LEFT VENTRICULAR GUIDEWIRE �PACING DURING TAVR
The Leaflet Lounge
Kristina Mack MSN-ED, BSN, RN, CCRN-CMC, CV-BC
September 2026
DISCLAIMER
THIS PRESENTATION IS FOR EDUCATIONAL USE ONLY. FOR MORE DETAILS AND UPDATES ON THE DEVICES PICTURED OR MENTIONED WITHIN THE TEXT, PLEASE SEE THEIR OFFICIAL WEBSITE.
WHAT IS LV GUIDEWIRE PACING?
Left ventricular guidewire pacing (LV-GW pacing) is a temporary pacing technique that sends the electrical signals directly through the metal guidewire placing it in the left ventricle of the heart.
This procedure is most used during Transcatheter Aortic Valve Replacement (TAVR) to provide rapid ventricular pacing during valvuloplasty and device deployment.
Instead of having to obtain another access as with traditional transvenous pacing, the physician can use the same access for pacing and for valve support.
3
(Image from ScienceDirect.com)
WHY IS RAPID VENTRICULAR PACING USED DURING TAVR?
Rapid ventricular pacing is used during TAVR procedures to temporarily drop cardiac output to stabilize deployment during valve replacement.
Normal cardiac contraction can push against the balloon or the new artificial valve and push it out of position leading to valve embolization.
4
ELECTRICAL PACING CIRCUIT
Key Points:
5
Connection | Location | Function |
Cathode (-) “Back to Black” | Connected to the external end of the guidewire positioned in the left ventricle. It becomes the temporary pacing electrode. It is the negative connection that delivers the pacing stimulus to the myocardium of the left ventricle. | It is the active pacing electrode. The distal curve of the guidewire contacts the left ventricular myocardium to the deliver the stimulus that causes ventricular depolarization (contraction) |
Anode (+) “Red to Ahead” | Connected to a return electrode to complete the circuit.
| Completes the electrical circuit by providing the return pathway through the patient’s tissue. |
REQUIRED EQUIPMENT
6
(Image from Medtronic)
LV VERSUS RV PACING
7
Category | LV Guidewire Pacing | RV Transvenous Pacing |
Access | Uses the existing arterial access and delivery guidewire positioned in the LV Usually avoids a separate venous access. | Requires separate venous access with a sheath and pacing catheter positioned in the RV. |
Pacing location | The LV guidewire acts as the negative electrode (cathode). A cutaneous or subcutaneous electrode acts as the positive return electrode (anode). | A temporary pacing catheter provides electrodes directly within the RV. |
Equipment | LV stiff guidewire, external pulse generator, pacing cable or sterile alligator clips, approved return electrode, and an insulating balloon or valve-delivery catheter. | Venous sheath, temporary transvenous pacing catheter, connecting cables, and external pulse generator. |
Procedural efficiency | May shorten the procedure and reduce fluoroscopy time, equipment use, and cost by eliminated RV lead placement. | Familiar and reliable, but additional venous access and lead positioning my increase procedural and fluoroscopy time. |
Capture characteristics | Capture depends on LV wire contact, insulation, correct polarity, and secure connections. Thresholds may be higher or more variable. | Usually provides predictable capture when the lead has stable contact with the right ventricular myocardium. |
Availability after deployment | Only available when the LV leads is in place, has adequate insulation, and the connections are maintained. | Can remain available after deployment and can be secured for continued pacing during recovery and post-procedural monitoring. |
Post-TAVR conduction support | Limited because the wire must eventually be removed. | Better suited when continued temporary pacing may be required as it can remain in place. |
Access complications | Avoids complications from an additional venous puncture. | Venous bleeding, hematoma, thrombosis, infection, and pneumothorax. |
Cardiac complications | Potential LV perforation, pericardial effusion, tamponade, ventricular ectopy/ ventricular arrhythmia, or mitral valve injury due to the procedural proximity to the valve’s sub-valvular apparatus and chordae tendineae. | Potential RV perforation, pericardial effusion, tamponade, ventricular arrhythmia, lead entanglement, or tricuspid valve injury. |
Pacing-system complications | High threshold, unreliable capture, incorrect polarity, inadequate insulation, guidewire movement, local muscle stimulation, or needle-related injury at the return electrode. | Lead displacement, lead malfunction, failure to capture, inappropriate sensing, or unstable lead positioning. |
Commonly favored when: | Some physicians may prefer a “minimalist” TAVR approach, capture is reliable, and the patient has a low anticipated need for continued pacing. | Baseline conduction disease, RBBB, anticipated high-grade AV block, unreliable LV capture, or likely need for pacing post-artificial valve deployment. |
LV GUIDEWIRE PACING VS. RV TRANSVENOUS PACING
PATIENT AND PROCEDURAL SELECTION
Patients will be evaluated prior to procedure day for:
8
(Image from Edwards Lifesciences)
PREPARATION AND SAFETY CHECK
9
(Image from Shutterstock)
PROCEDURAL WORKFLOW
The physician will position a stiff 0.035” LV support/ delivery guidewire inside the left ventricular apex and myocardial cavity.
A diagnostic catheter or the transcatheter heart valve delivery catheter will remain over the wire to ensure proper electrical insulation and focused current.
10
(Image from Springer Nature)
A return path must be established by using either:
The physician will then connect the external pacemaker box using a sterile disposable alligator clamp on the LV wire terminal.
(Blusztein et al., 2023)
Method | Description | Primary Benefits | Risks |
Grounding Pad | Large adhesive conductive pad that is placed directly on the patient’s skin. |
|
|
Subcutaneous Needle | 21 to 23-gauge needle that is inserted into the patient’s subcutaneous tissue typically in the femoral region. |
|
|
Intravascular Wire | A dedicated conductive wire that is placed directly inside of the femoral vein (J-wire) |
|
|
EFFECTIVE CAPTURE
Test pacing at a slower rate should be performed on each patient to ensure:
Electrical capture alone does not confirm adequate hemodynamic suppression.
13
(Blusztein et al., 2023)
DEPLOYMENT CADENCE
14
TROUBLESHOOTING
Troubleshooting typically involves checking the electrical connections, grounding, and wire positioning.
15
BENEFITS AND LIMITATIONS
Benefits:
Limitations:
16
POSSIBLE COMPLICATIONS
17
POST-TAVR RHYTHM MANAGEMENT
LV wire pacing does not replace post-procedure conduction monitoring. The wire must come out and cannot be secured in place.
Prior to the procedure, the heart team will evaluate each patient for their risk of post-procedural abnormal conduction issues and have a back up plan in place.
Post-procedure management includes discussion of telemetry, EKG assessment, criteria for continued temporary pacing, and escalation for high-grade AV block.