X' in Pump and Drug Library columns
97
10014Dosing unit mismatchUpdate order received for currently programmed drug, but the dosing units in the order do not match the currently programmed dosing unitsXAdd 'X' in Pump column
98
10015Concentration mismatchUpdate order received for currently programmed drug, but the concentration in the order does not match the currently programmed concentrationXAdd 'X' in Pump column
99
10016Therapy mismatchUpdate order received for a running continuous infusion, but the parameters in the order are for an intermittent infusionXAdd 'X' in Pump column
100
10017Program is identicalUpdate order received, but there were no changes to the existing order parametersXSee 9053, and add 'X' in Pump column