CPR Device – for training and practical use !
Parameters measured:
- Compression Depth
- Compression Rhythm
Compression Depth:
- Measured in cm
- Precision upto 2 digits (Eg: 6.57 cm)
Compression Rhythm:
There are two way the measurement of rhythm is implemented;
<insert app interface>
Working Principle
Hardware used:
MPU 6050 measures acceleration
If previous acceleration is –ve and current acceleration is 0
<Insert down motion>
Acceleration 🡪 Velocity 🡪 Distance
Compression depth
1 Compression
This creates a major problem – Integration Error and error drift; (discussed later).
The Double Integration Problem:
a = g (sinθ)
v = ∫ a . dt
x = ∫ v . dt
Assumption: Linear Acceleration = Accelerometer reading - Gravity
a = g (sinθ) (Orientation component)
Error is parabolic
The Solution to the double integration problem:
1) Calibration ; a = g (sinθ) to g … <approx.>
2) Substitute Integration with multiple additions
2.1 )
2.2)
Choosing the correct sampling period ts :
Selecting the appropriate (ts) is a tradeoff between computational latency and parameter resolution
Performance parameters:
Accuracy of compression depth:
Accuracy of rhythm:
Dimensions:
Cost: ₹ 868
< Demo >
Key features: