1 of 8

100

90

80

70

*Plateau Stability: *This require stability of BOTH the calculated saturation and the reference oximeters. For regulatory purposes, plateaus must be stable for >60 seconds prior to sampling. Labview will define this and green light the calculated SO2 once stable. However, ideally the reference oximeters will also reach stability for 60 seconds (also will be noted as green light in labview). In subjects with low perfusion, reference oximeter stability may require 5-8 minutes of a stable calculated saturation values. Desaturation plateaus should not exceed 10 minutes total.

  • ISO requires data points within 3% of the limits of the range, for example 70-73% and 97-100% if testing the range 70-100%
  • Sample time between blood draws must be >20s
  • Blood gas functional SaO2 may be run in duplicate and result discarded if discrepancy of >2%, otherwise results are averaged
  • This protocol is an example of targets. Actual plateaus may vary at PI discretion based on study subject physiology and safety.

Hypoxia Lab Protocol 2024

94%

1

>60s*

>20s

Last updated Nov 25, 2024

9 / 1-2.5

(Lpm)

Air / N2

9 / 3-5

9 / 8-9

9 / 9**

6 / 8-9**

CO2 0.3-0.5lpm

throughout

6. Drop Air to 3LPM x 4s between plateaus

SaO2 (%)

16 17 18 19

= Run 1

= Run 2

96%

100%

2 3 4 5

84%

20 21 21 23 24

86%

6 7 8 9 10

25 26 27 28 29

72%

76%

11 12 13 14 15

** Actual flow rates required may vary significantly and should be adjusted based on flows required for higher SaO2 plateaus. Record flows for each run. If required flows vary significantly from those here, then check for leaks!

(optional)

(optional)

(optional)

(optional)

2 of 8

OLD VERSIONS BELOW

3 of 8

100

90

80

70

*Plateau Stability: *This require stability of BOTH the calculated saturation and the reference oximeters. For regulatory purposes, plateaus must be stable for >60 seconds prior to sampling. Labview will define this and green light the calculated SO2 once stable. However, ideally the reference oximeters will also reach stability for 60 seconds (also will be noted as green light in labview). In subjects with low perfusion, reference oximeter stability may require 5-8 minutes of a stable calculated saturation values. Desaturation plateaus should not exceed 10 minutes total.

  • ISO requires data points within 3% of the limits of the range, for example 70-73% and 97-100% if testing the range 70-100%
  • Sample time between blood draws must be >20s
  • Blood gas functional SaO2 may be run in duplicate and result discarded if discrepancy of >2%, otherwise results are averaged
  • This protocol is an example of targets. Actual plateaus may vary at PI discretion based on study subject physiology and safety.

Hypoxia Lab Protocol 6.2023

94%

1

>60s*

>20s

Last updated November 15, 2023

9 / 1-2.5

(Lpm)

Air / N2

9 / 3-5

9 / 8-9

9 / 9**

6 / 8-9**

CO2 0.3-0.5lpm

throughout

Drop Air to 3LPM x 4s between plateaus

SaO2 (%)

14 15 16 17

= Run 1

= Run 2

96%

100%

2 3 4 5

84%

18 19 20 21

86%

6 7 8 9

22 23 24 25

72%

76%

10 11 12 13

** Actual flow rates required may vary significantly and should be adjusted based on flows required for higher SaO2 plateaus. Record flows for each run. If required flows vary significantly from those here, then check for leaks!

4 of 8

100

90

80

70

1

>60s

>20s

SaO2 (%)

14 15 16 17

= Run 1

= Run 2

100%

2 3 4 5

18 19 20 21

6 7 8 9

22 23 24 25

10 11 12 13

5 of 8

100

90

80

70

*Plateau Stability: *This require stability of BOTH the calculated saturation and the reference oximeters. For regulatory purposes, plateaus must be stable for >60 seconds prior to sampling. Labview will define this and green light the calculated SO2 once stable. However, ideally the reference oximeters will also reach stability for 60 seconds (also will be noted as green light in labview). In subjects with low perfusion, reference oximeter stability may require 5-8 minutes of a stable calculated saturation values. Desaturation plateaus should not exceed 10 minutes total.

