RAPID VENTRICULAR PACINGļæ½RVP DURING TAVR DEPLOYMENT
The Leaflet Lounge
Kristina Mack MSN-ED, BSN, RN, CCRN-CMC, CV-BC
August 2026
DISCLAIMER
The information in the presentation is for educational use only.
TAVR causes conduction disturbances mainly because part of the heartās electrical system lies just beneath the aortic valve. This is the exact area where the transcatheter valve is expanded and anchored.
After leaving the AV node, the bundle of His passes through the central fibrous body and along the membranous interventricular septum, just below the junction of the right and noncoronary aortic cusps. The proximal left bundle branch then runs superficially along the left ventricular outflow tract. A TAVR prosthesis extends from the aortic annulus into this outflow tract, placing its lower frame very close to the HisāPurkinje system (Chen et al., 2020).
(Nuche et al., 2024)
During balloon valvuloplasty and valve deployment, several processes can affect this tissue:
A recent prospective electrophysiology study found that TAVR can impair conduction at both the AV node and the His bundle/infra-nodal system. Many abnormalities improved afterward, consistent with reversible compression and edema in some patients (Medline et al., 2026).
Rapid ventricular pacing (RVP) is deliberately used during transcatheter aortic valve replacement (TAVR). RVP is used to temporarily drive the heart to a rate of 180-220 beats per minute to reduce cardiac output.
RVP will create a short, controlled period of very low cardiac output so the balloon or the prosthetic valve is not pushed out of position by a forceful heartbeat. Dislodgement of the prosthetic valve is called embolization (Matsuo et al., 2025).
The heart has not electrically āstoppedā; it is just contracting rapidly while producing minimal effective forward flow (Matsuo et al., 2025).
(Braasch et al., 2024)
We will use a traditional temporary right ventricular pacing wire.
RVP will be controlled by the circulating cath lab nurse during the procedure.
The physician may have the team perform a test run before crossing the aortic valve to verify stable capture thresholds and ensure safe backup pacing.
Image from Medtronic
CONSIDERATIONS
| Balloon-Expandable Valve | Self-Expanding Valve |
Deployment | Balloon inflation expands the valve essentially all at once. | Valve gradually expands as the delivery capsule is withdrawn. |
Pacing | Usually rapid ventricular pacing, commonly around 180-220 bpm | Often controlled, lower-rate pacing, commonly around 100-160 bpm, depending on the valve, patient, and operator |
Blood Pressure Goal | Marked reduction in cardiac output | Goal is generally stabilization rather than profound hypotension |
Speed | Very brief, usually only several seconds | Slower and more deliberate |
Position Checks | Position confirmed immediately before inflation | Position and implant depth can be repeatedly assessed during partial deployment |
Repositioning | Once balloon deployment occurs, it cannot simply be recaptured. | Many current self-expanding systems can be recaptured and repositioned before the point of no recapture |
Final Release | Inflate 𔪠hold briefly 𔪠completely deflate 𔪠stop RVP | Gradually unsheathe 𔪠confirm depth/position 𔪠proceed beyond recapture point 𔪠final release |
STEPS
For rapid ventricular pacing, the temporary pacemaker is typically set to:
Test ventricular capture
Before valve deployment, the team briefly activates the pacemaker at a slower rate and confirms:
This prevents discovering a failure to capture at the critical moment of valve deployment.
Position the valve or valvuloplasty balloon
The balloon or transcatheter valve is positioned across the native aortic annulus. Fluoroscopy and, when applicable, echocardiography are used to verify depth, alignment, and co-axiality.
The team coordinates the ***Deployment Cadence***.
Begin rapid ventricular pacing or controlled, lower-rate pacing
As the pacemaker rate is increased, the team watches for:
Inflate the balloon or deploy the valve
While the blood pressure and cardiac output are suppressed:
The pacing burst is kept as brief as possible. For balloon-expandable deployment, the balloon is inflated, held briefly, and then completely deflated.
Deflate first, then stop pacing!
Pacing is continued until the balloon is fully deflated.
Once the balloon is deflated:
Deployment Cadence:
The simplified sequence is:
Position valve ā start RVP ā confirm blood-pressure drop ā deploy valve ā deflate balloon ā stop RVP ā confirm recovery.
Assess the result and conduction system
After deployment, angiography, pressure measurements, and echocardiography may be used to assess:
The temporary wire can then function as backup pacing if TAVR has caused high-grade AV block. When there is no new conduction disturbance, it may be removed in the procedure room. With new bundle-branch block, PR/QRS prolongation, or transient or persistent complete heart block, a secure temporary lead may be retained while the patient is monitored.
