Final Year Project: :��Improving performance of CMOS-MEMS by adding trenches on membrane
STUDENT NAME:1)ER JIA LE 2)STEVEN YANG
PROFESSOR:MING-HUANG LI
�
1
CMUT Working Principle
By applying a small AC voltage signal, with a DC bias on the thin film structure, static electricity is generated, causing deformation and vibration of the thin film structure to produce ultrasonic waves.
When the pressure-induced displacement of the thin film structure occurs due to sound waves, the capacitance value changes. The resulting motion current can be processed through the circuit to obtain a sensing
2
Applications:Ultrasonic Probe
Ultrasound probes commonly used for medical diagnosis have frequency distributions approximately as follows:
3.5 MHz: General abdominal imaging, obstetrics and gynecology
5.0 MHz: Vascular surgery, breast surgery, obstetrics
7.5 MHz: Breast imaging, thyroid imaging
�
Research aim :design a resonant frequency within the range of 1-10 MHz while enhancing the output pressure during CMUT transmission.
3
Device design:Trenches design on membrane
4
The flexural rigidity of a plate:
參考:�Improving the Performance of Capacitive Micromachined Ultrasound Transducers using Modified Membrane and Support Structures
h:effective thickness
Structural design:
5
Conventional
2trenches
4 trenches
Process Flow:
6
Metal
Via
SiO2
conventional
7
Max-disp:0.1174μm
Min.disp:0.0464 μm
2 trenches with 3μm
8
Max-disp:0.0562 μm
Min.disp:0.1456 μm
4 trenches with 3μm
9
Max-disp:0.1888 μm
Min.disp:0.0778 μm
COMSOL simulation result
10
Type | Min. displacement [um] | Max displacement [um] | Improvement of min | Improvement of max |
Conventional | 0.046 | 0.117 | - | - |
2 trenches with 3μm | 0.056 | 0.146 | 21.12% | 24.02% |
2 trenches with 5μm | 0.057 | 0.149 | 22.84% | 27.17% |
4 trenches with 3μm | 0.078 | 0.189
| 72.89% | 64.17% |
4 trenches with 5μm | 0.083 | 0.201 | 78.88% | 71.04% |
Important parameters affecting displacement
Trenches Position :Trenches positioned at the edges maximize the reduction in rigidity.
Number of Trenches :The greater the number of trenches, the larger the displacement at the thin film edges, simultaneously reducing the resonance frequency.
Trenches Width :As the width of the trenches approaches the center of the thin film, the displacement at the center of the thin film also increases.
11
Measure in vacuum(Network Analyzer)
12
Conventional
Simulation frequency: 7.61MHz
Actual frequency:6.30MHz
2trenches
Simulation frequency:6.56MHz
Actual frequency:5.32MHz
4 trenches
Simulation frequency:5.63MHz
Actual frequency:4.72MHz
13
Conventional
2 trenches
type | Frequency(MHz) | Bandwidth(kHz) | Q | loss |
Conventional | 6.30 | 25.533 | 267.89 | -62.29 |
2trenches | 5.32 | 8.36 | 635.98 | -49.626 |
4 trenches | 4.72 | 20.51 | 230.04 | -59.464 |
4 trenches
Ultrasonic probe measurement:
1)The component is immersed in water, and a sinusoidal voltage is applied to the component.
2)Ultrasonic probes are used to measure the output voltage in water at different distances from the component.
3)The output voltage is converted into output pressure.
14
Measurement Data:
15
2.51MHz
BW=0.23MHz
2.10MHz
BW=0.23MHz
conventional
4 Trenches
Output Pressure vs. Distance :
16
Result Comparison
17
Parameter | Device type | Bias voltage | Frequency | Electrode to Resonator Gap | Array number | Distance | Output pressure |
Square shape(This work) | conventional | DC: 50V AC: 10V | 2.51MHz | 0.64μm | 100 | 1mm | 5.4kPa |
Square shape(This work) | 4 Trench design | DC: 50V AC: 10V | 2.10MHz | 0.64μm | 100 | 1mm | 8.04kpa |
NTHU[1] | Dual eletrode | DC:170V Impulse:170V | 1MHz | 0.64μm | 72 | 10mm | 12.72kpa |
NTHU[1] | Single electrode | DC:200V Impulse:170V | 2MHz | 0.64μm | 72 | 10mm | 43.9kpa |
GT[3] | Single electrode | DC:100V Impulse:100V | 3.5MHz | 0.6μm | 33 | 15mm | 2.24kpa |
Conclusion:
18
未來發展
19
感測原理:
20
共振頻率誤差原因:
21
1)晶圓廠製程後的薄膜實際厚度誤差:±10%的厚度
2)後製程中蝕刻可能有undercut或overetching的可能
3)模擬中的設定材料參數上與現實存在一定誤差
COMSOL中 Young‘s modulus設定:silicon:70 GPa Al:70 GPa
