_ENUM_GROUPSTOKENNUM_VALUEApplicationDescriptionTags
2
Enums with updated/added descriptions
3
_MDC_PUMP_DELIV_STAT pump-delivery-status-deliveringThe pump is currently delivering the fluid at the programmed rate. ds-ip-rel_2.0 MDC_DEV_PUMP_CURRENT_DELIVERY_STATUS
4
_MDC_PUMP_DELIV_STAT pump-delivery-status-not-deliveringThe pump is not currently delivering. See _MDC_PUMP_NOT_DELIVERING_REASONds-ip-rel_2.0
5
_MDC_PUMP_DELIV_STAT pump-delivery-status-kvoThe pump is currently delivering the fluid at a programmed KVO rate. ds-ip-rel_2.0
6
_MDC_PUMP_DELIV_STAT pump-delivery-status-flushingThe pump is currently flushing the fluid line. ds-ip-rel_2.0
7
_MDC_PUMP_DELIV_STAT pump-delivery-status-transitioningThe pump is currently tranistioning to a new rate, new device during a relay or infusate source. ds-ip-rel_2.0
8
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-primingPump is stopped because it completed priming or was stopped during during priming.ds-ip-rel_2.0
9
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-delayedPump is stopped while waiting for a programmed delay period to expire.ds-ip-rel_2.0
10
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-alarmingPump is stopped due to an alarm condition.ds-ip-rel_2.0
11
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-by-clinicianPump was stopped by a clinician.ds-ip-rel_2.0
12
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-powered-offPump is stopped because it was powered offds-ip-rel_2.0
13
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-not-specifiedPump is stopped due to an unspecified reason.ds-ip-rel_2.0
14
_MDC_PUMP_PROGRAM_COMPLETION_MODE pump-program-completion-mode-stopWhen the program has completed the pump will stop.ds-ip-rel_2.0MDC_DEV_PUMP_PROGRAM_COMPLETION_MODE Programmed operation of the pump when a delivery completes (e.g., KVO). {IHE PCD Infusion Pump Work Group }
15
_MDC_PUMP_PROGRAM_COMPLETION_MODE pump-program-completion-mode-kvoWhen the program has completed the pump will continue to deliver but at the kvo rate.ds-ip-rel_2.0
16
_MDC_PUMP_PROGRAM_COMPLETION_MODE pump-program-completion-mode-continueWhen the program has completed the pump will continue to deliver but at the same programmed rate.ds-ip-rel_2.0
17
_MDC_PUMP_PROGRAM_DELIVERY_MODE pump-program-delivery-mode-continuousThe programmed delivery mode for the infusion is a single rate for the duration of the program. The programmed mode could be continuous or intermittent.ds-ip-rel_2.0MDC_DEV_PUMP_PROGRAM_DELIVERY_MODE
18
_MDC_PUMP_PROGRAM_DELIVERY_MODE pump-program-delivery-mode-ramp-taperThe programmed delivery mode for the infusion is ramp-taper. This a specific type of multi-step where the platteau rate and time period are defined and the pump calculates the duration of each step on the ramp up to and taper down from the plateau.ds-ip-rel_2.0
19
_MDC_PUMP_PROGRAM_DELIVERY_MODE pump-program-delivery-mode-multi-stepThe programmed delivery mode for the infusion is multi-step, where the rate transitions to another rate after a period of time. There may be multiple steps defined, including a loading dose.ds-ip-rel_2.0
20
_MDC_PUMP_PROGRAM_DELIVERY_MODE pump-program-delivery-mode-multi-dosingThe programmed delivery mode for the infusion is multi-dosing, is where the same dose is repeatedly administered from the same container at a fixed interval. There may or may not be infusion activity between the doses.ds-ip-rel_2.0
21
_MDC_PUMP_PROGRAM_STATUS pump-program-status-not-startedThe infusion pump is programmed but has not yet started the fluid delivery to the patient.ds-ip-rel_2.0MDC_DEV_PUMP_PROGRAM_STATUS
22
_MDC_PUMP_PROGRAM_STATUS pump-program-status-activeThe infusion pump is programmed and is actively delivering the fluid to the patient.ds-ip-rel_2.0
23
_MDC_PUMP_PROGRAM_STATUS pump-program-status-completeThe infusion pump has completed delivery of the fluid to the patient.ds-ip-rel_2.0
24
_MDC_PUMP_SOURCE_INFO pump-source-info-primaryThe infusion pump is delivering a standard drug dose from the primary reservior.ds-ip-rel_2.0MDC_DEV_PUMP_ENABLED_SOURCES, MDC_DEV_PUMP_ACTIVE_SOURCES
25
_MDC_PUMP_SOURCE_INFO pump-source-info-secondaryThe infusion pump is delivering a standard drug dose from the secondary reservior.ds-ip-rel_2.0
26
_MDC_PUMP_SOURCE_INFO pump-source-info-pcaThe infusion pump is delivering a patient-initiated bolus drug dosing (e.g., PCA/PCEA)ds-ip-rel_2.0
27
_MDC_PUMP_SOURCE_INFO pump-source-info-loadingThe infusion pump is delivering a loading drug dose.ds-ip-rel_2.0
28
_MDC_PUMP_SOURCE_INFO pump-source-info-clinicianThe infusion pump is delivering a clinician initiated bolus drug dose ds-ip-rel_2.0
29
_MDC_PUMP_SOURCE_INFO pump-source-info-intermittentThe infusion pump is delivering a programmed bolus drug dose ds-ip-rel_2.0
30
_MDC_PUMP_STAT pump-status-infusingThe infusion pump mechanism is active.ds-ip-rel_2.0MDC_PUMP_INFUSING_STATUS
31
_MDC_PCA_LOCKOUT_STATUS pca-lockout-dose-limitThe PCA dose-limit lockout is in effect because the maximum cumulative amount of drug that can be administered in a period of time has been reached.
