_ENUM_GROUPSTOKENNUM_VALUEApplicationDescriptionTags
2
_MDC_PUMP_DELIV_STAT pump-delivery-status-deliveringThe pump is currently delivering the fluid at the programmed rate. ds-ip-rel_2.0 MDC_DEV_PUMP_CURRENT_DELIVERY_STATUSAdded description
3
_MDC_PUMP_DELIV_STAT pump-delivery-status-not-deliveringThe pump is not currently delivering. See _MDC_PUMP_NOT_DELIVERING_REASONds-ip-rel_2.0Added description
4
_MDC_PUMP_DELIV_STAT pump-delivery-status-kvoThe pump is currently delivering the fluid at a programmed KVO rate. ds-ip-rel_2.0Added description
5
_MDC_PUMP_DELIV_STAT pump-delivery-status-flushingThe pump is currently flushing the fluid line. ds-ip-rel_2.0Added description
6
_MDC_PUMP_DELIV_STAT pump-delivery-status-transitioningThe pump is currently tranistioning to a new rate, new device during a relay or infusate source. ds-ip-rel_2.0Added description
7
_MDC_PUMP_DELIV_STAT pump-delivery-status-primingThe pump is currently priming the fluid line. Though it is moving fluid, it is not intended to be delivered to the patient, but rather to fill the line, purge air, etc.ds-ip-rel_2.0new enum for group 8
8
_MDC_PUMP_DOSE_DELAY_INTERVAL_TYPEMDCX_DOSE_DELAY_INTERVAL_TYPEMDC_TIME_PD_DOSE_START_INTERVAL26926157998
9
_MDC_PUMP_DOSE_DELAY_INTERVAL_TYPE MDC_TIME_PCA_LOCKOUT26917157989
10
_MDC_PUMP_MODE pump-mode-multi-step4Deprecated 10MAY21, add code for "not in use"
11
_MDC_PUMP_MODE pump-mode-multi-dosing8Deprecated 10MAY21, add code for "not in use"
12
_MDC_PUMP_MODE pump-mode-ramp-taper9Deprecated 10MAY21, add code for "not in use"
13
_MDC_PUMP_MODE pump-mode-patient-dose10The infusion pump is programmed to allow patient-initiated drug dosing (e.g., PCA). Replaced pump-mode-pca that has the same value.Deprecated 10MAY21, add code for "not in use"
14
_MDC_PUMP_MODE pump-mode-loading-dose14Deprecated 10MAY21, add code for "not in use"
15
_MDC_PUMP_MODE pump-mode-continuous15Deprecated 10MAY21, add code for "not in use"
16
_MDC_PUMP_MODE pump-mode-piggyback16Deprecated 10MAY21, add code for "not in use"
17
_MDC_PUMP_MODE pump-mode-clinician-dose17Deprecated 10MAY21, add code for "not in use"
18
_MDC_PUMP_MODE pump-mode-kvo18Deprecated 10MAY21, add code for "not in use"
19
_MDC_PUMP_MODE pump-mode-flushing19Deprecated 10MAY21, add code for "not in use"
20
_MDC_PUMP_MODE pump-mode-concurrent20Deprecated 10MAY21, add code for "not in use"
21
_MDC_PUMP_MODE pump-mode-priming21The pump is currently priming the fluid line. Though it is moving fluid, it is not intended to be delivered to the patient, but rather to fill the line, purge air, etc.Deprecated 10MAY21, add code for "not in use"
22
_MDC_PUMP_MODE pump-mode-other22The pump is programmed in an operational mode that does not match one of the defined modes.Deprecated 10MAY21, add code for "not in use"
23
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-standby5Pump has been programmed and "paused" for an indeterminate period (possibly with configured max period limit).ds-ip-rel_2.0MDC_DEV_PUMP_NOT_DELIVERING_REASON
24
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-between-doses9Pump has been programmed and started but is in patient-dose mode or multi-dosing mode and not currently delivering fluid.ds-ip-rel_2.0
25
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-switching-source11The pump is momentarily stopped while transitioning between infusate sources.
