Image Quality and Artifacts in CT
SAHANA KAYASTHA
MSC.MIT 1ST YR
ROLL NO:26
MMC,IOM
1
11/16/2021
Overview
2
11/16/2021
Introduction
3
11/16/2021
Image Quality
4
11/16/2021
5
11/16/2021
Image Quality
6
11/16/2021
Spatial resolution
7
11/16/2021
Spatial Resolution
Spatial Resolution: how it can be measured?
Spatial resolution of an image is measured in two ways:
Spatial resolution
10
11/16/2021
Spatial Resolution
Spatial Frequency
Fig: A simplified illustration of spatial frequency.
12
11/16/2021
13
11/16/2021
14
11/16/2021
0.0125mm
Spatial Resolution
Positioning and alignment, CT number accuracy and slice thickness
Low contrast resolution
CT number uniformity assessment
High contrast (spatial) resolution
CT ACR 464 Phantom
Spatial Resolution
CT ACR 464 Phantom
Spatial Resolution
CT ACR 464 Phantom
Spatial Resolution�
Analyzing the spread of information using Modulation Transfer Function
19
11/16/2021
Spatial Resolution
Modulation Transfer Function
Fig: graphing the MTF of two separate CT systems, An MTF curve extending to the right indicates a system with higher spatial resolution capabilities.
21
11/16/2021
Fig: showing limiting resolution on a given CT system, at an MTF equal to 0.1. In this example, the limiting resolution of scanner A is 4.3 and scanner B is 5.0.
Modulation Transfer Function
An MTF curve that is higher at low spatial frequencies indicates better contrast resolution
Modulation Transfer Function
Spatial Resolution
CT Spatial Resolution
X
Y
Z
Factors affecting spatial resolution
Pixel Size,FOV and matrix Size
Fig: Two small objects in the patient showing how pixel size affect the resolution
Focal Spot Size
Sampling (Detector width and Detector Spacing)
30
11/16/2021
Detector pitch(spacing)
pitch
32
11/16/2021
Slice thickness
Slice thickness
34
11/16/2021
Isotropic Spatial Resolution
Advantages :-
Isotropic Spatial Resolution
Image or Slice thickness
Fig: The MTF shows higher spatial resolution along z for thinner slice images, and the z-axis resolution degrades (the MTF amplitude is reduced) as the reconstructed slice thickness increases.
Slice Sensitivity Profile
Slice Sensitivity Profile
Slice Sensitivity Profile
Slice Sensitivity Profile
Fig: illustrates that in one multi-detector scanner, the table speed can affect the slice sensitivity profile and effective slice thickness. The top curve was obtained using 4 x 5mm collimation mode, pitch of 0.75 (table speed of 15 mm/rot) and a nominal reconstructed slice thickness of 5mm; the calculated FWHM was 5.31 mm. The bottom curve was obtained using the same parameters except that the pitch was increased to 1.5 (30 mm/rot) and the resulting calculated FWHM was 6.24 mm; a 17% increase.
Tradeoff Between Spatial Resolution and Slice thickness
Reconstruction Filter
Reconstruction Filter
FIGURE :CT spatial resolution phantom, consisting of 4–12 line-pairs per centimeter (from American College of Radiology accreditation phantom), reconstructed using standard (A) and bone (B, high-resolution) filters.
Contd…
Fig: shows the MTF of two different reconstruction filters measured on a multidetector scanner (GE LightSpeed Qx/I
46
11/16/2021
smooth
medium
sharp
Patient motion
47
11/16/2021
Contrast Resolution
Contrast Resolution
Fig: No large differences are noted in mass density and effective atomic number among tissues, but the differences are greatly amplified by computed tomography imaging.
Contrast resolution
50
11/16/2021
ACR Phantom
different sizes and with a small difference in density(typically frim 4 to 10 HU) from the
background
Contrast Resolution
Contrast resolution
52
11/16/2021
Factors affecting Contrast Resolution
Factors affecting Contrast Resolution
Noise
Where, xi is each CT value, x is the mean CT value
& n is the no. of CT values averaged
Noise
Noise
Noise
59
11/16/2021
Temporal resolution
Factors Affecting Temporal Resolution
Linearity
Linearity
Fig:Computed tomography (CT) linearity is acceptable if a graph of average CT number versus the linear attenuation coefficient is a straight line that passes through 0 for water.
