ABCDEFGHIJKLMNOPQRSTUVWXYZ
1
CategorySourceIHE CodeIHE TextComments/ExamplesIHE Example
2
OfflineConnectivity9001Pump is alarmingPump cannot accept order while currently...Ex: alarming/being programmed
3
Unavailable StateCurrent infusion9002Infuser cannot accept programInfuser is in a state where it cannot accept program; e.g., alarming or in standby
4
NoncompliantOrder (APR)9003Care area or profile not selectedUsed when required by pump
5
NoncompliantOrder (APR)9004Pump is in delayed startPump cannot accept order while infusion is currently...Ex: delayed/alarming/uninterruptable
6
Outside LimitsOrder (APR)9005Infuser/channel is currently infusing
7
OfflineConnectivity9006Current medication cannot be interruptedMedication on primary cannot be interrupted by a piggyback
8
OfflineConnectivity9007Pump cannot accept…Ex: titrate/bolus/piggyback(9013) ...order without a primary infusion program
9
NoncompliantOrder (APR)9008Tubing, cassette, or syringe not installed on pumpUsed when required by pump
10
NonmatchOrder (APR)9009Bolus order type is not supportedEx: bolus/secondary/loading/multi … order is not supported by pumpORC-1 = “CH” not supported by pump
11
NonmatchOrder (APR)9010Multistep order type is not supportedApplies to multistep infusion only
12
NonmatchOrder (APR)9011Medication does not support bolusEx: bolus/secondary/loading/multi … order is not supported by current infusionMedication in drug library not configured for bolus
13
NonmatchOrder (APR)9012Medication is not configured for loading doseApplies to multistep infusion only
14
UnsupportedCurrent infusion9013Medication does not support multistepApplies to multistep infusion only
15
Outside LimitsOrder (APR)9014Unable to program medication as piggybacke.g., medication not configured for piggyback administration in drug library
16
OfflineConnectivity9015Ex: bolus/secondary/loading/multi … order is not supported by scanned medication
17
Misc.9016Dose/Rate Units do not match drug libraryUnable to match scanned medication to….Ex: Library/Current Infusion/Current Patient Infoe.g., ordered units = mL/hr; drug library units = mcg/kg/min
18
Unavailable StateDevice9017Unable to match medication to drug librarye.g., Medication does not exist in drug library
19
NonmatchOrder (APR)9018Parameter (ParameterName) does not match drug librarye.g., Give Strength Units (RXG-18) = mg, drug library = mEq
20
NoncompliantOrder (APR)9019Medication does not match the currently programmed orderApplies to bolus from existing infusion only
21
NoncompliantOrder (APR)9020Medication concentration does not match the currently programmed orderApplies to bolus from existing infusion only
22
Nonmatch9021Bolus dose units do not match the currently programmed ordere.g., “mg” instead of “mL”
23
NoncompliantOrder (APR)9022Parent order ID does not match the currently programmed orderApplies to bolus from existing infusion only
24
Misc.9023Patient weight mismatche.g., patient weight known by pump differs from weight sent in PCD-03
25
Outside LimitsOrder (APR)9024Patient ID mismatche.g., patient ID known by pump differs from ID sent in PCD-03
26
NoncompliantOrder (APR)9025Care area or profile mismatchCare area or profile not cleared on pump; or pump is set to a different care area
27
NoncompliantOrder (APR)9026Dose rate or VTBI exceeds maximumScanned order exceeds….Ex: Max Dose, Max Rate, Min Dose, VTBI, container volumee.g., greater than pump maximum
28
NoncompliantOrder (APR)9027Drug library hard limit exceedede.g., dose exceeds maximum allowable for medication
29
Unmet PrereqsDevice9028Not enough fluid or medication in container to deliver bolusApplies to bolus from existing infusion only
30
Unavailable StateDevice9029Parameter (ParameterName) outside of allowable range (MinValue to MaxValue)e.g., ordered rate greater than pump maximum
31
Unavailable StateCurrent infusion9030Invalid program parameter(s) (ParameterName1, ParameterName2, ...)Invalid fielde.g., volume units are not mL
32
NoncompliantOrder (APR)9031Invalid units for parameter (ParameterName)Invalid fielde.g., Give Strength Volume Units (RXG-24) contains a medication unit value instead of volume units
33
Unavailable StateCurrent infusion9032Requested infusion program is staleInvalid fieldThe value of ORC.9 is older than what the IOC will allow to program the pump
34
Unavailable StateDevice9033Unable to parse multistep programIs this categorized correctly?Applies to multistep infusion only
35
Outside LimitsOrder (APR)9034Patient ID missing or invalidInvalid or Missing field, split?
36
NonmatchOrder (APR)9035Missing required program parameter(s) (ParameterName1, ParameterName2, ...)Missing fielde.g., Give Amount Minimum (RXG-5) - volume to be infused - is missing
37
NonmatchOrder (APR)9036Patient weight missingMissing fieldDrug is weight-based or BSA-based, but patient weight OBX is missing
38
NonmatchOrder (APR)9037Patient height missingMissing fieldDrug is BSA-based, but patient height OBX is missing
39
UnsupportedDevice9038Lockout interval missingMissing fieldLockout Interval is required when Patient Dose is present
40
UnsupportedCurrent infusion9039Dose limit missingMissing fieldPCA dose limit is required but missing
41
NonmatchOrder (APR)9040Patient BSA missingMissing fieldDrug is BSA-based but BSA OBX is missing
42
NoncompliantOrder (APR)9041Unknown infuser or channele.g., incorrect infuser or channel ID
43
UnsupportedCurrent infusion9042Infuser is powered off
44
UnsupportedCurrent infusion9043Infuser is offline or unable to connect to infusere.g., infuser not on network or weak wireless signal
45
UnsupportedDevice9044Request timed out
46
Unmet PrereqsDeviceOther erroris anyone using this code and how?Used when other errors are not applicable
47
UnsupportedOrder (APR)Program rejected by useris this really a error code?Program rejected by user prior to starting infusion
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100