  • ISO requires data points within 3% of the limits of the range, for example 70-73% and 97-100% if testing the range 70-100%
  • Sample time between blood draws must be >20s
  • Blood gas functional SaO2 may be run in duplicate and result discarded if discrepancy of >2%, otherwise results are averaged
  • This protocol is an example of targets. Actual plateaus may vary at PI discretion based on study subject physiology and safety.

Hypoxia Lab Protocol 6.2023

92%

1

>60s*

>20s

Last updated June 15, 2023

9 / 1-2.5

(Lpm)

Air / N2

9 / 3-5

9 / 8-9

9 / 9**

6 / 8-9**

CO2 0.3-0.5lpm

throughout

Drop Air to 3LPM x 4s between plateaus

SaO2 (%)

14 15 16 17

= Run 1

= Run 2

100%

96%

2 3 4 5

82%

18 19 20 21

87%

6 7 8 9

22 23 24 25

73%

77%

10 11 12 13

** Actual flow rates required may vary significantly and should be adjusted based on flows required for higher SaO2 plateaus. Record flows for each run. If required flows vary significantly from those here, then check for leaks!

6 of 8

100

90

80

70

Example: Target samples per plateau = 5; Total number plateaus = 5; Total samples = 25; plateau and sample sizes above are an example, not intended to be a limit. Each plateau should be stable for >30s and time between samples should be >20s

Deasturation plateaus should not exceed 10 minutes. This protocol is provided as an example of targets. Actual plateaus may vary at PI discretion based on study subject physiology and safety.

*Time to stability of plateaus may vary, especially if low perfusion

** Actual flow rates required may vary significantly and should be adjusted based on flows required for higher SaO2 plateaus. Record flows for each run. If required flows vary significantly from those here, then check for leaks!

Open Ox Protocol

92%

1

>60s

>20s

Last updated May 5, 2023

9 / 1-2

(Lpm)

Air / N2

9 / 3-5

9 / 8-9

9 / 9**

6 / 8-9**

CO2 0.5

throughout

Allow calc sat to stabilize AND reference oximeters to stabilize before sampling

Drop Air to 3LPM x 4s between plateaus

SaO2 (%)

14 15 16 17

= Round 1

= Round 2

100%

96%

2 3 4 5

82%

18 19 20 21

87%

6 7 8 9

22 23 24 25

75%

75%

10 11 12 13

7 of 8

SaO2 (%)

100

90

80

70

99-97%

92-85%

84-78%

77-70%

Example: Target samples per plateau = 5; Total number plateaus = 5; Total samples = 25; plateau and sample sizes above are an example, not intended to be a limit. Each plateau should be stable for >30s and time between samples should be >20s

Deasturation plateaus should not exceed 10 minutes. This protocol is provided as an example of targets. Actual plateaus may vary at PI discretion based on study subject physiology and safety.

*Time to stability of plateaus may vary, especially if low perfusion

Example from ISO 80601-2-61:2017(E)

97-92%

2

3

4

5

1

7

14

15

16

6

9

17

18

19

8

11

20

21

22

10

13

23

24

25

12

>30s

>20s

Last updated November 15, 2022

8 of 8

100

90

80

70

100%

95%

Example: Target samples per plateau = 4; Total number plateaus = 4; Total samples = 26; plateau and sample sizes above are an example, not intended to be a limit. Each plateau should be stable for >30s and time between samples should be >20s.

Deasturation plateaus should not exceed 10 minutes. This protocol is provided as an example of targets. Actual plateaus may vary at PI discretion based on study subject physiology and safety.

*Time to stability of plateaus may vary, especially if low perfusion

Hypoxia Lab 2020 ISO 80601-2-61:2017(E)

85%

75%

100%

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

17

18

19

21

22

23

25

26

16

20

24

>30s*

>20s*

Last updated November 15, 2022

SaO2 (%)