The scientific community has been witness to the rise and fall and rise again of supercritical fluid chromatography (SFC) as a robust analytical tool. Due to advancements in instrumentation, such as back pressure regulators, we can begin to take SFC techniques out of the niche and place them amongst the ranks of GC and HPLC. In order to accomplish this, we need to establish a set of column parameters that will allow for the most efficient separations.Define column parameters for analytical scale SFC based on:

    • particle type
    • particle diameter
    • column dimensions

(Consider including a short section called “Key Aspects” here that summarizes your poster in a bullet point fashion)

​

ACT Program

Dr. Paige Wicker & Nazmee Parveen

Joisha Girme, Anna Romero, Ravi Ramkissoon 

​

Acknowledgements

​

Introduction

Methods

Acknowledgements

Results/Expected Results

The Synthesis and Extraction of Endocrine Disrupting Steroids and The Kinetic Rates under Varying External Stressors 

  Destini Black, Aaliyah Devore, Phat Dinh, A. Paige Wicker, Nazmee Parveen, Frank W. Foss, Kevin A. Schug

  Department of Chemistry and Biochemistry, The University of Texas at Arlington

​

Sample Preparation

Instrumentation

Conclusions/Future Work

MS Detection

LCMS-8040

Ionization Source

ESI (+)

Mode

MRM

Precursor: 315.25 m/z

Product: 97.00 m/z (CE=-24 eV)

and 109.00 m/z (CE=-26 eV)

Nebulizing Gas Flow

2 L/min

DL Temperature

250° C

Heat Block Temperature

400° C

Drying Gas Flow

15 L/min

Background:

  • Reported side effect of birth control are nausea, dizziness, and  psychological effects
  • Objective: To find a natural alternative for the intake of progesterone & to synthesize a degraded form of ethinyl estradiol. 

​

Synthesis 

Quantitative Study 

Kinetic 

3 step synthesis 6-⍺-hydroxy-ethinylestradiol

LC Parameters

Extraction and quantitation of progesterone from egg

Progesterone &  6-⍺-hydroxy-ethinylestradiol degradation under mimicked human conditions 

MS

Column

Restek Raptor C18 2.7 μm, 100 x 2.1 mm.

Mobile Phases

Solvent A: 5 mM ammonium acetate, 0.1% formic acid in water

Solvent B: 5 mM ammonium acetate, 0.1% formic acid in methanol

Flow Rate

0.2 mL/min

Injection Volume

2 μL

Gradient

85% B → 100% B for 6 min, hold 4 min

Synthesis Schematics

  • NMR after every synthetic step
  • Expand the gradient to isolate progesterone peak
  • Determine LOD for quantitative study
  • Generate a standard addition calibration curve and study matrix effects
  • Test degradation in differing pH and temperature

​

References

1. Zia et al. Eur. J. Dent. 2019, 13(1), 124-128.

2. Kutztown University Mass Spectrometry. https://www.kutztown.edu/academics/colleges-and-departments/liberal-arts-and-sciences/departments/physical-sciences/chemistry-and-biochemistry/instrumentation/mass-spectrometry/lc-ms.html (accessed 2022-10-24).

3. Shimadzu-LCMS-8040 Liquid Chromatography. https://www.environmental-expert.com/downloads/shimadzu-model-lcms-8040-triple-quadrupole-liquid-chromatograph-mass-spectrometer-lc-ms-ms-brochure-813164 (accessed 2022-10-24).

4. Li et al. Food Sci. Nutr. 2019, 7(7), 2319-2326.

5. Weber et al. J F.Prakt.Chemie.1989, 331(2), 223-230.

6. York et al. Food Chem. 2020, 326(126939).

7. Shen et al. J. Chem. Eng. 2017, 327, 597-607.

8. Ekhato et al. Pub. Med. 2002, 67(3-4), 165-174.

9. Nejedly et al. Spinger. 2017, (24), 23196-23206.

Figure 6 (Expected Kinetic). Simple photodegradation of 17-alpha-ethinylestradiol in purified and seawater (25 °C). Purified water—black line; seawater (4 g L−1)—blue line; seawater (32 g L−1)—red line.9

Figure 1. Schematic presentation of a typical nuclear magnetic resonance spectrometer.1

​

Figure 5 (Quantitative Study). External calibration curve of progesterone from 12.5-200 ppb

Weigh

Vortex

Centrifuge

Extract supernatants

Nitrogen stream

Syringe filtration

LC-MS/MS

Figure 4 (Quantitative Study). Chromatogram of 200 ppb progesterone

Area: 5,160,000±30,000

Figure 2. LC-MS/MS system with a triple quadrupole mass analyzer.2,3