The Anatomy of a Spectrum
S. Donnell
AAVSO Spectroscopy Section Lead
Version 1.0
Astronomical Spectroscopy
It Starts with the Raw 2D Spectrum
2
Blue End of Spectrum
Red End of Spectrum
The 1D Profile of the Spectrum
3
Absorption Lines
Noise
Sky Background and Sensor Artifacts are Included in the Spectrum
Sky Background Emission Lines
4
Hot Pixels
The Atmosphere Modifies the Spectrum Through Wavelength-Dependent Extinction
Zenith Angle = 80 deg
Zenith Angle = 60 deg
Zenith Angle = 40 deg
Zenith Angle = 0 deg
All wavelengths are reduced in intensity as the zenith angle increases
Shorter wavelengths (blue) are reduced more than longer wavelengths (red) for all zenith angles
The reduction in blue intensity relative to red increases with increasing zenith angle
5
Non-Optimal Orientation of the Slit Reduces Flux at the Blue and Red Ends of the Spectrum
A slit oriented perpendicular to the horizon will pass most or all of the light across the spectrum.�
A slit oriented horizontal to the horizon will have a decrease in intensity in the blue and red ends.�
Mostly affects slit spectrometers with broad spectral ranges and narrow slits.
6
The Spectrum is Modified by Effects of the Camera Sensor
The sensitivity of the sensor to incident light (AKA response or quantum efficiency) varies by wavelength
Sensors tend to be less responsive in the blue and red ends of the visible spectrum
Different sensor have different spectral response curves
7
The Spectrum is Modified by Effects of the Imaging System
Vignetting - a reduction in brightness toward the edge of the field of view compared to the center
Apparent in a spectrum of a uniform continuum light source
As a result, spectra will be fainter in the blue and red ends of the spectrum compared to the middle green part
Also includes variations in pixel-to-pixel sensitivity of the sensor
8
The Spectrometer Introduces Geometric Distortions
Tilt
Smile
9
Smile refers to the curvature of the spectral lines introduced by the spectrograph configuration.
A wavelength calibration frame is typically used to detect and correct for this distortion.
Tilt refers to the deviation of the spectrum and corresponding lines from the horizontal. Correcting for tilt is desired for summing values along the vertical axis
And, After All the Above…Noise
10
Steps to Obtain a Quality Spectrum
11
Removing the Sky Background
Process is to define a set of regions that
12
Reducing Noise and Removing Sensor Artifacts
What remains is your set of target frames with the noise from the camera reduced and bad pixels disregarded
* Bias frames are used to scale the dark frames for various target frame exposure times.
13
Correct for Pixel Sensitivity Variations and Optical Path Distortions
14
Calibrate the Spectrum for Wavelength
15
Calibrate the Spectrum for Wavelength
16
Correct for Atmosphere and Instrument Response Effects
Causes
17
Difference Between Corrected and Uncorrected Spectrum
18
Steps to Obtain a Quality Spectrum
✅ Rotate the slit
✅ Remove the sky background
✅ Remove sensor artifacts (hot/cold pixels)
✅ Reduce noise
✅ Correct for variations in pixel-to-pixel sensitivity and optical effects
✅ Correct for geometric distortions in the spectrum
✅ Calibrate the spectrum for wavelength
✅ Correct for atmospheric effects
✅ Correct for instrument response
19