PHOPML: Basement Precision Measurement Lab Equipment Maintenance Downtime Log
2
3
Color Code DefinitionsDOWN - Equipment not usable
4
UP - Equipment is operational
5
Partially Operational - Equipment may be usable contact Lab Manager
6
DateEquipmentStatusFollow on NotesSystem Update
7
09/06/12JEOLupthe non stop low volume alarm sound was due to low battery capacity, and the batteries were replacedservice completed
8
09/04/12JEOLdownAlarm sound due to ground wire connection to external feed through, and wouldn't stopwaiting for service
9
05/30/12Supra 55VPup EBL installed, and completed. Still needs to do test runservice completed
10
05/25/12JEOL upMay 18 problem addressed, Routin maintenance completedservice completed
11
05/24/12Supra 55VPdownEBL upgrade service in progress
12
05/18/12JEOLdownx-axis stretched, need recal, contrast near max out waiting for service
13
04/23/12Supra 55VPuptip was replaced, Raith was faulty and will reschedule for the installationservice completed
14
04/03/12JEOLupPMT replaced, and pump oil filled upservice completed
15
03/30/12JEOL downcontrast max out, very noisywaiting for service
16
03/23/12Supra 55VPdownRaith ebeam blanker and NPGS installation, replace aperture setservice in progress
17
02/03/12Supra 40VPupNPGS hard drive restored, a backup drive was installedservice completed
18
01/30/12Supra 40VPonly imaging
EBL is down
NPGS computer hard drive failureIT service in progress
19
01/06/12Supra 55VPuptip and aperture replaced, vac board and Zeiss PC replaced
EDAX only spectrum works, not mapping
Zeiss service completed
20
12/08/11Supra 55VPdowntip failed, PC hard drive failedservice in progress
21
11/07/11Supra40VPupjoystick pad replacedservice completed
22
11/04/11Supra40VPdownjoystick pads doesn't respond, waiting for servicepart on order
23
11/01/11Supra55VPupbroken stage part replacedservice completed
24
10/04/11Supra 55VPdownstage parts damaged, need multiple parts from manufacturewaiting for parts
25
08/11/11Supra 40VPupRaith blanker replacedservice comopleted
26
08/05/11Supra55VPpartially operationalZeiss software (user files and loadlock macro) problems solved, imaging is functional. Communication to EDAX is not up yet, EDAX is downwiting for EDAX service
27
08/05/11Supra 40VPpartially operarional blanker wire part replaced, system bakeout completed, imaging is functional. Raith blanker not working, waiting for replacement.waiting for part
28
08/01/11Supra 40VPdownwire connected to blanker plate replaced; being baked out for 60 hoursservice in progress
29
08/01/11Supra55VPdowncommunication issues in loading Zeiss software; IT and Zeiss service are working on itservice in progress
30
07/29/11Supra 55VPdownHard drive needs to be reformated, IT is working on it.
Zeiss service needs to reload software afterwards
service in progress
31
07/29/11Supra 40VPdownfeedthrough of Raith replaced, still not working. Found wire connected to blanker plate was damagedpart on order
32
07/22/11Supra 40VPdowncolumn power supply to 8KV was faulty, also insulator at Raith feedthrough not proper which caused arcing parts being ordered
33
07/22/11Supra 55VPupIT is still working on PC attacked by virus, but functional with new user password
34
07/21/11JEOL upJEOL is in working conditionservice completed
35
07/18/11JEOL downdisplay monitor quitwiatin for service
36
07/18/11Supra55VPupPump was fixed, some software issue, the computer was slowservice completed
37
07/15/11Supra 55VPdownvac control board replaced, pump still not workingwaing for service
38
07/13/11Supra 55VPdownpump not working, gun vacuum low waiting for service
39
07/11/11Supra 40VPupstage board and joy stick pad were replaced. Zeiss
repeatability test past
service completed
40
06/24/11Supra 40VPoperational with
stage problem
Raithe blanker replaced, y-lead screw replaced, hard panel replaced,
stage problem still exists for EBL
waiting for service
41
06/17/11Supra 40VPdownRaith failed (beam arcing, focusing problem)service in progress
42
05/13/11Supra 40VPoperational with stage problem Stage problem recurred, imaging OK but EBL not acurateservice in progress
43
05/05/11Zeiss Supra 40VPupthe stage problem was repairedservice completed
44
05/02/11Zeiss Supra 40VPdownstage lockup during initializationwaiting for service
45
04/25/11Dimension AFMupthe piezo was replaced and shipped back service completed
46
03/15/11Dimension AFMdownbroken piezo, needs to ship afm back to replace itin service
47
03/08/11Dimension AFMuplaser adjusting knobs repaired from vender, but still not working. Additional repair of exterder box needed.service completed
48
01/17/11Zeiss Supra 40VPdownstage got stuck in Y Axiswaiting for service
49
12/22/10Dimension AFMdownlaser adjusting knobs not workingneeds to ship back to Vender for repair
50
12/03/10Zeiss Supra 40VPUPNoise promblems remianed after bake out. Service found the Raith blanker which was replaced end of September, was faulty. Service replaced the Raith and the tool is back functioning.service completed
51
12/01/10Zeiss Supra 40VPdownservice did further cleaning in upper column, and baked out for 2 daysin service
52
11/23/10Zeiss Supra 40VPoperational with intermitent problemservice only had time to repair stucked stage, the clicking and image shifting issue remain.waiting for more service
53
11/22/10Zeiss Supra 40VPdownstage got stuck in Y Axiswaiting for service
54
11/16/10Zeiss Supra 40VP
operational with intermitent problemintermitent imaging shifting resurfaced when using low KV, EBL at 30KV no problem. Service came in to replace one of the isolation damplers,
still hear clicking sound occasionally, may need to replace more dampers.
need additional service
55
11/11/10Zeiss Supra 40VPup
Cap replaced, tip was centered, many parameters were realigned, and scroll pump was also replaced due to being noisy.service in progress
(needs islolation damper )
56
11/9/2010Zeiss Supra 40VPpatially operational for EBL
waiting for service
57
11/08/10Zeiss Supra 40VP down for service during daytime
partoally operational at night time for EBL
power supply was replaced, but found not the cause of problem.
