ABCDEFGHIJKLMNOPQRSTUVWXYZAAABACADAE
1
HL7
Error Code
HL7 Error DescriptionResolutionExamplePump
Gateway
Drug Library
F2F Proposed Category
WG Proposed CategoryGroup sortWG Discussion 13MAR23WG discussion through 13FEB23KurtAl
2
8801CCA limit exceededEnsure that the CCA is correct and that the rate and patient weight are appropriateXXXOutside Limits880110.8801
3
8802Infuser limit exceededEnsure rate is < 999 mL/hrXOutside Limits9014.1Same as 9014?
4
8803A different medication is currently programmedStop the pump and resubmit smart pump programXUnavailable State9002.1 Related to 9002, in this case it may not be infusing, just programmed
5
8804Infuser is partially programmedClear the current program then resubmit smart pump programXUnavailable State8822.1
6
8805Line is in standbyCancel standby and resubmit smart pump programXXUnavailable State9032.1Add new code for Pump is in standbySame as 9032? delayed start = standby?
7
8806Patient BSA outside of allowable rangeEnsure ordered patient BSA is appropriate for the current CCAXXOutside Limits9005.1add as new codeB Braun reports as "9005^Parameter (patient BSA) outside of allowable range"
8
8807Patient height outside of allowable rangeEnsure ordered patient height is appropriate for the current CCAXXOutside Limits9005.2add as new codeB Braun reports as "9005^Parameter (patient height) outside of allowable range"
9
8808Patient weight outside of allowable rangeEnsure ordered patient weight is appropriate for the current CCAXXOutside Limits9005.3add as new codeB Braun reports as "9005^Parameter (patient weight) outside of allowable range"
10
8809Dose is invalidThe ordered dose cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXNoncompliant8809
11
8810Program type is not supported (possible cause: infusion cannot be interrupted by piggyback)Loading Dose, Multistep and Bolus require manual programmingXXNonmatch9030.1don't need 8810, use 9030B Braun reports primary doesn't allow secondary as "9030^Current medication cannot be interrupted"
12
8811Rate on Line A and Line B must each be at least 0.5 mL/hr and the total of rates on both lines cannot exceed 500 mL/hCombined rates of Lines A and B must be between
0.5 and 500 mL/hr
XNoncompliant8811
13
8812Unable to program piggyback on primary lineSelect Line B and resubmit smart pump program.XXNoncompliant9030need to investigate origin of this one is it the same as 9030
14
8813Ordered patient BSA does not match BSA currently programmed on pump (clear program on pump then try again)Pump is currently using a different BSA value. Clear the current program then resubmit smart pump program.XNonmatch10002.1B Braun reports as "10002^Patient BSA mismatch"same as 10002
15
8814Ordered patient height does not match height currently programmed on pump (clear program on pump then try again)Pump is currently using a different patient height value. Clear the current program then resubmit smart pump program.XNonmatch10001.1B Braun reports as "10001^Patient height mismatch"same as 10001
16
8815Default drug library installed on infuserReturn pump to Biomed for drug library downloadXDrug Library Error8815
17
8816Unable to change rate or titrate a completed infusionProgram a new bag or increase VTBI manually.XNoncompliant8816
18
8817Rate change program is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XNoncompliant8818.1More generic condition than 8818
19
8818Unable to change rate or titrate an unknown medicationNot allowed when medication is not in the pump drug library. Program pump manually.XNoncompliant8818Pump is running a manually programmed infusion outside of the library Similar to 8817 but more specific.