26
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-primingPump is stopped because it completed priming or was stopped during during priming.ds-ip-rel_2.0Added description
27
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-delayedPump is stopped while waiting for a programmed delay period to expire.ds-ip-rel_2.0Added description
28
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-alarmingPump is stopped due to an alarm condition.ds-ip-rel_2.0Added description
29
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-by-clinicianPump was stopped by a clinician.ds-ip-rel_2.0Added description
30
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-powered-offPump is stopped because it was powered offds-ip-rel_2.0Added description
31
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-not-specifiedPump is stopped due to an unspecified reason.ds-ip-rel_2.0Added description
32
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-flushingPump is stopped because it completed the flush or was stopped during the flush.ds-ip-rel_2.0new enum for group 8
33
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-switching-deviceThe pump is stopped while transitioning between devices during a relay.new enum for group 8
34
_MDC_PUMP_PROGRAM_COMPLETION_MODE pump-program-completion-mode-stopWhen the program has completed the pump will stop.ds-ip-rel_2.0MDC_DEV_PUMP_PROGRAM_COMPLETION_MODE Programmed operation of the pump when a delivery completes (e.g., KVO). {IHE PCD Infusion Pump Work Group }Added description
35
_MDC_PUMP_PROGRAM_COMPLETION_MODE pump-program-completion-mode-kvoWhen the program has completed the pump will continue to deliver but at the kvo rate.ds-ip-rel_2.0Added description
36
_MDC_PUMP_PROGRAM_COMPLETION_MODE pump-program-completion-mode-continueWhen the program has completed the pump will continue to deliver but at the same programmed rate.ds-ip-rel_2.0Added description
37
_MDC_PUMP_PROGRAM_DELIVERY_MODE pump-program-delivery-mode-continuousThe programmed delivery mode for the infusion is a single rate for the duration of the program. The programmed mode could be continuous or intermittent.ds-ip-rel_2.0MDC_DEV_PUMP_PROGRAM_DELIVERY_MODEAdded description
38
_MDC_PUMP_PROGRAM_DELIVERY_MODE pump-program-delivery-mode-ramp-taperThe programmed delivery mode for the infusion is ramp-taper. This a specific type of multi-step where the platteau rate and time period are defined and the pump calculates the duration of each step on the ramp up to and taper down from the plateau.ds-ip-rel_2.0Added description
39
_MDC_PUMP_PROGRAM_DELIVERY_MODE pump-program-delivery-mode-multi-stepThe programmed delivery mode for the infusion is multi-step, where the rate transitions to another rate after a period of time. There may be multiple steps defined, including a loading dose.ds-ip-rel_2.0Added description
40
_MDC_PUMP_PROGRAM_DELIVERY_MODE pump-program-delivery-mode-multi-dosingThe programmed delivery mode for the infusion is multi-dosing, is where the same dose is repeatedly administered from the same container at a fixed interval. There may or may not be infusion activity between the doses.ds-ip-rel_2.0does this need to be deprecated?
41
_MDC_PUMP_PROGRAM_STATUS pump-program-status-not-startedThe infusion pump is programmed but has not yet started the fluid delivery to the patient.ds-ip-rel_2.0MDC_DEV_PUMP_PROGRAM_STATUSAdded description
42
_MDC_PUMP_PROGRAM_STATUS pump-program-status-pausedThe infusion pump is programmed and was started but is not currently actively delivering the fluid to the patient.new enum for group 8
43
_MDC_PUMP_PROGRAM_STATUS pump-program-status-activeThe infusion pump is programmed and is actively delivering the fluid to the patient.ds-ip-rel_2.0Added description
44
_MDC_PUMP_PROGRAM_STATUS pump-program-status-completeThe infusion pump has completed delivery of the fluid to the patient.ds-ip-rel_2.0Added description
45
