Subcutaneous Device for Asthma Attacks Prevention
María del Valle Blázquez Romero, Adrián Gómez Aldudo, Sebastián Huerta, Gabriel Iglesias, Antonio Sánchez Salgado
Asthma Statistics
Asthma patients are forced to carry their inhalers everywhere and if they were to suffer from an asthma attack, they would have to self-administer the treatment which increases the likelihood of human-error leading to inaccurate drug administration.
Prior Art:
Design Goals:
#1 Must be minimally-invasive
#2 Must have dimensions no greater than 3 mL
#3 Must reduce bronchiole inflammation
#4 Must be biocompatible
#5 Must not compromise user’s quality of life
Inaccurate Administration of Asthma Medication
|
| Concept Variants | |||
Selection Criteria | Reference (Inhaler) | Subcutaneous device | Micro-particles filter device | Bronchial Mesh | Low-level laser therapy |
Biocompatibility | 0 | 3 | 3 | 2 | 3 |
Minimally-invasive | 0 | 2 | 2 | 2 | 1 |
Ease of maintenance | 0 | 4 | 2 | 2 | 3 |
Physical durability | 0 | 5 | 3 | 1 | 1 |
Time of development | 0 | 3 | 2 | 5 | 3 |
Novelty | 0 | 3 | 4 | 5 | 5 |
Cost | 0 | 3 | 2 | 1 | 3 |
|
|
|
|
|
|
Total score | 0 | 26 | 20 | 18 | 19 |
Ranking | 5 | 1 | 2 | 4 | 3 |
Continue? | No | Yes | Revise | No | Revise |
Pugh Chart Pointed to a Subcutaneous Device
Block Diagram, Critical Subcomponents
Detection
Biosensor wireless signal transmission
Circuit activates battery & motor
Container
Threaded bar rotates
Piston goes down
Highly concentrated solution of bronchodilator salbutamol
Energy
Stimulus (cytokines concentration)
Bronchodilator drug release
Cytokines detection triggers the response
Block Diagram & Critical Subcomponents
Block Diagram & Critical Subcomponents
Block Diagram & Critical Subcomponents
Battery
Motor
H-Bridge
Switch
Potentiometer
Circuit Detects Signal and Activates Motor
Ø = 3.5 mm
Ø = 0.1 mm
10 mm
50 mm
Cylindrical container
Truncated cone
Piston
Threaded bar
Structural Design and Dimensions
Design Goal #2: “Must have dimensions no greater than 3 mL”
3D SolidWorks Animation
Design Goal #4: “Must be biocompatible”
SolidWorks Design & Material Selection Analysis
Dose Release Calculation
Design Goal #3: “Must reduce bronchiole inflammation”
Dose Release Calculation
Design Goal #3: “Must reduce bronchiole inflammation”
Minimally-invasive procedure
Design Goal #1: “Must be minimally-invasive”
Limitations: Next Steps
Conclusions
Thanks!