20
8819Rate change or titration cannot specify a durationProgram pump manually. Review medication order and pump drug library and correct as needed.XNoncompliant10003.1B Braun reports as "10003^Dose over time infusion can only be modified by manual programming"Applies to dose and dose over time infusions
21
8820Keypad is currently lockedUnlock keypad and resubmit smart pump program.XUnavailable State8820new code to replace this one and 10018 "pump is locked and cannot be programmed."Similar to 10018
22
8821Rate unit is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XXXNoncompliant9008.1B Braun reports as "9004^Invalid program parameter (Give Rate Units)" [but may be changing to use 9008]
23
8822Programming not completePrevious program was not completely cleared. Resubmit smart pump program.XUnavailable State8822Same as 8804
24
8823Secondary program rejected because primary infusion is completePrimary line must be active (infusion not complete) to program secondary.XUnavailable State8823
25
8824Medication amount unit is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XXNoncompliant8824
26
8825Unable to program as secondaryProgram pump manually.XXNoncompliant9013.1same as 9013B Braun reports as "9013^Unable to program medication as piggyback"same as 9013? Deliver alone mode may not be a piggyback
27
8826CCA change is in progressTo use new CCA, stop the pump and clear the current program(s). Resubmit smart pump program.XUnavailable State8826
28
8827Smart pump programming not enabledPump gateway is not enabled for smart pump programming.XUnavailable State8827Connectivity = Unable to send order to pump
29
8828Infuser does not support smart pump programmingReturn pump to Biomed for configuration.XUnavailable State8828Connectivity = Unable to send order to pump

B Braun reports as "10064^Auto-programming is not enabled on the pump"
30
8829The patient ID is blankCheck sender's interface settingsXNoncompliant9031.1remove and use 9031B Braun reports as "9031^Patient ID is missing or invalid"same as 9031?
31
8830The patient ID is not supportedCheck sender's interface settingsXNoncompliant9031.2remove and use 9031B Braun reports as "9031^Patient ID is missing or invalid"same as 9031?
32
8831Invalid message versionCheck sender's interface settingsXNoncompliant8831Need to understand how this occurs with ICU Med pumpsB Braun reports as ERR-3 = 200^Unsupported message type^HL70357 - assumes an HL7 level error eg not a PCD03. 8831 might be a TF version related error after the HL7 is processed. (valid PCD03)
33
8832Malformed smart pump programCheck sender's interface settingsXNoncompliant8832Replace with unable to parse smart pum program message new code
34
8833Patient dose lockout period below 5 minute minimumThe order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8833
35
8834Patient dose lockout period exceeds 120 minute maximumThe order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8834
36
8835Invalid maximum dose limit lockout period. Valid hours: 1/4/6/12The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8835
37
8836Missing maximum dose limit or maximum dose limit lockout period (valid hours: 1/4/6/12)The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8836
38
8837Missing patient dose or patient dose lockout period (valid minutes: 5-120)The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8837
39
9001Unknown infuser or channelCheck pump barcode, resubmit smart pump program. If unsuccessful then manually program or return pump to Biomed.e.g., incorrect infuser or channel IDXConnectivityConnectivity9001Connectivity = Unable to send order to pump
40
9002Infuser/channel is currently infusingStop the line and resubmit the smart pump program. (possible cause: a different medication is currently programmed)XXUnavailable StateUnavailable State9002
41
9003Missing required program parameter(s) (ParameterName1, ParameterName2, ...)Program pump manually. Review medication order and pump drug library and correct as needed.e.g., Give Amount Minimum (RXG-5) - volume to be infused - is missingXXXNoncompliantNoncompliant9003
42
9004Invalid program parameter(s) (ParameterName1, ParameterName2, ...)Program pump manually. Review medication order and pump drug library and correct as needed.e.g., volume units are not mLXXXNoncompliantNoncompliant9004