32
_MDC_PCA_LOCKOUT_STATUS pca-lockout-not-specifiedThe pump is locked out of administering additional doses due to an unspecified reason.
33
_MDC_PCA_LOCKOUT_STATUSpca-lockout-noneThe pump is not locked out of administering additional doses.
34
New enums
35
_MDC_PUMP_DELIV_STAT pump-delivery-status-primingThe pump is currently priming the fluid line. Though it is moving fluid, it is not intended to be delivered to the patient, but rather to fill the line, purge air, etc.ds-ip-rel_2.0new enum for group 8
36
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-flushingPump is stopped because it completed the flush or was stopped during the flush.ds-ip-rel_2.0new enum for group 8
37
_MDC_PUMP_NOT_DELIVERING_REASON pump-stopped-switching-deviceThe pump is stopped while transitioning between devices during a relay.new enum for group 8
38
_MDC_PUMP_PROGRAM_STATUS pump-program-status-pausedThe infusion pump is programmed and was started but is not currently actively delivering the fluid to the patient.new enum for group 8
39
_MDC_PUMP_PROGRAM_STATUSpump-program-status-not-programmedThe infusion pump is not completely programmed.new enum for group 8
40
_MDC_PUMP_SOURCE_INFO pump-source-info-primingThe infusion pump is currently priming the fluid line. This fluid is not intended to be delivered to the patient, but rather to fill the line, purge air, etc.new enum for group 8
41
_MDC_PCA_LOCKOUT_STATUS pca-lockout-patient-dose-intervalThe PCA patient-dose lockout is in effect because the minimum amount of time that must pass after the last patient dose has been reached.MDC_PCA_LIMIT_LOCKOUT_STATUSnew enum for group 9
42
_MDC_PCA_LOCKOUT_STATUSpca-lockout-max-patient-dosesPatient has received the maximum number of allowable doses in the time period and is locked out from receiving more doses until the time period has expirednew enum for group 10
43
_MDC_INFUS_ORDER_TYPEclinician-doseAn infusion order type for a bolus dose in addition to the base infusion that is initiated by a clinician.new enum for group 8, used in TF vol 2 section 3.3.4.4.10.2 Bolus from existing infusion
44
_MDC_INFUS_ORDER_TYPEloading-doseAn infusion order type for a bolus in addition to the base infusion initiated prior to the start of the base infusion.new enum for group 8, used in TF vol 2 section 3.3.4.4.10.2 Bolus from existing infusion
45
_MDC_INFUS_ORDER_TYPEcontinuousAn infusion order type for the base infusion where the fluid or medication is administered at a consistent rate. new enum for group 8, used in TF vol 2 section 3.3.4.4.10.2 Bolus from existing infusion
46
_MDC_INFUS_ORDER_STEP_TYPEloading-doseAn infusion order step initiated prior to a continuous step.new enum for group 8, used in TF vol 2 section3.3.4.4.10.3 Multistep
47
_MDC_INFUS_ORDER_STEP_TYPEcontinuousAn infusion order step where the fluid or medication is administered at a consistent rate for the duration of the infusion step.new enum for group 8, used in TF vol 2 section3.3.4.4.10.3 Multistep
48
Enums that can/should be deprecated
49
_MDC_PCA_LOCKOUT_STATUS pca-lockout-patient-doseDeprecated, see pca-lockout-patient-dose-intervalMDC_PCA_LIMIT_LOCKOUT_STATUS
50
_MDC_PUMP_MODE pump-mode-multi-step4Deprecated 10MAY21, add code for "not in use"
51
_MDC_PUMP_MODE pump-mode-multi-dosing8Deprecated 10MAY21, add code for "not in use"
52
_MDC_PUMP_MODE pump-mode-ramp-taper9Deprecated 10MAY21, add code for "not in use"
53
_MDC_PUMP_MODE pump-mode-patient-dose10The infusion pump is programmed to allow patient-initiated drug dosing (e.g., PCA). Replaced pump-mode-pca that has the same value.Deprecated 10MAY21, add code for "not in use"
54
_MDC_PUMP_MODE pump-mode-loading-dose14Deprecated 10MAY21, add code for "not in use"
55
_MDC_PUMP_MODE pump-mode-continuous15Deprecated 10MAY21, add code for "not in use"
56
_MDC_PUMP_MODE pump-mode-piggyback16Deprecated 10MAY21, add code for "not in use"
57
_MDC_PUMP_MODE pump-mode-clinician-dose17Deprecated 10MAY21, add code for "not in use"
58
_MDC_PUMP_MODE pump-mode-kvo18Deprecated 10MAY21, add code for "not in use"
59
_MDC_PUMP_MODE pump-mode-flushing19Deprecated 10MAY21, add code for "not in use"
60
_MDC_PUMP_MODE pump-mode-concurrent20Deprecated 10MAY21, add code for "not in use"
61
_MDC_PUMP_MODE pump-mode-priming21The pump is currently priming the fluid line. Though it is moving fluid, it is not intended to be delivered to the patient, but rather to fill the line, purge air, etc.Deprecated 10MAY21, add code for "not in use"
62
_MDC_PUMP_MODE pump-mode-other22The pump is programmed in an operational mode that does not match one of the defined modes.Deprecated 10MAY21, add code for "not in use"
63
_MDC_PUMP_SOURCE_INFO pump-source-info-secondary-piggybackThis term is not needed, can be withdrawn 10MAY21
64
65
66
67
68
69
Red text is new or changed
70
Yellow highlight has been reviewed by WG
71
Existing descriptions are in black text
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100