1 of 16

Initial Testing Plan

Assistive Hip Exoskeleton

2 of 16

Design Requirements Summary

CR# 

Customer  Requirement 

CR1 

Improved Walking Assistance 

CR2 

Comfortability 

CR3 

Low Profile Design 

CR4 

Safe Operation 

CR5 

Within Budget 

ER# 

Engineering  Requirement 

ER1 

NEMA 1 enclosed system 

ER2 

10 Nm moment about the hip joint 

ER3 

35°/15° + 15° (min + 15 = max) flexion/extension 

ER4 

50° + 15° (min + 15 = max) abduction/adduction 

ER5 

50° + 15° (min + 15 = max) internal/external rotation 

ER6 

<3.0 kg exoskeleton weight 

ER7 

Project cost < $2000 

ER8 

<40 mm lateral plane protrusion 

ER9 

Maximum Safe Motor Torque Output 

3 of 16

Top Level Testing Summary

Experiment/Test 

Relevant DRs 

Thermal Safety 

CR4, CR2  

NEMA 1 Enclosure 

CR4, CR3, ER1, ER8 

(Fitting) ISO 13482:2014 

CR1, CR4, CR2, ER3, ER4, ER5, ER6, ER8 

Walk Test 

CR1, CR2, CR3, ER8, ER9, ER2  

4 of 16

Detailed Testing Plans (NEMA 1)

  • CR & ER Requirements
    • CR4 - Safe Operation 
    • CR3 - Low Profile Design 
    • ER1 - NEMA 1 enclosed system 
    • ER8 - <40 mm lateral plane protrusion 
  • Procedure:
    • Dirt poured top, sides, and bottom of Enclosure
    • Remove Device top of enclosure measure grams of dirt entered device

  • Measurements Variables:
    • Dirt (g) into the device
    • Device is enclosed
  • Calculated Variables:
    • ER1 NEMA 1 Enclosure Standard (Pass/fail)
    • ER2 Low profile
  • Resources/Equipment:
    • Prototype
    • Housing Prototype
    • Dirt (30g)
    • Lab Facilities

5 of 16

Detailed Testing Plans (Thermal Safety)

  • CR & ER Requirements
    • CR4 - Safe Operation
    • CR3 -Comfortable
  • Procedure:
    • Power on the device
    • Wait for 5–10-minute of usage
    • Ensure the max torque is applied
    • Collect Data in 60 second Increments

  • Measurements Variables:
    • Temperature of motor in C°
  • Calculated Variables:
    • Temperature of motor in C° < 33 C°
  • Resources/Equipment:
    • Prototype
    • Lab Software operating
    • Lab Facilities
    • Participant

6 of 16

Detailed Testing Plans – (Fitting Test)

CR & ER Requirements

    • CR2- Comfortability 
    • CR3-Low Profile Design 
    • CR4-Safe Operation 
    • ER3- 35°/15° Flexion/Extension 
    • ER4- 50° /15 Abduction/Adduction 
    • ER5- 50°/15° Internal/External Rotation 
    • ER8 - <3.0 kg exoskeleton weight 

Procedure

    • Potential hazards of device explained orally and written before device is donned by participant (ISO-3482:2014)
    • Angles of movement Measured
    • Weight of Device measured

Measurements Variables

    • Device Protrusion in mm
    • Comfortability Survey
    • Flexion/extension in degrees
    • Abduction/adduction in degrees
    • Internal/external rotation in degrees

Calculated Variables

    • Device Protrusion in mm
    • Comfortability Survey
    • Flexion/extension in degrees
    • Abduction/adduction in degrees
    • Internal/external rotation in degrees

Resources/Equipment

    • Fully assembled device
    • Participant
    • Comfortability Survey
    • Measurement Device
    • Written hazards Review and consent form
    • Lab Facilities

7 of 16

Detailed Testing Plans – (Walk Test)

CR & ER Requirements

  • CR1- Improved Walking Assistance 
  • CR2- Comfortability 
  • CR3- Low Profile Design
  • ER2 - 10 Nm moment about the hip joint 
  • ER8- <40 mm lateral plane protrusion 
  • ER9 - Maximum Safe Motor Torque Output

 

Procedure:

  • Participant will walk 6 meters without device on twice
  • Participant will don device
  • Device will be powered on
  • Max toque will be provided via control software
  • Participant will walk 6 meters with device powered on
  • Assistive Torque will be monitored through lab control software
  • Participant takes off device
  • During testing video footage should be captured

Measurements Variables:

    • Torque Output in Nm
    • Participant Comfortability

Calculated Variables:

    • Amount of gait assistance provided from Torque in Nm
    • Comfortability on scale 1-10

Resources/Equipment:

    • Fully Assembled Prototype
    • Lab control software app
    • Participant
    • Comfortability Survey
    • Written hazards Consent
    • Lab Facilities

8 of 16

Specification Sheet Preparation

Customer Requirement

Met/Not Met (Yes/No)

Client Approved (Yes/No)

CR1 -Improved Walking Assistance

 

 

CR2-Comfortability

 Yes

 

CR3 - Low Profile Design

 Yes

 

CR4 -Safe Operation

 

 

CR5 -Within Budget

 Yes

 

9 of 16

Specification Sheet Preparation

Engineering Requirement

Target

Tolerance

Measured/ Calculated

ER met? (yes/no)

Client Acceptable (yes/no)

ER1 -NEMA 1 enclosed system

0 g dirt

+ 5 g

 1 g 

 Yes

Yes

ER2- Moment about the hip joint

>10 Nm

- 1 Nm

16 Nm

 Yes

Yes

ER3 - flexion/extension

35°/15° 

+/- 5°

 40°/20°

 Yes

Yes 

ER4- abduction/adduction

50°/15° 

+/- 5°

 56°/15°

 Yes

Yes

ER5 - Internal/external rotation

>50°/15°

-0°

 90°/90°

 Yes

Yes

ER6 Device weight

<3kg

+ 0.3 kg

2.8 kg

 Yes

Yes 

ER7 Project cost < $2000

<$2000

+ $0

 $1362.5

 Yes

Yes

ER8 40 mm lateral plane protrusion

40 mm

+/- 5 mm

 38 mm

 Yes

 Yes

ER 9 Maximum Safe Motor Torque Output

<9 Nm

+ 0 Nm

8 Nm

Yes

Yes

10 of 16

Initial Results (50%)

NEMA 1 – Complete

Fitting Test – Complete

Thermal Safety – Incomplete

Walk Test – Incomplete

11 of 16

NEMA 1 Test Results

  • Device and dirt was weighed after each test
  • Scale measurement uncertainty +/-0.01g
  • No more than 2g of dirt entered the device 
  • Brushes were effective in pushing out the dirt

Weight of Device Between Procedures

Before Trial 1

After Trial 1

Before Trial2

After Trial 2

0.37kg

Device

0.37kg

Device

0.37kg

Device

0.37kg

Device

32g

Dirt

32g

Dirt

32 g

Dirt

31 g

Dirt

12 of 16

13 of 16

Fitting Test – Results

(Forward and backward motion)

(Lateral motion)

(Foot rotation)

Three degrees of Freedom Testing

Condition

Goal Angle

Measured Angle

Flexion/Extension

35°/15°

40 °

Abduction/Adduction

50°/15°

57°

Internal/External rotation

50°/15°

90°

14 of 16

Comfortability Survey

  • Is there any source of discomfort?
  • Did the device weigh you down?
  • Did the device interfere with range of motion?
  • Did protrusion interfere with normal motion?
  • On a scale of 1-10 rate comfort

15 of 16

QFD

16 of 16

Questions?