_MDC_PUMP_PROGRAM_STATUSpump-program-status-not-programmedThe infusion pump is not completely programmed.new enum for group 8
46
_MDC_PUMP_SOURCE_INFO pump-source-info-primaryThe infusion pump is delivering a standard drug dose from the primary reservior.ds-ip-rel_2.0MDC_DEV_PUMP_ENABLED_SOURCES, MDC_DEV_PUMP_ACTIVE_SOURCESAdded description
47
_MDC_PUMP_SOURCE_INFO pump-source-info-secondaryThe infusion pump is delivering a standard drug dose from the secondary reservior.ds-ip-rel_2.0Added description
48
_MDC_PUMP_SOURCE_INFO pump-source-info-pcaThe infusion pump is delivering a patient-initiated bolus drug dosing (e.g., PCA/PCEA)ds-ip-rel_2.0Added description
49
_MDC_PUMP_SOURCE_INFO pump-source-info-loadingThe infusion pump is delivering a loading drug dose.ds-ip-rel_2.0Added description
50
_MDC_PUMP_SOURCE_INFO pump-source-info-clinicianThe infusion pump is delivering a clinician initiated bolus drug dose ds-ip-rel_2.0Added description
51
_MDC_PUMP_SOURCE_INFO pump-source-info-intermittentThe infusion pump is delivering a programmed bolus drug dose ds-ip-rel_2.0Added description
52
_MDC_PUMP_SOURCE_INFO pump-source-info-secondary-piggybackThis term is not needed, can be withdrawn 10MAY21
53
_MDC_PUMP_SOURCE_INFO pump-source-info-primingThe infusion pump is currently priming the fluid line. This fluid is not intended to be delivered to the patient, but rather to fill the line, purge air, etc.new enum for group 8
54
_MDC_PUMP_STAT pump-status-infusingThe infusion pump mechanism is active.ds-ip-rel_2.0MDC_PUMP_INFUSING_STATUSAdded description
55
_MDC_PUMP_STAT pump-status-not-infusingThe infusion pump mechanism is stopped.ds-ip-rel_2.0
56
_MDC_PCA_LOCKOUT_STATUS pca-lockout-patient-doseDeprecated, see pca-lockout-patient-dose-intervalMDC_PCA_LIMIT_LOCKOUT_STATUSis there a related event?
57
_MDC_PCA_LOCKOUT_STATUS pca-lockout-patient-dose-intervalThe PCA patient-dose lockout is in effect because the minimum amount of time that must pass after the last patient dose has been reached.MDC_PCA_LIMIT_LOCKOUT_STATUSis there a related event?
58
_MDC_PCA_LOCKOUT_STATUS pca-lockout-dose-limitThe PCA dose-limit lockout is in effect because the maximum cumulative amount of drug that can be administered in a period of time has been reached.is there a related event?
59
_MDC_PCA_LOCKOUT_STATUS pca-lockout-not-specifiedThe pump is locked out of administering additional doses due to an unspecified reason.is there a related event?
60
_MDC_PCA_LOCKOUT_STATUSpca-lockout-noneThe pump is not locked out of administering additional doses.is there a related event?
61
_MDC_PCA_LOCKOUT_STATUSpca-lockout-max-patient-dosesPatient has received the maximum number of allowable doses in the time period and is locked out from receiving more doses until the time period has expiredis there a related event?
62
_MDC_INFUS_ORDER_TYPEclinician-doseAn infusion order type for a bolus dose in addition to the base infusion that is initiated by a clinician.new enum for group 8, used in TF vol 2 section 3.3.4.4.10.2 Bolus from existing infusion
63
_MDC_INFUS_ORDER_TYPEloading-doseAn infusion order type for a bolus in addition to the base infusion initiated prior to the start of the base infusion.new enum for group 8, used in TF vol 2 section 3.3.4.4.10.2 Bolus from existing infusion
64
_MDC_INFUS_ORDER_TYPEcontinuousAn infusion order type for the base infusion where the fluid or medication is administered at a consistent rate. new enum for group 8, used in TF vol 2 section 3.3.4.4.10.2 Bolus from existing infusion
65
_MDC_INFUS_ORDER_STEP_TYPEloading-doseAn infusion order step initiated prior to a continuous step.new enum for group 8, used in TF vol 2 section3.3.4.4.10.3 Multistep
66
_MDC_INFUS_ORDER_STEP_TYPEcontinuousAn infusion order step where the fluid or medication is administered at a consistent rate for the duration of the infusion step.new enum for group 8, used in TF vol 2 section3.3.4.4.10.3 Multistep