43
9005Parameter (ParameterName) outside of allowable range (MinValue to MaxValue)Ensure rate and patient weight are appropriate for the current CCA. Rate cannot exceed 999 mL/hr.e.g., ordered rate greater than pump maximumXXXOutside LimitsOutside Limits9005
44
9006Infuser is powered offTurn on pump, manually select CCA and resubmit smart pump programXConnectivityConnectivity9006Connectivity
45
9007Infuser is offline or unable to connect to infuserCheck communication icons on pump before resubmitting program, or program pump manually.e.g., infuser not on network or weak wireless signalXConnectivityConnectivity9007Connectivity
46
9008Invalid units for parameter (ParameterName)Concentratione.g., Give Strength Volume Units (RXG-24) contains a medication unit value instead of volume unitsXXXNoncompliantNoncompliant9008
47
9009Dose/Rate Units do not match drug librarye.g., ordered units = mL/hr; drug library units
=mcg/kg/min
XXNonmatchNonmatch9009
48
9010Unable to match medication to drug library Check CCA. Review medication order and pump drug library and correct if needed. (possible cause: medication not in CCA)e.g., Medication does not exist in drug libraryXXNonmatchNonmatch9010
49
9011Patient weight mismatchPump is currently using a different weight value. Stop the pump, clear the current program then resubmit smart pump program. Ordered patient weight does not match weight currently programmed on pump (clear program on pump then try again)e.g., patient weight known by pump differs from weight sent in [PCD-03]XNonmatchNonmatch9011
50
9012Patient ID mismatche.g., patient ID known by pump differs from ID sent in [PCD-03]XXNonmatchNonmatch9012
51
9013Unable to program medication as piggybackInfuse in concurrent mode or use a separate pump.e.g., medication not configured for piggyback administration in drug libraryXXUnsupportedUnsupported9013
52
9014Dose rate or VTBI exceeds maximume.g., greater than pump maximumXXOutside LimitsOutside Limits9014Make more general, remove 8802How does this differ from 9005?
53
9015Request timed outResubmit smart pump program or program manually.XXConnectivityConnectivity9015
54
9016Other errorResubmit smart pump program or program manually.Used when other errors are not applicableXXXMisc.Misc.10050.3Update description to be like 10200
55
9017Infuser cannot accept program Ensure CCA is selected and resubmit smart pump program. (possible causes: CCA or New Patient not selected on infuser)Infuser is in a state where it cannot accept program; e.g., alarming or in standbyXUnavailable StateUnavailable State9017Assumes that pump is configured to support auto programming but not in a state to accept it at this time. Temporary condition.
56
9018Parameter does not match drug libraryProgram pump manually. Review medication order and pump drug library and correct as needed.e.g., Give Strength Units (RXG-18) = mg, drug library = mEqXXNonmatchNonmatch9018
57
9019Patient weight missingProgram pump manually. Review medication order and interface message and correct as needed.Drug is weight-based or BSA-based, but patient weight OBX is missingXXNoncompliantNoncompliant9019
58
9020Patient height missingDrug is BSA-based, but patient height OBX is missingXXNoncompliantNoncompliant9020
59
9021Care area or profile mismatchCare area or profile not cleared on pump; or pump is set to a different care areaXNonmatchNonmatch9021
60
9022Requested infusion program is staleThe value of ORC.9 is older than what the IOC will allow to program the pumpXXNoncompliantNoncompliant9022
61
9023Program rejected by userResubmit smart pump program or program manually.Program rejected or canceled by user prior to starting infusionXMisc.Misc.9023
62
9024Drug library hard limit exceededDose or rate is outside the drug library limit for the medication. Review medication order and pump drug library and correct as needed.e.g., dose exceeds maximum allowable for medicationXXOutside LimitsOutside Limits9024DIfferent from 9005 in that 9024 is related to library configuration vs pump Max. 9024 could also be that the value is below the minimum limitsHow does this differ from 9005? Because it is a drug library hard limit instead of pump limit?
