Validation
Of Bowie Dick Test & Biological Indicator Test
Wan Maiza Binti Wan Zahari
Penolong Pengarah Kanan Kejururawatan
Bahagian Kejururawatan KKM
(Merangkap Ketua Perkhidmatan Pensterilan Berpusat (KKM)
Introduction of Bowie & Dick Test
Mr. J. Dick in 1963.
How Bowie & Dick Test Function
Steam Penetration
Air Removal
Composition of the Bowie-Dick test pack
2017 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI ST79:2017
Composition & Challage
Pass
Fail
Pass – Sterilizer is functioning in good conditions
Pass – Sterilizer is showing sign of failure. Sterilizer servicing is required.
Fail – Sterilizer must be shutdown, pending repair and may need re-qualification if repair is deemed significant.
Overexposure – gray or silvery appearance, overexposure (time and/or temperature) is indicated; a retest is required since the test is invalid.
Why is it important?�
BI
Attest
Spore test
Biological Indicator
Conventional BI
Rapid BI
48 hours
24 hours
“Hantar Lab”
Conventional BI
48 hours
24 hours
“Hantar Lab”
&
Ampoule Media
Spore Disc
Geo Bacillus Stearo Thermophilus
Spore
CONTROL TUBE
INCUBATE 60°C
24 Hours
POSITIVE RESULT
PROCESSED TUBE
INCUBATE 60°C
24 Hours
NEGATIVE RESULT
TERRJAH
Rapid BI
3 hours
1 hour
30 Min
20 Min
2 hours
30 Min
Media
INCUBATE 60°C
RESULT READING & VALIDATION
BY INCUBATOR
FLUORESCENT READING
TERRJAH