202609030900-0940clean optics and bleach lines
3
202608191100-1200Bleached and emptied sediment
4
202608051100-1200Bleached, emptied sediment, and cleaned nitrate, fluorescence, and ...
5
202607211600-1630Emptied debubbler of sediment
6
202606161600-1630Emptied debubbler of sediment
7
202605261000-1100Bleached and emptied sediment
8
202605041200-1300Bleached and cleaned nitrate and fluorescence
9
202601261000-1700Got CTD back online. Cleaned optics
10
202508131100 - 1700calibrated ph probe and transmissometer ,
11
202508121100-1600changed all cables to SBE19, corrosion damage
12
202507091500-1700added debubler
13
202504111000-1200calibrated ph probe
14
202504091000removed transmisometer
15
202503241010-1055cleaned with bleach, removed flow meter. check for outflow in large clear tube left of ph probe
16
202503171500-1530Cleaned, no bleach
17
202503111250-1330Cleaned, no bleach
18
202503101100 - 1230
19
202503071700Reset circuit breaker on weather station - TC
20
202503031530-1400Cleaned. No bleach, fluometer stuck at zero, but there's flow-will keep an eye on it
21
20250222830reset circuit breaker on weather station (had been off since 02/18), invalid PAR readings after data transmission restarted
22
202502201500-1530cleaned, no bleach
23
202502101450-1515cleaned, no bleach
24
202502031515-1545cleaned
25
202501301315-1345cleaned
26
202501131200bad numbers. hose had disconnected due to pressure buildup
27
202412181500-1530cleaned. Flow meter stuck at 1.5. Talked to Jose about the issue
28
20241210-202412110900-1500Repairs to leakon Sandholdt road, data appears to be affected (warm, low DO, low pH)
29
202412111340-1420cleaned, tightened sediment trap base
30
202412051545-1630cleaned, backwash overflow water. Called James and backwash needs to be cleaned. seawater system still functional
31
202411211430-1500cleaned, tubing leaking at t-section
32
202411141340-1410cleaned
33
202411071320-1355cleaned
34
202410211330-1430fixed flow, removed nitrate, cleaned
35
202410101300-1330Repaired hose break that occurred around 2100 on 10/09/24. Cleaned
36
20240927900-1000cleaned and trained
37
202409170900-1600swapped SBE19, calibrated ph
38
202409110900-0930cleaned
39
2024082314:00-14:30flushed
40
20240820took offline to install new hardware. optode is back in
41
202408080900-0930flushed, calibrated SUNA. SUNA is noisy. Bulb likely failing.
42
202407261600-1620cleaned
43
202407081130-1200cleaned
44
202407021530cleaned
45
202406261500removed non responsive optode
46
202405100900-0930cleaned
47
202405021315-1400cleaned. found the flow off and a leak in some tubing. tube was replaced
48
202404221530-1600cleaned and flushed system. suna was fouled reading 0
49
202404100900-1000cleaned and tested synchro instrument
50
202404051520-1550cleaned and flushed
51
20240402910checked wires to pH probe. there was a possible short. issue resolved
52
202403291500-1530cleaned. pH reading low. keep watch possible corroded wire
53
202403201010-1045cleaned
54
202403121600-1630cleaned flow was found off
55
202402281330-1400flushed system. nitrate reading high. possible calibration needed
56
202402161115-1145flushed and cleaned
57
202402091215-1245cleaned optics, flushed system, flow was found off.
58
202401251515-1530cleaned optics, flushed system
59
202401121630-1700cleaned optics
60
202401051650-1715found flow off. flushed system
61
202312261140CTD memory full on 12/19. cleared on 26th
62
202312181300cleaned, system looked clean flow was good
63
202312081030-1050cleaned
64
202312011645-1715cleaned
65
202311211223system was clogged. zero flow
66
202311131400-1430cleaned and put fluorometer back inline. new calibration results tomorrow
67
202310271015-1045cleaned and flushed. upon arrival flow was off
68
202310180850-0920flushed system cleaned optics
69
202310121345-1500cleaned and calibrated ph
70
202310021040-1100cleaned fluoro, trans and pH, flushed system
71
202309211050-1115cleaned fluoro, trans and pH
72
202308310900-1020cleaned optics, flushed system
73
202308161030-1130cleaned, calibrated SUNA
74
202308091000-1400cleaned, calibrated pH and ODO. Took grab sample
75
202308021230-130Flushed system cleaned optics
76
202307190915-0940Flushed system cleaned optics
77
2023712915-945cleaned optics, flushed SUNA, connected to it via phone to run script. Successful
78
20230705100-140Flushed system cleaned optics
79
202306271230-130Flushed system cleaned optics
80
20230621900optide serial unplugged
81
202306201300Breaker tripped and power had been lost for a few days
82
202306160920-1000flushed system cleaned optics
83
20230612SUNA back online (bad connection)
84
20230607flushed system cleaned optics
85
202305311000-1030flushed system cleaned optics
86
20230524900-1000flushed system cleaned optics
87
20230517900-1000flushed system cleaned optics
88
202305111220pH output did not match the readout. Corrosion in one connection. cleaned and that fixed the problem.
89
202305101000-1700calibrated SUNA and optode. Grab samples taken. system flushed.
90
202305030915-1000flushed system cleaned optics
91
202304051000-1100flushed system cleaned optics
92
202303311550-1615flushed system cleaned optics
93
20230322945-1015flushed system cleaned optics
94
20230307Cleaned system
95
202302201521CTD stopped responding on 2/19. Cleared memory even though there was plenty and restarted logging. This fixed the issue.
96
202302140920-1630Cleaned, calibrated ph probe. Issues reconnecting solved.
97
202302081122-1140routine cleaning
98
202301251050-1120flush, clean optics and empty sediment trap
99
202301200930-0955Routine cleaning and set up new PVC connector with Max
100
202301030930-1030flushed lines. cleaned optics