63
9025Lockout interval missingLockout Interval is required when Patient Dose is presentXXNoncompliantNoncompliant9025
64
9026Dose limit missingPCA dose limit is required but missingXXNoncompliantNoncompliant9026
65
9027Patient BSA missingProgram pump manually. Review medication order and interface message and correct as needed.Drug is BSA-based but BSA OBX is missingXXNoncompliantNoncompliant9027
66
9028Tubing cassette or syringe not installed on pumpInstall and resubmit smart pump program.Used when required by pumpXUnmet PrereqsUnmet Prereqs9028
67
9029Care area or profile not selectedSelect CCA and resubmit smart pump program.Used when required by pumpXUnavailable StateUnavailable State9029
68
9030Current medication cannot be interruptedInfuse in concurrent mode or use a separate pump.Medication on primary cannot be interrupted by a piggybackXUnavailable StateUnavailable State9030
69
9031Patient ID missing or invalidXXNoncompliantNoncompliant9031
70
9032Pump is in delayed startCancel delayed start and resubmit smart pump program.XUnavailable StateUnavailable State9032
71
9033Pump is alarmingResolve alarm then resubmit smart pump program.XXUnavailable StateUnavailable State9033
72
9034Not enough fluid or medication in container to deliver bolusApplies to bolus from existing infusion onlyXOutside LimitsOutside Limits9034
73
9035Parent order ID does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatchNonmatch9035
74
9036Medication does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatchNonmatch9036
75
9037Medication concentration does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatchNonmatch9037
76
9038Bolus order type is not supportedORC-1 = “CH” not supported by pumpXXXUnsupportedUnsupported9038
77
9039Medication does not support bolusMedication in drug library not configured for bolusXXUnsupportedUnsupported9039
78
9040Bolus dose units do not match the currently programmed ordere.g., “mg” instead of “mL”XXNonmatchNonmatch9040
79
9041Unable to parse multistep programApplies to multistep infusion onlyXXNoncompliantNoncompliant9041
80
9042Medication is not configured for loading doseApplies to multistep infusion onlyXXUnsupportedUnsupported9042
81
9043Medication does not support multistepApplies to multistep infusion onlyXXUnsupportedUnsupported9043
82
9044Multistep order type is not supportedApplies to multistep infusion onlyXXUnsupportedUnsupported9044
83
9050Malformed MessageXXNoncompliant8832.1Replace with unable to parse smart pum program message new codeFrom Baxter
84
9051Processing ErrorXXXNoncompliant8832.3Replace with unable to parse smart pum program message new codeFrom Baxter
85
9052Device Alias does not match PumpXXMisc.10072.2From Baxter
86
9053Infusion Order Update with no ChangesXXNoncompliant10017.1
B Braun's "10017^Program is identical" may be one reason for an invalid update order
From BaxterSame as 10017
87
9054Order Number MismatchXXNonmatch8832.4Need more info, might be same as 9035From Baxter
88
9055Unknown ErrorXXXMisc.10050.1Use 9016Generic error like 9016 and 10050, 10200From BaxterMay be same as 10050
89
10000Pump stopped communicating during auto-programmingUnable to determine outcome of programming attempt because the pump stopped communicatingConnectivity10000
90
10001Patient height mismatchIf an update order and the new height doesn't match the originalXNonmatch10001.1same as 8814 related to 9011(weight)Add 'X' in Pump column
91
10002Patient BSA mismatchIf an update order and the new BSA doesn't match the originalXNonmatch10002.2same as 8813 related to 9011(weight)Add 'X' in Pump column
92
10003Dose over time infusion can only be modified by manual programmingXNoncompliant10003.2See 8819, and add 'X' in Pump column
93
10004Syringe pump does not allow a secondaryXNoncompliantNot the same as 9013Similar to 8825Add 'X' in Pump column
94
10005Timeout waiting for order reviewUser did not review and accept the order before timeoutXNoncompliant10005Add 'X' in Gateway column
95
10006Drug not found in selected care unitXXDrug Library Error10006Add 'X' in Pump and Drug Library columns
96
10007Secondary orders are only allowed if the primary infusion is programmed as continuousXNoncompliant10007Add 'X' in Pump column
97
10008Concentration mismatchConcentration in order does not match concentration(s) in drug library for selected care unitXXNonmatch10008Add 'X' in Pump and Drug Library columns
98
10009Therapy not supportedIntermittent order received for drug name that is defined in drug library as continuous, or vice versaXXNoncompliant10009Add 'X' in Pump and Drug Library columns
99
10010Incorrect pump typeOrder received for drug that is in drug library but not enabled for the target pump typeXXNonmatch10010Add 'X' in Pump and Drug Library columns
100
10011Concentration missingOrder received for drug that is defined with a concentration, but the order had rate units of ml/hr and no concentrationXXNoncompliant10011Add 'X' in Pump and Drug Library columns