0HL7
Original Code
New codeHL7 Error DescriptionResolutionExamplePump
Gateway
Drug Library
F2F Proposed Category
WG Proposed CategoryGroup sortWG Discussion 23MAY23Updated DescriptionAl
2
190019001Unknown infuser or channelCheck pump barcode, resubmit smart pump program. If unsuccessful then manually program or return pump to Biomed.e.g., incorrect infuser or channel IDXConnectivityConnectivity9001
3
190029002Infuser/channel is currently infusingStop the line and resubmit the smart pump program. (possible cause: a different medication is currently programmed)XXUnavailable StateUnavailable State9002
4
190039003Missing required program parameter(s) (ParameterName1, ParameterName2, ...)Program pump manually. Review medication order and pump drug library and correct as needed.e.g., Give Amount Minimum (RXG-5) - volume to be infused - is missingXXXNoncompliantNoncompliant9003
5
190049004Invalid program parameter(s) (ParameterName1, ParameterName2, ...)Program pump manually. Review medication order and pump drug library and correct as needed.e.g., volume units are not mLXXXNoncompliantNoncompliant9004
6
190059005Parameter (ParameterName) outside of allowable range (MinValue to MaxValue)Ensure rate and patient weight are appropriate for the current CCA. Rate cannot exceed 999 mL/hr.e.g., ordered rate greater than pump maximumXXXOutside LimitsOutside Limits9005
7
190069006Infuser is powered offTurn on pump, manually select CCA and resubmit smart pump programXConnectivityConnectivity9006
8
190079007Infuser is offline or unable to connect to infuserCheck communication icons on pump before resubmitting program, or program pump manually.e.g., infuser not on network or weak wireless signalXConnectivityConnectivity9007
9
190089008Invalid units for parameter (ParameterName)Program pump manually. Review medication order and pump drug library and correct as needed.e.g., Give Strength Volume Units (RXG-24) contains a medication unit value instead of volume unitsXXXNoncompliantNoncompliant9008
10
190099009Dose/Rate Units do not match drug librarye.g., ordered units = mL/hr; drug library units
=mcg/kg/min
XXNonmatchNonmatch9009
11
190109010Unable to match medication to drug library Check CCA. Review medication order and pump drug library and correct if needed. (possible cause: medication not in CCA)e.g., Medication does not exist in drug libraryXXNonmatchNonmatch9010
12
190119011Patient weight mismatchPump is currently using a different weight value. Stop the pump, clear the current program then resubmit smart pump program. Ordered patient weight does not match weight currently programmed on pump (clear program on pump then try again)e.g., patient weight known by pump differs from weight sent in [PCD-03]XNonmatchNonmatch9011
13
190129012Patient ID mismatche.g., patient ID known by pump differs from ID sent in [PCD-03]XXNonmatchNonmatch9012
14
190139013Unable to program medication as piggybackInfuse in concurrent mode or use a separate pump.e.g., medication not configured for piggyback administration in drug libraryXXUnsupportedUnsupported9013
15
190149014Dose rate or VTBI exceeds maximume.g., greater than pump maximumXXOutside LimitsOutside Limits9014Make more general and remove 8802 or make more specific to match the exampleDose rate or VTBI exceeds pump maximum
16
190159015Request timed outResubmit smart pump program or program manually.XXConnectivityConnectivity9015
17
190169016Other errorResubmit smart pump program or program manually.Used when other errors are not applicableXXXMisc.Misc.10050.3Update description to be like 10200 "Other error (error code = <varies>)""Other error [(internal error code = <varies>)]" Where the internal error code is optional and can be included for debugging purposes if available.
18
190179017Infuser cannot accept program Ensure CCA is selected and resubmit smart pump program. (possible causes: CCA or New Patient not selected on infuser)Infuser is in a state where it cannot accept program; e.g., alarming or in standbyXUnavailable StateUnavailable State9017Assumes that pump is configured to support auto programming but not in a state to accept it at this time. Temporary condition.Infuser is currently in a state when it cannot accept the program.
19
190189018Parameter does not match drug libraryProgram pump manually. Review medication order and pump drug library and correct as needed.e.g., Give Strength Units (RXG-18) = mg, drug library = mEqXXNonmatchNonmatch9018
20
190199019Patient weight missingProgram pump manually. Review medication order and interface message and correct as needed.Drug is weight-based or BSA-based, but patient weight OBX is missingXXNoncompliantNoncompliant9019
21
190209020Patient height missingDrug is BSA-based, but patient height OBX is missingXXNoncompliantNoncompliant9020
22
190219021Care area or profile mismatchCare area or profile not cleared on pump; or pump is set to a different care areaXNonmatchNonmatch9021
23
190229022Requested infusion program is staleThe value of ORC.9 is older than what the IOC will allow to program the pumpXXNoncompliantNoncompliant9022
24
190239023Program rejected by userResubmit smart pump program or program manually.Program rejected or canceled by user prior to starting infusionXMisc.Misc.9023
25
190249024Drug library hard limit exceededDose or rate is outside the drug library limit for the medication. Review medication order and pump drug library and correct as needed.e.g., dose exceeds maximum allowable for medicationXXOutside LimitsOutside Limits9024Different from 9005 in that 9024 is related to library configuration vs pump Max. 9024 could also be that the value is below the minimum limitsProgrammed value is outside of the Drug library hard limits for the medication.
