CD-RCM: Generalizable Continuous-Depth Novel View Synthesis for Reflectance Confocal Microscopy
1
Authors: Tooba Imtiaz1, Milind Rajadhyaksha2, Kivanc Kose*,2, and Jennifer Dy*,1
2026
1
2
Presenter: Tooba Imtiaz
Reflectance Confocal Microscopy (RCM)
2
en-face @ surface
? unknown depths
en-face @ depth 3μm
en-face @ depth 6μm
en-face @ depth 200μm
z
200μm
Sparse, anisotropic z-stack
? unknown depths
? unknown depths
z
non-invasive, in vivo imaging
0.5μm
Reflectance Confocal Microscopy (RCM)
3
z
0.5μm
Sagittal View
Coronal View
Angular cross-section
Cross-Sectional Views
z-stack
Reflectance Confocal Microscopy (RCM)
4
z
0.5μm
z-stack
Sagittal View
Coronal View
Angular cross-section
Cross-Sectional Views
Clinician must scroll through 50–65 slices;�mentally reconstruct 3D structure
3D tissue architecture is precomputed and presented; no mental reconstruction needed!
Matches vertical section geometry of H&E histopathology; clinicians read it with existing training
En-face geometry has no equivalent in histopathology; clinicians must translate b/w imaging modes
Structural inconsistencies across depth (registration errors, acquisition noise) immediately visible as discontinuities
Motion artifacts and low-signal slices appear as individual bad frames; can be missed/misinterpreted
DEJ, the primary diagnostic landmark for melanoma staging, renders as an identifiable transition
DEJ appears as gradual textural shift across many slices; no clear boundary visible in any one frame
Comparison of Imaging Paradigms
5
Modeling RCM Acquisition as Axial Translation
6
Camera Pose
cᵢ ← (cᵢ − c̄) / c∞
Shared normalized coordinate system preserves relative depth.
Camera intrinsics K
- Identity rotation: R = I3
- Pure z-translation:
- World-to-camera transform:
🡪 Camera pose (c2w):
constant axial step-size
slice index
🡪 Standardized poses:
Stack Canonicalization
and
constant scaling factor
Method
7
Self-attention
Patchify
Transformer Block (x 24)
Input I2 �& Plücker Rays
Predicted Target�Ît
Input I1 �& Plücker Rays
Ground Truth�It
xk
xk+1
yl
ŷl
MLP
Target Plücker Rays
RCM Probe
Z
Skin section (in vivo)
Linear
Unpatchify
Linear
LoRA-adapted Dinov3, trained on dermatology images
Method
8
z
INPUT
3 sparse slices
CD-RCM
OUTPUT
predicted targets
z
INPUT
3 consecutive slices
CD-RCM
continuous depth interpolation
Training Time
Inference Time
OUTPUT
Results
9
Qualitative comparison of NDS methods. CD-RCM best preserves fine cellular structures and depth-dependent morphology, producing predictions closest to GT.
Quantitative comparison. CD-RCM outperforms classical interpolation and learned baselines across photometric (PSNR, SSIM) and perceptual (LPIPS, LSPF) metrics
Qualitative comparison of NDS methods. CD-RCM best preserves fine cellular structures and depth-dependent morphology, producing predictions closest to GT.
CD-RCM – Results
10
CD-RCM: Generalizable Continuous-Depth Novel View Synthesis for Reflectance Confocal Microscopy
11
2026