IMPORTANT RISKS
Because RVP intentionally causes profound transient hypotension, prolonged or repeated pacing can contribute to myocardial ischemia or injury, especially in patients with poor ventricular function or coronary disease.
Other potential problems include failure to capture and ventricular arrhythmias. A conventional right-ventricular temporary wire also carries risks such as venous bleeding, lead displacement, right-ventricular perforation, and cardiac tamponade. This is why the team verifies capture beforehand and minimizes the number and duration of pacing bursts (Matsuo et al., 2025).
The left bundle branch is exposed along the upper ventricular septum, so a new left bundle branch block is a common finding. More extensive injury can produce high-grade or complete AV block.
This is particularly dangerous in someone with a pre-existing right bundle-branch block. Because the right bundle is already nonfunctional, new injury to the left-sided conduction system may eliminate the remaining pathway to the ventricles, leading to complete heart block (Maznyczka et al., 2026).
There is an increased risk of conduction disturbances if:
(Miyashita et al., 2023)
(Bartusik-Aebisher et al., 2025)
Resources:
Bartusik-Aebisher, D., Serafin, I., & Aebisher, D. (2025). New-Onset Left Bundle Branch Block and Other Conduction Disturbances After TAVR: Incidence, Predictors, and Clinical Implications.Ā Prosthesis,Ā 7(4), 71. https://doi.org/10.3390/prosthesis7040071
Braasch, M. C., Alakhtar, A. M., Zajarias, A., & Kaneko, T. (2024). Transcatheter aortic valve replacement embolization: A fleeing, formidable, yet defeatable foe.Ā JTCVS Structural and Endovascular,Ā 4, 100030. https://doi.org/10.1016/j.xjse.2024.100030
Chen, S., Chau, K. H., & Nazif, T. M. (2020). The incidence and impact of cardiac conduction disturbances after transcatheter aortic valve replacement.Ā Annals of Cardiothoracic Surgery,Ā 9(6), 452ā467. https://doi.org/10.21037/acs-2020-av-23
Matsuo, K., Arai, T., Nagai, M., Hori, Y., Hoya, H., Gatate, Y., Yoshitake, A., & Nakano, S. (2025). Impact of Rapid Pacing Time on Myocardial Injury in Transcatheter Aortic Valve Implantation for NonāEnd-stage Renal Disease Patients.Ā CJC Open,Ā 7(12), 1592ā1601. https://doi.org/10.1016/j.cjco.2025.08.011
Maznyczka, A., Pilgrim, T., Philippon, F., RodĆ©s-Cabau, J., & Backer Ole, De. (2026). Conduction disturbances after transcatheter aortic valve implantation.Ā European Heart Journal,Ā 47(30), 4032ā4054. https://academic.oup.com/eurheartj/article-abstract/47/30/4032/8490579?redirectedFrom=fulltext&login=false
Medline, A., Waks, J. W., DāAvila, A., Tung, P., Clarke, J.-R., Poulin, M.-F., Pinto, D. S., Ultimo, B., Kiernan, K., Laham, R. J., & Buxton, A. E. (2026). Electrophysiological observations during and after transcatheter aortic valve replacement.Ā Heart Rhythm, S1547-5271(26)02369-6. https://doi.org/10.1016/j.hrthm.2026.05.015
Miyashita, H., Moriyama, N., Sugiyama, Y., Jalanko, M., Dahlbacka, S., VƤhƤsilta, T., Vainikka, T., ViikilƤ, J., & Laine, M. (2023). Conduction Disturbance After Transcatheter Aortic Valve Implantation With Self- or Balloon-Expandable Valve According to the Implantation Depth.Ā The American Journal of Cardiology,Ā 203, 17ā22. https://doi.org/10.1016/j.amjcard.2023.05.025
Nuche, J., Ellenbogen, K. A., Mittal, S., Windecker, S., Benavent, C., Philippon, F., & Josep RodĆ©s-Cabau. (2024). Conduction Disturbances After Transcatheter Aortic Valve Replacement.Ā ŠŗŠ°ŃŠ“ŠøŠ¾Š»Š¾Š³ŠøŃ ŃŠ·Š±ŠµŠŗŠøŃŃŠ°Š½Š°,Ā 17(22), 2575ā2595. https://doi.org/10.1016/j.jcin.2024.07.032
Thank you!