Further diagnosis was performed by service, and contamination was one of major reasons
service in progress
58
11/04/10Zeiss Supra 40VP
partially
operational for EBL
control-interface problem addressed (water level low in chiller), waiting for 8KV power supply service scheduled early next week
59
11/04/10Zeiss Supra 40VPdowncontrol interface non-functional - and EBL and imaging are both downservice has been contacted, waiting for update on service
60
10/29/10Zeiss Supra 40VPpartialy operationalimage shifted, wobbling, noise problem was diagnosed from a fauty 8KV power supply. Tool works good when using 20KV and above for EBL.waiting for part - to be installed 11/9/10
61
10/27/10Zeiss Supra 40VP
downimage shifted, wobbling, noisewaiting for service
62
10/02/10Zeiss supra 40VP
up gun isolation valve replaced, bake-out overnightservice completed
63
09/24/10Zeiss Supra40VP
downRaith Beam Blanker was replaced on 23rd , but found gun vacuum not sufficient, waiting for service
64
9/9/2010Zeiss Supra 40VP
partially operationalRaith beam blanker not working correctlydiagnosis in progress
65
07/12/10Zeiss Supra 40VP
upone of the dampers was replacedservice completed
66
07/05/10Zeiss Supra 40VPpartially
operatiional
clicking sounds from air pot leaking, image wobblingwaiting for
service
67
06/29/10Zeiss Supra 40VPupPM and tip replacement completedservice completed
68
06/25/10Zeiss Supra40VPdownPM and tip replacement
69
05/19/10Zeiss Supra 40VP
upstage mannually lowered, VP detector was in touch with tall sample.
Ground cable temperarily connected, new one will be replaced later
service completed
70
05/18/10Zeiss Supra 40VP
downstage touching, not able to initialize, ground cable brokenservice call placed
71
05/10/10JEOL 6100upnitrogen gas line pressure not enough, correctedService Completed
72
05/07/10JEOL 6100downrough pump failed waiting for service
73
04/23/10Zeiss Supra 40VPUPIT installed flash drive scan program, and updated McAfee on both Zeiss and Nabity computers. Zeiss User files are still corrupted, It may need to create new account for each user.service completed
74
04/16/10Zeiss Supra 40VPupZeiss computer hard drive replaced, software reloaded, there are still problems with user file setting possibly due to corrupted user files from old hard drive, Zeiss and ebl are functionalservice in progress
75
04/14/10Zeiss Supra 40VP
downZeiss computer kept freezing up due to virus introduced from Flash drive by user, hard panal failed, and eventually failed hard driveservice in progress with IT and Zeiss
76
03/25/10Zeiss Supra 40VPup
turbo pump, and vacuum controller replacedservice completed
77
03/22/10Zeiss Supra 40VPdownvacuum not sufficient waiting for service
78
02/25/10JEOL 6100upwater chiller was replacedservice completed
79
02/10/10JEOL 6100downvacuum insuficient
waiting for service
80
02/02/10Zeiss Supra 40VPupTurbo pump was bad, replaced with refurbished one by Zeiss
service completed
81
02/01/10Zeiss Supra 40VPdownChamber not being pumped down properly
waiting for service
82
01/17/10Zeiss Supra 40VPupservice came in same day, and fixed the frozed stageservice completed
83
10/29/09Veeco AFM
Dimension
upafm computer is workingservice coompleted
84
10/23/09Veeco AFM
Dimension
downcomputer virus problem from memory stickIT service in progress
85
09/02/09Zeiss Supra 40VPupturbo pump replaced, air-leak noise fixedservice completed
86
08/28/09Zeiss Supra 40VPdowntip replace and vibration, air-leak issues afterwardsservice in progress
87
07/30/09JEOL 6100upParts were replaced, and TV mode is workingservice completed
88
07/20/09Zeiss Supra 40VPupdirt and residues found on stage and moving device areaservice completed
89
07/19/09Zeiss Supra 40VPdownStage doesn't move, and doesn' tinitialize
90
06/25/09Zeiss Supra 40VPUPservice completed
91
06/23/09Zeiss Supra 40VPDownrepair of stage "lock-up", stage motor replaced, tension tape both directions replaced, stage board replaced, new hard drive installedservice in progress
92
06/16/09JEOL 6100downTV mode not workingservice in progress
93
06/12/09Zeiss Supra 40VPPartially OperationalHard Panel replaced, new version of SMT software installedservice in progress
94
06/08/09JEOL 6100UPIdentified possible cause. Parts are being ordered. It's funtioning at the momentparts on order
95
06/05/09JEOL 6100downTV mode not working
96
06/05/09Zeiss Supra 40VPPartially
operational
Hard Panel not functioning, no EBL communication
Imaging working by using mouse to do all functions
Service in progress
97
05/18/09JEOL 6100DownTV mode blurry images, mag & scale bar not shown sometimesService call made
98
05/13/09JEOL 6100UPproblems was addressedService call completed
99
05/11/09JEOL 6100DOWNProblems with imagining. Service has been called.Service in progress
100
05/04/09Zeiss Supra 40VPPartially OperationalRotation of the stage is not operational. Waiting for service and parts.Service in progress