1 of 37

Shower energy systematics in hdgeant/4

Richard Jones, University of Connecticut

January 2, 2018

2 of 37

Purpose

  • Claims are made that 2𝛾 mass spectra in the calorimeters are different in hdgeant4 simulations than they are with hdgeant.
  • May be due to inconsistent energy scales in the two simulations.
  • May originate in either fcal, bcal or both

2

  1. Compare reconstructed shower energy with generated.
  2. Same conditions (control.in) in hdgeant and hdgeant4 sims.
  3. Single-photon simulation -- gets rid of position resolution, track matching efficiency, other possible reasons for bias in a 2𝛾 mass spectrum.
  4. Separate BCal and FCal showers, check for systematics with polar angle.

Plan:

3 of 37

Part 1: FCal showers

  • uniformly generated in energy [0, 2] GeV
  • uniformly generated in polar angle [0, 20°], azimuthal angle [0, 360°]
  • sliced into bins in polar angle generated
  • plot and fit reconstructed shower energy vs generated.
    • standard control.in physics settings
    • all photons shot from center of target (0, 0, 65) cm
    • material interactions (ie. conversions) allowed to happen
    • run number 51000 (with magnetic field, no DIRC radiators)
    • 100,000 photons generated, 92% are reconstructed as single shower
    • only first reconstructed shower is included in plots
    • quantity plotted is average reconstructed shower energy

3

4 of 37

Fcal: 3° - 5°

4

5 of 37

Fcal: 5° - 7°

5

6 of 37

Fcal: 7° - 9°

6

7 of 37

Fcal: 9° - 11°

7

8 of 37

Fcal: 11° - 13°

8

9 of 37

Part 2: BCal showers

  • uniformly generated in energy [0.5, 1.5] GeV
  • uniformly generated in polar angle [22°, 67°], azimuthal angle [0, 360°]
  • sliced into bins in polar angle generated
  • plot and fit reconstructed shower energy vs generated.
    • standard control.in physics settings
    • all photons shot from center of target (0, 0, 65) cm
    • material interactions (ie. conversions) allowed to happen
    • run number 51000 (with magnetic field, no DIRC radiators)
    • 100,000 photons generated, ~90% are reconstructed as single shower
    • only first reconstructed shower is included in plots
    • quantity plotted is average reconstructed shower energy

9

10 of 37

BCal: 20°-25°

10

11 of 37

BCal: 25°-30°

11

12 of 37

BCal: 30°-35°

12

13 of 37

BCal: 35°-40°

13

14 of 37

BCal: 40°-45°

14

15 of 37

BCal: 45°-50°

15

16 of 37

BCal: 50°-55°

16

17 of 37

BCal: 55°-60°

17

18 of 37

BCal: 60°-65°

18

19 of 37

BCal: 65°-70°

19

20 of 37

Conclusions

  1. significant systematic bias is seen as a function of polar angle
  2. bias visible both in the fcal and the bcal
  3. agreement between hdgeant and hdgeant4 is quite good:
    1. very good in the Fcal
    2. difference in the Bcal in the y-offset of the E-calibration
  4. source of difference in the Bcal not clear -- what would produce an offset?
  5. in all cases, size of E-systematic is greater than the hdgeant/4 difference.

20

21 of 37

Nonlinear effects

  • Next I repeat all of the plots shown above with quadratic fits
  • Provides comparison in nonlinear response between hdgeant and hdgeant4
  • The data are all the same as shown before, only the fit is different

21

22 of 37

Fcal: 3° - 5°

22

23 of 37

Fcal: 5° - 7°

23

24 of 37

Fcal: 7° - 9°

24

25 of 37

Fcal: 9° - 11°

25

26 of 37

Fcal: 11° - 13°

26

27 of 37

Bcal: 20° - 25°

27

28 of 37

Bcal: 25° - 30°

28

29 of 37

Bcal: 30° - 35°

29

30 of 37

Bcal: 35° - 40°

30

31 of 37

Bcal: 40° - 45°

31

32 of 37

Bcal: 45° - 50°

32

33 of 37

Bcal: 50° - 55°

33

34 of 37

Bcal: 55° - 60°

34

35 of 37

Bcal: 60° - 65°

35

36 of 37

Bcal: 65° - 70°

36

37 of 37

Conclusions regarding nonlinear response

  • good consistency between hdgeant and hdgeant4 in both fcal and bcal.
  • no residual nonlinearity is observed in the bcal, at least at angles < 70°
  • very little residual nonlinearity is observed in the fcal
    • quadratic coefficient probably rises to +0.01 around 9°
    • then it dives to negative values above 11°, as showers leak out the sides of the fcal
    • this latter effect is better treated as an energy leakage than a “nonlinear effect”
  • the fcal energy scale still needs work as a polar angle dependent correction
  • proposal -- should be applied to the block energies in mcsmear as f(r)

37