26
190259025Lockout interval missingLockout Interval is required when Patient Dose is presentXXNoncompliantNoncompliant9025
27
190269026Dose limit missingPCA dose limit is required but missingXXNoncompliantNoncompliant9026
28
190279027Patient BSA missingProgram pump manually. Review medication order and interface message and correct as needed.Drug is BSA-based but BSA OBX is missingXXNoncompliantNoncompliant9027
29
190289028Tubing cassette or syringe not installed on pumpInstall and resubmit smart pump program.Used when required by pumpXUnmet PrereqsUnmet Prereqs9028
30
190299029Care area or profile not selectedSelect CCA and resubmit smart pump program.Used when required by pumpXUnavailable StateUnavailable State9029
31
190309030Current medication cannot be interruptedInfuse in concurrent mode or use a separate pump.Medication on primary cannot be interrupted by a piggybackXUnavailable StateUnavailable State9030
32
190319031Patient ID missing or invalidXXNoncompliantNoncompliant9031
33
190329032Pump is in delayed startCancel delayed start and resubmit smart pump program.XUnavailable StateUnavailable State9032
34
190339033Pump is alarmingResolve alarm then resubmit smart pump program.XXUnavailable StateUnavailable State9033
35
190349034Not enough fluid or medication in container to deliver bolusApplies to bolus from existing infusion onlyXOutside LimitsOutside Limits9034
36
190359035Parent order ID does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatchNonmatch9035
37
190369036Medication does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatchNonmatch9036
38
190379037Medication concentration does not match the currently programmed orderApplies to bolus from existing infusion onlyXXNonmatchNonmatch9037
39
190389038Bolus order type is not supportedORC-1 = “CH” not supported by pumpXXXUnsupportedUnsupported9038
40
190399039Medication does not support bolusMedication in drug library not configured for bolusXXUnsupportedUnsupported9039
41
190409040Bolus dose units do not match the currently programmed ordere.g., “mg” instead of “mL”XXNonmatchNonmatch9040
42
190419041Unable to parse multistep programApplies to multistep infusion onlyXXNoncompliantNoncompliant9041
43
190429042Medication is not configured for loading doseApplies to multistep infusion onlyXXUnsupportedUnsupported9042
44
190439043Medication does not support multistepApplies to multistep infusion onlyXXUnsupportedUnsupported9043
45
190449044Multistep order type is not supportedApplies to multistep infusion onlyXXUnsupportedUnsupported9044
46
290xxPump or line/channel is in standbyCancel standby and resubmit smart pump programXXUnavailable State9032.1
Add new code for Pump is in standby
47
3880690xxPatient BSA outside of allowable rangeEnsure ordered patient BSA is appropriate for the current CCAXXOutside Limits9005.1add as new code
48
4880790xxPatient height outside of allowable rangeEnsure ordered patient height is appropriate for the current CCAXXOutside Limits9005.2add as new code
49
5880890xxPatient weight outside of allowable rangeEnsure ordered patient weight is appropriate for the current CCAXXOutside Limits9005.3add as new code
50
61001290xxInfusion cannot be programmed as primaryDrug in order is defined as secondary only, but assignment as primary was attemptedXXNoncompliantnew code
51
71005290xxInvalid patient dataPump cannot process the patient-related parameters in the order commandXMisc.Add as a distinct error
52
890xxUnable to parse smart pump program message.XNoncompliantReplaces 8832, "9050, 9051"
53
990xxPump is locked and cannot be programmed.Unlock pump and resubmit smart pump program.XUnavailable StateReplaces 8820 and 10018
54
10881390xxOrdered patient BSA does not match BSA currently programmed on pump (clear program on pump then try again)Pump is currently using a different BSA value. Clear the current program then resubmit smart pump program.XNonmatch10002.1
B Braun reports as "10002^Patient BSA mismatch"
55
11881490xxOrdered patient height does not match height currently programmed on pump (clear program on pump then try again)Pump is currently using a different patient height value. Clear the current program then resubmit smart pump program.XNonmatch10001.1
B Braun reports as "10001^Patient height mismatch"
56
12880390xxA different medication is currently programmedStop the pump and resubmit smart pump programXUnavailable State9002.1add to 90xxPump is currently programmed with a different medication. Clear the pump and resubmit the program.