Uniformity
The CT No. measurement should not change with the location of the selected region of interest (ROI) or with the phantom position relative to the isocentre of the scanner.
Artifacts In CT
Classification of Artifacts
Appearance | Cause |
Streaks | -Improper sampling data, partial volume average, Pt motion, Beam hardening, Noise, spiral/ helical, Mechanical failure |
Shading | -Partial volume averaging , Beam hardening, Spiral/ Helical, Scatter radiation, Off focal radiation , Improper projection |
Rings/Bands | -Bad detector channel |
Classification of Artifacts
Fig:Different appearances of artifacts. A, Streak. B, Ring. C, shading.
Classification of Artifacts
Beam Hardening Artifacts
Beam Hardening Artifacts
Fig:Effect of beam hardening as the x-ray beam traverses different object sizes. A, Original spectrum. B, After traversing 15 cm water. C, After traversing 30 cm water.
Beam Hardening Artifacts
Uncorrected
Corrected
71
11/16/2021
Beam Hardening Correction
Beam Hardening Correction
Beam Hardening Correction
Fig:CT images of the posterior fossa show the that occurs between dense objects when only calibration correction is applied (a) and the reduction in artifacts when iterative beam hardening correction is also applied
Partial Volume Artifacts
Partial Volume Artifacts
Remedy
Partial Volume Artifacts
Photon Starvation Artifacts
Photon Starvation Artifacts
Photon Starvation Artifacts
Fig:Projection data as they might appear for a horizontal x-ray beam passing through the shoulders. Diagrams show the data in their original form (a) and with adaptive filtration (b).
Fig:Original axial CT images (top) and coronal reformatted images (bottom) in their original form (a) and after reconstruction with multidimensional adaptive filtration (b).
Undersampling/Aliasing
Fig:CT image of a Teflon block in a water phantom shows aliasing (arrow) due to undersampling of the edge of the block
Undersampling/Aliasing
Edge Gradient Artifact
Fig:The irregular shading in the left lobe of the liver (indicated by arrows) in this image
Classification of Artifacts
84
11/16/2021
Metal Artifacts
Metal Artifacts
Metal Artifacts
Fig: CT images of a patient with metal spine implants, reconstructed without any correction (a) and with metal artifact reduction (b
Motion Artifacts
Fig: The diagonal shading degrading this image is caused from patient movement during the scan
Motion Artifacts
Remedy
Motion Artifacts
Fig: Patient peristaltic motion. A, Without compensation. B, With correction
Out of field/Incomplete Projection Artifacts
Reduction:-
Classification of Artifacts
Ring Artifacts
Ring artifacts
Avoidance of Ring Artefacts
Tube Arcing
Remedy:
Tube Arcing
Fig: CT tube arcing artifact seen
Fig: Plot of generator kV versus time in images showing drop of voltage corresponding to appearance of tube arcing artifact (A and B)
Classification of Artifacts
Helical Artifacts in the Axial Plane
100
11/16/2021
Fig: Consecutive axial CT images from a helical scan of a cone-shaped phantom lying along the scanner axis.
Helical Artifacts in the Axial Plane
101
11/16/2021
Fig: Series of CT images from a helical scan of the abdomen shows helical artifacts (arrows).
Helical Artifacts in the Axial Plane
To keep helical artifacts to a minimum
102
11/16/2021
Cone Beam Effect
Cone Beam Effect
Cone Beam Effect
Fig:CT images from data collected by an outer detector row (a) and an inner detector row (b) show cone beam artifacts around a Teflon rod, which was positioned 70 mm from the isocenter at an angle of 60° to the scanner axis.
Wind Mill Artifact
Remedy
Avoidance and Correction of Helical and Cone Beam Artifacts
Classification of artefacts
Stair Step Artifacts
Stair Step Artifacts
Remedy:
Zebra Artifacts
Fig: Maximum intensity projection image obtained with helical CT shows zebra artifact
Summary
112
11/16/2021
Image Quality
Spatial Resolution
Contrast Resolution
Temporal Resolution
Noise
Linearity
Uniformity
Artifact
113
11/16/2021
Conclusion
115
11/16/2021
References�
Thank you
116
11/16/2021