Guideline Update: Sterilization 2025

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

These slides are adapted from the webinar Guideline Update: Sterilization 2025 presented September 10, 2024. To listen to the webinar, visit https://www.aorn.org/education/education-for-individuals/recorded-webinars

2 of 32

Amber Wood is a Senior Perioperative Practice Specialist at AORN where she has served as lead author for several AORN Guidelines. Amber offers clinical information to members via the AORN Consult Line and contributes regularly to the Clinical Issues column in the AORN Journal. She has served as a member of the Association for the Advancement of Medical Instrumentation (AAMI), is a liaison to the CDC Healthcare Infection Control Practices Advisory Committee (HICPAC) and is a fellow of the Association for Professionals in Infection Control and Epidemiology (APIC).

Developed and originally presented by

Amber Wood, MSN, RN, CNOR, CIC, FAPIC

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

3 of 32

��������Objectives��

Outcome

  1. Discuss changes to guideline recommendations.
  2. Describe examples of guideline key takeaways in practice.
  3. Discuss implementation strategies for guideline updates.

​

Eighty percent of perioperative RN attendees will gain knowledge of evidence-based practices related to sterilization and transfer this knowledge into clinical practice.

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

4 of 32

Surgical Site Infection (SSI)

MORTALITY

2- TO 11-FOLD INCREASE

75% DEATHS DIRECTLY ATTRIBUTED TO SSI

LENGTH OF STAY

EXTENDED 9.7 DAYS

COST

$3.3 BILLION ANNUALLY IN US

$20,000 PER ADMISSION

NHSN 2024; Berrios-Torres 2017

Purpose

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

5 of 32

Process devices based on intended use

Sterilization provides the highest level of assurance that surgical items are free of viable microbes

Single-use

Terminal Sterilization

Liquid Chemical Sterilization

Automated HLD

Manual HLD

Margin of Safety

Rutala 2008; Rowan 2023; Rutala 2023

1.1-2

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

6 of 32

What’s new?

Short-cycle sterilization

Additive manufacturing

Transport vehicles

Water quality

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

7 of 32

Manufacturer’s Instructions for Use (IFU)

Device

Packaging

Sterilizer

ANSI/AAMI ST79; ANSI/AAMI ST8; FDA 2015

1.4

Written Clarification

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

8 of 32

Ophthalmic Instruments

Toxic Anterior Segment Syndrome (TASS)

    • Chemical sterilant residue

Unless validated for ocular toxicity, avoid low-temperature sterilization methods

    • Ethylene oxide (EO), hydrogen peroxide

Park 2018; Chang 2018

1.4.5

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

9 of 32

Ophthalmic Instrument Cleaning and Sterilization Task Force

Chang 2018; Chang 2018

5.5

Practice

Short-cycle sterilization

Study

Sterilization effective regardless of interruption of the full drying time

Position

When used in accordance with the IFU of FDA-cleared sterilizers, short-cycle sterilization is appropriate for routine use between sequential same-day ophthalmic cases

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

10 of 32

Short-Cycle Sterilization

Centers for Medicare and Medicaid Services (CMS) 2014; Chang 2018

5.5

Glossary

A sterilization method for a wrapped or contained load that is sterilized in accordance with the device, sterilizer, and packaging manufacturer’s instructions for use; includes a dry time; and is packaged in a sterile barrier that permits storage for use at a later time. – AORN

Correct use of a sterilization cycle for a wrapped/contained load that meets the device manufacturer’s IFU is the equivalent of terminal sterilization and is not IUSS if it includes a dry time and is packaged in a wrap or rigid sterilization container intended to be stored for later use. – CMS

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

11 of 32

Table 1. Differences Between Terminal, Immediate Use, and

Short-Cycle Sterilization 

Terminal

Short-Cycle

IUSS

Packaging

 Wrapped or contained for storage  

Contained for IUSS

Exposure time

Full

Dry time*

Full

Shortened

 

0-1 minute

 

Moisture is expected and accepted inside the package

No

Yes

The item is suitable for storage after the cycle

Yes

No

* Full dry time means the entire dry time recommended in the device manufacturer’s IFU. When a shortened dry time is not provided in the manufacturer’s IFU, the organization is responsible for verifying that a shortened dry time is capable of producing a dry package.

