ABCDEFGHIJKLMNOPQRSTUVWXYZAAABACADAEAFAG
1
Test report
version: 1author:Branimir
2
No.itemTestAssociated failure modeTest descriptionResultsOK/ NOKopen issuesTesterDate
Machine version
ResultsOK/ NOKopen issuesTesterDate
Machine version
3
ventilation performance alarms
yesterday OK today leaky setup
prototype 7 + industrial PCB + FW 2.33
v7 + FW 2.
4
1PEEP alarm
1. set peep level 2. introduce PEEP fault by reducing mechanical PEEP setting 3. set PEEP back and trigger inhalations 4.
Branimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
Branimir/Tom/Bruno
09/04/2020
prototype IX PCB prototype FW 2.13
5
2Residual volume alarmBranimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
6
3Volume within threshold
1. set volume to narrow limits around currently achieved volume (e.g. +-30mL) 2. increase pressure for overvolume 3. decrease to go below volume
Branimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
7
4Pressure within threshold
1. normal operation 2. volume limit from previous test 3. set pressure to moderate value, rate to 0.3 and pressure limits to 1 cmH20 3.
Branimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
8
5BPM alarm1. set IE too highBranimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
9
10
machine reliability alarms
11
6arduino not found1. disconnect ARduino cable 2. re-connect and hit reset errormachine keeps on running gives PC error and
weird GUI graphs - could be disabled when this alarm is detected
Branimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
machine keeps on running gives PC error and
weird GUI graphs - could be disabled when this alarm is detected
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
12
7failed sensor calibrationBranimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
13
8disconnected sensor1. unplug sensor flow USB 2. unplug sensor pressure USBBranimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
14
9external power disconnect1. shut off external poweralarm on PC and Arduino lightBranimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
alarm on PC and Arduino light
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
15
10Branimir/Tom10/04/2020
prototype 7 + industrial PCB + FW 2.33
Branimir/Tom/Bruno
09/04/2020
prototype 7 FW 2.34
16
11
17
12
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100