57
88888801CCA limit exceededEnsure that the CCA is correct and that the rate and patient weight are appropriateXXXOutside Limits8801
58
88888802Infuser limit exceededEnsure rate is < 999 mL/hrXOutside Limits9014.1
59
88888804Infuser is partially programmedClear the current program then resubmit smart pump programXUnavailable State8822.1
60
88888809Dose is invalidThe ordered dose cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXNoncompliant8809
61
88888811Rate on Line A and Line B must each be at least 0.5 mL/hr and the total of rates on both lines cannot exceed 500 mL/hCombined rates of Lines A and B must be between
0.5 and 500 mL/hr
XNoncompliant8811
62
88888812Unable to program piggyback on primary lineSelect Line B and resubmit smart pump program.XXNoncompliant9030
need to investigate origin of this one is it the same as 9030
63
88888815Default drug library installed on infuserReturn pump to Biomed for drug library downloadXDrug Library Error8815
64
88888816Unable to change rate or titrate a completed infusionProgram a new bag or increase VTBI manually.XNoncompliant8816
65
88888817Rate change program is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XNoncompliant8818.1More generic condition than 8818
66
88888818Unable to change rate or titrate an unknown medicationNot allowed when medication is not in the pump drug library. Program pump manually.XNoncompliant8818
Pump is running a manually programmed infusion outside of the library Similar to 8817 but more specific.
67
88888819Rate change or titration cannot specify a durationProgram pump manually. Review medication order and pump drug library and correct as needed.XNoncompliant10003.1
B Braun reports as "10003^Dose over time infusion can only be modified by manual programming"
68
88888821Rate unit is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XXXNoncompliant9008.1
B Braun reports as "9004^Invalid program parameter (Give Rate Units)" [but may be changing to use 9008]
69
88888822Programming not completePrevious program was not completely cleared. Resubmit smart pump program.XUnavailable State8822
70
88888823Secondary program rejected because primary infusion is completePrimary line must be active (infusion not complete) to program secondary.XUnavailable State8823
71
88888824Medication amount unit is invalidProgram pump manually. Review medication order and pump drug library and correct as needed.XXNoncompliant8824
72
88888826CCA change is in progressTo use new CCA, stop the pump and clear the current program(s). Resubmit smart pump program.XUnavailable State8826
73
88888827Smart pump programming not enabledPump gateway is not enabled for smart pump programming.XUnavailable State8827
74
88888828Infuser does not support smart pump programmingReturn pump to Biomed for configuration.XUnavailable State8828
75
88888831Invalid message versionCheck sender's interface settingsXNoncompliant8831
Need to understand how this occurs with ICU Med pumps, this one is probably an HL7 accept ACK error
76
88888833Patient dose lockout period below 5 minute minimumThe order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8833
77
88888834Patient dose lockout period exceeds 120 minute maximumThe order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8834
78
88888835Invalid maximum dose limit lockout period. Valid hours: 1/4/6/12The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8835
79
88888836Missing maximum dose limit or maximum dose limit lockout period (valid hours: 1/4/6/12)The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8836
80
88888837Missing patient dose or patient dose lockout period (valid minutes: 5-120)The order cannot be programmed. Program pump manually. Review medication order and pump drug library and correct as needed.XXOutside Limits8837
81
88889052Device Alias does not match PumpXXMisc.10072.2
82
88889053Infusion Order Update with no ChangesXXNoncompliant10017.1
B Braun's "10017^Program is identical" may be one reason for an invalid update order
83
88889054Order Number MismatchXXNonmatch8832.4
Need more info, might be same as 9035
84
888810000Pump stopped communicating during auto-programmingUnable to determine outcome of programming attempt because the pump stopped communicatingConnectivity10000
85
888810001Patient height mismatchIf an update order and the new height doesn't match the originalXNonmatch10001.1
86
888810002Patient BSA mismatchIf an update order and the new BSA doesn't match the originalXNonmatch10002.2
87
888810003Dose over time infusion can only be modified by manual programmingXNoncompliant10003.2
88
888810004Syringe pump does not allow a secondaryXNoncompliantNot the same as 9013
89
888810005Timeout waiting for order reviewUser did not review and accept the order before timeoutXNoncompliant10005
90
888810006Drug not found in selected care unitXXDrug Library Error10006
91
888810007Secondary orders are only allowed if the primary infusion is programmed as continuousXNoncompliant10007
92
888810008Concentration mismatchConcentration in order does not match concentration(s) in drug library for selected care unitXXNonmatch10008
93
888810009Therapy not supportedIntermittent order received for drug name that is defined in drug library as continuous, or vice versaXXNoncompliant10009
94
888810010Incorrect pump typeOrder received for drug that is in drug library but not enabled for the target pump typeXXNonmatch10010
95
888810011Concentration missingOrder received for drug that is defined with a concentration, but the order had rate units of ml/hr and no concentrationXXNoncompliant10011
96
888810013Patient profile not found in selected care unitOrder received for drug that is in drug library, but is not configured for the patient profile selected by the userXXNonmatch10013
97
888810014Dosing unit mismatchUpdate order received for currently programmed drug, but the dosing units in the order do not match the currently programmed dosing unitsXNonmatch10014
98
888810015Concentration mismatchUpdate order received for currently programmed drug, but the concentration in the order does not match the currently programmed concentrationXNonmatch10015
99
888810016Therapy mismatchUpdate order received for a running continuous infusion, but the parameters in the order are for an intermittent infusionXNonmatch10016
100
888810017Program is identicalUpdate order received, but there were no changes to the existing order parametersXNonmatch10017