1 of 17

MLAB 1415: Hematology

Quality Control

&

Sysmex- 430 XN

2 of 17

Quality Control (QC)

  • What is QC?
    • Analyzer performance against a known standard (expected range)
      • Is it in-range?
        • Patient results will be valid
      • Is it out-of-range?
        • Patient results may be compromised
    • Commercial Controls:
      • 3 levels (low, normal, high), each vial has 3 mL of control material
      • Stored at 2-8 0C
      • Values stored in instrument computer
      • Levey-Jennings graph generated and stored for each parameter

3 of 17

Quality Control

4 of 17

QC Preparation

5 of 17

Linearity�(Reportable Range/ AMR)

  • Instruments are calibrated for each analyte with a range that is clinically relevant.
  • With concentrations above or below the reportable range or Linearity range, the result does not correspond with the calibration curve in linear fashion.
  • Results outside of linearity are NOT acceptable.
  • Linearity ranges vary by instrument.
    • Example:

Parameter

Coulter STKS

Advia

WBC

0.0 -99.9 X 103 /µL

0.02-400 X 103 /µL

6 of 17

XN- 430 Reportable Ranges/Linearity

Parameter

Reportable Range/ Units

WBC

0.04-440.00 x 10 3/μL

RBC

0.02-8.60 x 10 6/μL

HGB

0.1-26.0 g/dL

HCT

0.2-74.5 %

PLT

2- 5,000 10 3/μL

7 of 17

Sysmex- XN-430�Principle

  • Quantitative automated hematology analyzer for the in vitro diagnostic use in determining whole blood parameters.
  • Closed tube analysis
  • 6-part differential
  • The instrument uses direct measurements by
    • Impedance- RBC/ PLT
    • Flow cytometry- WBC DIFF
    • SLS (hemoglobin method)- HGB
  • Hematocrit is calculated by the RBC pulse height detection method.

8 of 17

Sysmex 430 XN�Parameters

Parameters

WBC

Neut #

RBC

Lymph#

HGB

Mono #

HCT

Eo#

MCV

Baso#

MCH

Neut %

MCHC

Lymph %

PLT

Mono %

RDW

Eo %

MPV

Baso %

9 of 17

Automation Specs

Number of samples which can be analyzed per hour: 60

Number of tests performed on this analyzer: 1- CBC (see parameters slide)

Quality control: Three levels run each day of patient testing, or following instrument repair, reagent lot change, or when indicated.

Calibration: Shift in control values, replacement of instrument parts, reagent number change, or when indicated

Preventative maintenance: Daily: start up, weekly cleaning with Cellclean and shut down, general cleaning

Acceptable samples: whole blood from an EDTA tube ran within 4 hours of collection

Limitations: Lipemia, cold agglutinins, high WBC’s

10 of 17

Sample Requirements

  • EDTA 2K/3K Whole Blood
  • Analysis within 4 hours of draw

  • Unacceptable specimens
    • Clotted, or those containing clots
    • Grossly hemolyzed
    • Lipemic, icteric, cold agglutinins

11 of 17

Start- Up

  1. If the system is turned OFF
  2. Press the green power button on the front of the instrument
  3. Check the status indicator light to see when ready
  4. If needed, enter user/password
  5. Instrument initialization=background check
    1. See next slide

12 of 17

Background Check

  • Background check verifies the rinse, temperature, and stabilization are in optimal conditions. The background check can be done up to a maximum of three times.

13 of 17

Sample Processing- �

Full tube

Microtainers

Note- Raised bottom tubes have a 250 μL minimum

14 of 17

Reagents

Reagent

Function

Open Expiration

Cellpack DCL

Diluent

60 days

Sulfolyser SLS (1.5L)

Lysing reagent that releases the hemoglobin to be measured.

60 days

Lysercell WDF

Lyses the RBCs, then with Fluorocell dyes WBCs to categorize into WBC populations.

90 days

Fluorocell WDF

Used to stain WBCs for determination of differential count

90 days

15 of 17

Reagent Replacement

16 of 17

Daily Maintenance/Shutdown

17 of 17

Weekly Maintenance