Precision Under Pressure: AR Simulation for Postpartum Cesarean Hysterectomy
I. THE PROBLEM
1. Meinig School of Biomedical Engineering, Cornell University, Ithaca, New York, United States of America�2. Weill Cornell Graduate School of Medical Sciences, New York, New York, United States of America
IV. NEEDS STATEMENT
A way to train physicians and residents in postpartum cesarean hysterectomy to improve provider comfort with the procedure and reduce severe maternal morbidity.
II. OUR SOLUTION- THE HYBRID HAPTIC AR APPROACH
.
III. SYSTEM ARCHITECTURE
Real silicone skin and uterus phantom. Surgeon feels true tissue resistance during incision.
2. TRACKING LAYER
HoloLens 2 infrared tracking detects retroreflective markers on surgical tools, enabling real-time identification of scalpel and forceps from a single camera.
3D tool position is computed at 30 FPS using stereo triangulation with a 99.6 mm baseline, validated against hardware specs.
A Bosch BNO055 IMU uses built-in sensor fusion to track real-time tool orientation (pitch, roll, yaw).
Holographic overlay of uterus, bladder, and uterine arteries on phantom with surgical-grade registration.
3. AR LAYER
Real-time alerts: red warning for danger zone entry, green cue for correct dissection plane.
Depth-triggered reveal of full anatomy within 2 seconds of incision for intraoperative realism.
VI. TRACKING PERFORMANCE AND VALIDATION
VII. AR EXPERIENCE AND RESULTS
VIII. FUTURE WORK
Omar Abououf 1, Daniella Ameh 1 , Shreya Dandwate 1 , Dr. Newton DeFaria 1, Dr. Suelia Fleury Rosa1,
Jordan McMahan1, Jack Thompson 1Dr. Tamatha Fenster 2, Dr. Anna Daoud 2��
V. PHYSICAL PHANTOM
To replicate postpartum tissue, we evaluated 8 silicone formulations using dial indicator force measurement against an available skin reference model.
Figure 5.1 Dial indicator force reference testing against available skin model.
Figure 7.2 IR Blob detection (AHAT Camera)
Figure 6.1 Wearable Surgical Tracking Prototype
Figure 7.1 Unity - Operation Room
Metric | Result |
Stereo baseline | 99.6 mm HoloLens 2 spec |
3D position tracking | 30 FPS real-time |
Orientation tracking | 100 Hz, zero drift observed |
Collision detection | 60 Hz response |
Tool ID (scalpel vs forceps) | Single camera feed via tetrad geometry |
External cameras | None |
Figure 7.3 Left Normal postpartum uterine anatomy rendered at procedure start, showing organ boundaries and surgical landmarks. Right: Active hemorrhage state triggered by incorrect plane entry, uterine vasculature turns red. Correct tool placement for 3 seconds resolves the bleed and confirms procedural success.