5.5

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

12 of 32

NEW AORN Recommendation

5.5

Perform short-cycle sterilization only when all of the following conditions are met:

    • the device, packaging, and sterilizer manufacturer’s written IFU include instructions for short-cycle sterilization or terminal sterilization (eg, wrapped cycle);
    • the device manufacturer’s written instructions for cleaning, cycle type, exposure times, temperature settings, and drying times are readily available and followed;
    • items are placed in a containment device that has been FDA-cleared and validated for terminal sterilization;
    • the packaging manufacturer’s written IFU for cycle type, exposure times, temperature settings, and drying times are readily available and followed; and
    • if the drying time is shortened, the organization performs verification testing that the shortened time is capable of producing dry packages as described in ANSI/AAMI ST8.

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

13 of 32

AORN Recommendation: IUSS

5.4

Perform IUSS only when all of the following conditions are met:

    • the device and sterilizer manufacturer’s written IFU include instructions for IUSS;
    • the device manufacturer’s written IFU for cleaning, cycle type, exposure times, temperature settings, and drying times (if recommended) are readily available and followed;
    • items are placed in a containment device that has been validated for IUSS and cleared by the FDA for this purpose and in a manner that allows steam to contact all instrument surfaces;
    • the rigid sterilization container manufacturer’s written IFU are followed; and
    • measures are taken to prevent contamination during removal from the sterilizer and transfer to the sterile field.

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

14 of 32

NEW IUSS Study

Tantillo 2022

5.4

Propensity score-matched cohort study

    • Retrospective
    • 10 hospitals, same organization
    • Total knee arthroplasty, total hip arthroplasty, laminectomy, spinal fusion

70,600 patients

    • IUSS Group: 3,526 patients
    • Non-IUSS Group: 3,526 patients
    • Total: 7,052 patients

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

15 of 32

IUSS Study Results

Tantillo 2022

5.4

More infections in the IUSS group

Before matching

    • Groups imbalanced
      • Confounding variables
      • Procedure duration and type
      • IUSS group had more laminectomies and longer duration

No statistically significant difference

After matching

    • IUSS Group: infection rate 1.73% (n = 61)
    • Non-IUSS Group: infection rate 1.42% (n = 50)

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

16 of 32

IUSS Study Conclusions

Tantillo 2022

5.4

IUSS is safe to perform in orthopedic procedures when personnel adhere to the manufacturer’s IFU and guidelines for IUSS from AORN and AAMI.

    • Complying with protocols for cleaning and decontamination, sterilization, and aseptic transfer
    • Reviewing the IFU for sterilization parameters and verifying that parameters for sterilization were achieved
    • Logging and tracking IUSS reasons that can be traced to the patient for surveillance

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

17 of 32

Cooling Instruments after IUSS

Roper 2018

5.3.4

Prevent thermal injury

Instrument use

Cooling time

Most efficient method

    • Pressure
    • Contact time
    • Density
    • Material
    • Ventilation
    • Immersion in 3 L of sterile water

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

18 of 32

Additive Manufacturing

3D Printing

FDA 2017

1.6

  • Powder bed fusion
  • Stereolithography
  • Fused filament fabrication
  • Liquid-based extrusion

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

19 of 32

Sterilization Method

FDA 2017

1.6

Material

Polymer

Metal

Printing Pattern

Porous

Voids

Tortuous

Chemical Reaction

Residue

Device integrity and performance

    • Ability to resist deformation

Penetration of the sterilant

    • Removal of the sterilant

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

20 of 32

Patient-Matched Devices

FDA 2017

1.6.3

Glossary

Based on patient-specific measurements or imaging

Establish process for determining shelf life

    • Event-related sterility
    • Manufacturer expiration date
      • Date of patient’s imaging or finalization of design
      • Surgeon may need to evaluate for anatomical changes before procedure

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

21 of 32

Sterilization of 3D Printed Devices

FDA 2017; Nack 2020

1.6

Manufacturer

    • Perform validated process to remove residual manufacturing material
      • Excess material
      • Support structures
      • Lubricants
    • Provide cleaning and sterilization IFU to end user

