1 of 15

Matchless Mine Tailings

Experiment 7 Analytical Chemistry

Arielle Boucher & Ivy Goldson

2 of 15

Where is it from?

The Matchless Mine, Leadville CO

​

​

3 of 15

Research question

Is the lead concentration in the leadville mine tailings collected from the tabor mines acceptable by EPA standards?

4 of 15

Procedural Summary

Week 1

  1. Acid washing glassware
  2. Digestion of samples
  3. Method blank
  4. filtration

​

Week 2

  1. Dilution of digestion filtrate
  2. Creating standard additions
  3. Using spectroscopy to test standard additions

​

Week 3

  1. Re filtration of digestion samples
  2. Making of spiked solutions
  3. Using ICP-OES to run the samples.

5 of 15

What is GFAAS and how was it used in this experiment?

  • Excites atomized analyte
  • Cathode composition emits emissions that correspond with analyte
  • Emissions measured to find concentration
  • LOD 1-100 micrograms/L

Graphite Furnace Atomic Absorption Spectroscopy

6 of 15

GFAAS Standard Additions

Sample 2

​

Slope:

​

x-intercept:

​

Sample 1

​

Slope:

​

x-intercept:

​

7 of 15

GFAAS Blank

  • Matrix affects lowered the slope
    • Galena (PbS)
  • Causes amount of metal in blank to differ significantly with a 95% confidence interval from 0

​

8 of 15

GFAAS Results

Sample 1

​

Concentration of lead in soil:

42.6 mg/kg

​

Color: yellow

​

Texture: fine and powdery

Sample 2

​

Concentration of lead in soil:

119 mg/kg

​

Color: grey

​

Texture: course and metallic

9 of 15

What is ICP-OES and how was it used in this experiment?

  • Excites each element and separates them into spectral lines
  • Used to find the total metal concentrations in the samples

10 of 15

Sample 1

​

Concentration of lead in soil:

40.65 mg/kg

​

Color: yellow

​

Texture: fine and powdery

Sample 2

​

Concentration of lead in soil:

95.5 mg/kg

​

Color: grey

​

Texture: course and metallic

The samples were taken only a few inches apart, but their lead concentrations are drastically different

11 of 15

Limit of Detection

LOD

LOD=3*STDEV STDEV=6.402*10^-3

LOD=3*(6.402*10^-3)=1.92*10^-2 mg/L

​

​

​

  • Based on this LOD there was a measurable amount of lead in the sample recovered

​

12 of 15

Spike Recovery

  • Blank recovery: 97%
    • Indicated the method does not interfere
  • Sample 1 recovery: 348 %
    • uhoh
  • Sample 2 recovery: 207%
    • uhoh
  • The matrix is definitively interfering with the analysis.
  • There must be something else in the soil causing this interference
    • Galena (PbS)

​

13 of 15

Sample 1: 40.65 mg/kg

​

Sample 2: 95.5 mg/kg

​

​

​

​

​

​

Sample 1: 42.6 mg/kg

​

Sample 2: 119 mg/kg

​

​

​

​

​

​

ICP-OES

GFAAS

Method Comparison

​

  • Matrix effects in ICP-OES > GFAAS
    • ICP-OES Spike Recovery vs tcalc compared to ttable
  • Therefore GFAAS results are more reliable

14 of 15

Final Results

Epa standards: 1200 ppm

Our results: 42.6-119 ppm

Conclusion: The amount of lead is well within epa standards and not a hazard to Leadville residents. There is not that much lead in Leadville.

15 of 15