MIT: https://e-vent.mit.edu/clinical/key-ventilation-specifications/
2
AAMI: https://www.aami.org/docs/default-source/standardslibrary/200410_cr501-2020_rev1-2.pdf?sfvrsn=699e62b7_2
3
Testing Details
4
Person Testing:
Fraser Midstokke
5
Date:
6/3/2020
6
Rev#: 15C1
7
Test Lung: Michigan Intruments 3600i
8
Input Pressure:2.2psi
9
Lung Compliance:.02 L
10
Respiratory Rate (RR) (breaths per minute)Pass/FailNotes
11
(breaths per minute): between 6 – 40. Note that the low RRs of 6 – 9 are only applicable to Assist Control.(inflations/min preferably adjustable in steps 50 of no more than 2 inflations/min)Passsteady at ~23bpm
12
<<678910111213141516171819202122232425262728293031323334353637383940 >>
13
X
14
Tidal VolumePass/FailNotes
15
Tidal Volume (TV) (air volume pushed into lung): between 200 – 800 mL based on patient weight. (ml ±10 % in no more than steps of 50 ml)Pass~450
16
<<200250300350400450500550600650700750800 >>
17
X
18
I/E RatioPass/FailNotes
19
(inspiratory/expiration time ratio): recommended to start around 1:2; best if adjustable between range of 1:1 – 1:4*.Pass1:1-2
20
<<1:11:21:31:4 >>
21
XX
22
Inspiratory Pressure LimitPass/FailNotes
23
Airway pressure must be monitored continually. Maximum pressure should be limited to 40 cm H2O at any time - inspiratory pressure limit (15 to 40) cmH2O preferably adjustable in steps of no more than 5 cmH2OPass~25
24
<<152025303540 >>
25
X
26
PEEPPass/FailNotes
27
PEEP of 5–15 cm H2O required; many patients need 10–15 cmH2O (no more than 5 cmH2O steps )PassBetween ~5 to 6
28
<<567891011121314151617181920 >>
29
XX
30
Trigger SensitivityPass/FailNotes
31
When a patient tries to inspire, they can cause a dip on the order of 2 to 7 cm H2O, with respect to PEEP pressure (not necessarily equal to atmospheric pressure).Working on a way to test this
32
<<234567 >>
33
34
Plateau PressurePass/FailNotes
35
Plateau pressure should be limited to max 30 cm H2ONot sure this is relevant to our prototype since ours is pressure bassed.
36
Can we do this?
YesNo
37
38
FiO2 (Oxygen)Pass/FailNotes
39
FiO2 over the range of 21 % (ambient) to 95 % of the source oxygen concentration input to the EUV 44 in no more than 10 % stepsFailNot testing at this time.
40
<<213141516171819195 >>
41
42
Passive Mechanical Blow-off ValvePass/FailNotes
43
The use of a passive mechanical blow-off valve fixed at 40 cm H2O is strongly recommended. This is integrated into most manual resuscitators.FailNot a feature at this time.
44
Can we do this?
YesNo
45
46
Clinician required readingsPass/FailNotes
47
Clinician require readings of plateau pressure and PEEP (refer to Clinical tab)
Can we do this?
YesNoFailNot a feature at this time.
48