End User

    • Perform cleaning and sterilization according to the manufacturer’s IFU for the specific device, packaging, and sterilizer equipment
    • Items cannot be assumed to be sterile unless the manufacturers’ IFU have been followed

On-Site Manufacturing

    • Regulatory requirements
    • Experimental devices
    • Institutional Review Board

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

22 of 32

Sterile Processing Area Design

  • Two-room configuration recommended
    • Separate decontamination room and clean workroom
  • One-room design acceptable when equipment limited to table-top sterilizer
    • 4 feet minimum
    • Functional workflow patterns to separate dirty from clean areas

2.3

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

23 of 32

Transport Between Facilities

Standardization and oversight

Transport vehicles

    • Cargo area
      • surfaces durable to withstand cleaning and disinfection
      • controlled environment (temperature, humidity)
      • complete separation of clean/sterile and contaminated items (transport carrier)
      • minimizes movement (shock, vibration)
    • Ease of loading and unloading
    • Attended in unsecured areas

Compliance with regulatory requirements

    • US Department of Transportation
    • Occupational Safety and Health Administration

3.3

Glossary

ANSI/AAMI ST79; Cardone 2020; US Department of Transportation 2022

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

24 of 32

Water Quality

ANSI/AAMI ST108

10.16

12.8

Critical for effective sterilization

Monitor and control the quality of the water supply system that supplies steam generators (boilers)

Emergency water supply plan

    • Influences steam quality and purity
    • Routinely monitor
    • Seasonal fluctuations
    • Construction or building operation changes
    • Facility’s emergency operations plan

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

25 of 32

Wet Packs

ANSI/AAMI ST79; van Doornmalen 2019; Barbosa Rodrigues 2022

Can indicate a problem

    • Steam supply, sterilizer function, load configuration

Investigate & take corrective action

    • Date and time
    • Load configuration
    • Number and description of trays reported as wet
    • Cycle parameters
    • Type of sterile barrier system
    • Identification of the personnel who packaged the items
    • Identification of the personnel who removed the load from the sterilizer
    • Tray contents, configuration, and weight
    • Sterilizer performance
    • Room temperature and humidity

New publications

    • Heavy trays often implicated
    • Urge caution with tray liners

5.7

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

26 of 32

Sterilization Monitoring

Physical Monitors

    • Review at the end of each cycle
    • Printout: time, temperature, pressure

Chemical Indicators

    • External
      • Type 1 [tape]
    • Internal
      • Type 5 or 6 preferred
    • Do not place on sterile field if failed or questionable

Biological Indicators

    • Review for each load containing an implant
    • Quarantine until results known
    • Report positive results immediately and take corrective actions

ANSI/AAMI ST79

10.3-7

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

27 of 32

Routine Efficacy Testing

Sterilization Method

Frequency

Recommendation

Steam sterilizers

Weekly, preferably each day the sterilizer is used

10.7.1

Low-temperature hydrogen peroxide sterilizers

Each day the sterilizer is used, preferably with every load

10.10.2

Peracetic acid sterilant systems

Each day the sterilizer is used

10.11.2

Ethylene oxide sterilizer

Every load

10.9.2

Dry-heat sterilizers

Weekly, preferably each day the sterilizer is used

10.12.2

Table 3

ANSI/AAMI ST79; ST41; ST58; Rutala 2008

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

28 of 32

Single-Use Devices

FDA requirements for reprocessing single-use devices

    • Form interdisciplinary team to evaluate feasibility
    • Policies and procedures
    • Education
    • Quality measures

Patient education

FDA 2018; Grantcharov 2019

1.5

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

29 of 32

Guideline Essentials

https://www.aorn.org/essentials

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

30 of 32

Questions?

FAQs

AORN Nurse Consult Line

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

31 of 32

References

  • AORN Guideline for Sterilization
  • ANSI/AAMI ST79, ST8, ST108, ST41, ST58
  • Barbosa Rodrigues S, Queiroz de Souza R, Uchikawa Graziano K, Sidnei Erzinger G. Specialists’ opinion regarding factors related to wet loads after steam sterilization. J Hosp Infect. 2022;120:117-122.
  • Berríos-Torres SI, Umscheid CA, Bratzler DW, et al. Centers for Disease Control and Prevention Guideline for the Prevention of Surgical Site Infection, 2017. JAMA Surg. 2017;152(8):784-791.
  • Cardone A, Grous CA. Relocating sterile processing activities to an off-site facility: cost, design, and project management considerations. AORN J. 2020;112(1):30-38.
  • Chang DF, Hurley N, Mamalis N, Whitman J. Evaluation of ophthalmic surgical instrument sterility using short-cycle sterilization for sequential same-day use. Ophthalmology. 2018;125(9):1320-1324.
  • Chang DF, Mamalis N; Ophthalmic Instrument Cleaning and Sterilization Task Force. Guidelines for the cleaning and sterilization of intraocular surgical instruments. J Cataract Refract Surg. 2018;44(6):765-773.
  • Change in Terminology and Update of Survey and Certification (S&C) Memorandum 09-55 Regarding Immediate Use Steam Sterilization (IUSS) in Surgical Settings. August 29, 2014. Accessed June 26, 2024. https://www.cms.gov/Medicare/Provider-Enrollment-and-Certification/SurveyCertificationGenInfo/Downloads/Survey-and-Cert-Letter-14-44.pdf
  • Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling: Guidance for Industry and Food and Drug Administration Staff. March 2015. Accessed June 26, 2024. US Food & Drug Administration, Center for Devices and Radiological Health. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/reprocessing-medical-devices-health-care-settings-validation-methods-and-labeling
  • Technical Considerations for Additive Manufactured Medical Devices. US Food & Drug Administration. 2017. Accessed June 26, 2024. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/technical-considerations-additive-manufactured-medical-devicesFDA. 2018. CPG Sec. 300.500 *Reprocessing of Single Use* Devices.
  • CPG Sec. 300.500 *Reprocessing of Single Use* Devices. US Food & Drug Administration. March 18, 2005. Accessed June 26, 2024. https://www.fda.gov/media/71769/download#:~:text=Reprocessed%20SUDs%20should%20be%20capable,the%20reprocessors%20to%20be%20reprocessed

​

​

Copyright © 2024 AORN, Inc. All rights reserved. Used or adapted with permission.

32 of 32

References

  • Grantcharov P, Ahmed S, Wac K, Rivas H. Reprocessing and reuse of single-use medical devices: perceptions and concerns of relevant stakeholders toward current practices. Int J Evid Based Healthc. 2019;17(1):53-57.
  • Nack B, Nowakowski E, Nicholson F. A central sterile processing and hospital epidemiology and infection control collaboration to ensure safe patient care. AORN J. 2020;112(1):8-14.
  • Surgical Site Infection Event (SSI). National Healthcare Safety Network. January 2024. Accessed June 26, 2024. https://www.cdc.gov/nhsn/pdfs/pscmanual/9pscssicurrent.pdf
  • Park CY, Lee JK, Chuck RS. Toxic anterior segment syndrome – an updated review. BMC Ophthalmol. 2018;18(1):276.
  • Roper D, McClean A, Hand CJ. Too hot to handle: how quickly do surgical instruments cool sufficiently to use safely following sterilisation using the Medical Device Decontamination Capability (Forward)? J R Nav Med Serv. 2018;104(2):87-92.
  • Rowan NJ, Kremer T, McDonnell G. A review of Spaulding's classification system for effective cleaning, disinfection and sterilization of reusable medical devices: viewed through a modern-day lens that will inform and enable future sustainability. Sci Total Environ. 2023;878:162976.
  • Rutala WA, Weber DJ; Healthcare Infection Control Practices Advisory Committee. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008. Centers for Disease Control and Prevention. 2008. Updated June 2024. Accessed June 26, 2024. https://www.cdc.gov/infection-control/media/pdfs/Guideline-Disinfection-H.pdf
  • Rutala WA, Boyce JM, Weber DJ. Disinfection, sterilization and antisepsis: an overview. Am J Infect Control. 2023;51(11S):A3-A12.
  • Tantillo TJ, Stapleton EJ, Frane N, et al. The association of immediate-use steam sterilization with the incidence of orthopaedic surgical site infections: a propensity score-matched cohort study. J Bone Joint Surg Am. 2022;104(11):988-994.