2 of 37

Medical Device From Chula 1997-2017

  • Most devices/ Projects are started as students / residents projects
  • Funded by
    • Ratchadabhiseksompoj Fund
    • NRCT
    • Self-fund
  • EEG/PSG
  • Bililight
  • Portable ECG
  • Pulse oximeter(NRCT)
  • Bedside Monitor & Vital-sign Tele-medcine(NRCT)
  • Defibrillator(NRCT)
  • Uro-flow meter
  • OSLER
  • SOS
  • ANES-Train of Four(Royal Col. Anes)
  • Portable Sleep Screener
  • TRCS-HighFlow (Intronic)
  • Transport Ventilator (NRCT)

3 of 37

EEG 1997-2000

  • 1997 : 4th year Engineering Sutdent Project, continue in 1998 & 1999
  • Validated 1999 @Chula
  • Ratchadbhisek Sompoj Fund & The Supporting Funds for The Electrical Engineer Students
  • 1St price NRCT 2000
  • Engineering Colleagues
    • Cherdkul Sopawanich
    • Manop Wonsaisuwan
    • Kongpop U-Yen
    • Manisa-U-Yen
    • Kanawat Senanan
    • Taweesak Thanthipwan
    • Pished Bunnun

4 of 37

EEG-PSG 2000-2005

  • 4 Machines sales 2005-2007
  • Linux/Window operating system
  • 2/4 still operational
  • All Manual assembly / Multiple discrete parts

​

5 of 37

EEG-PSG Present

  • USB power, IEC 601.1
  • Surface mounted, Factory asembly
  • Complete from electrodes, amplifier, effort belts, thermistor, pressure air flow & CPAP titrator, Bodyposition, SpO2 & ETCO2

6 of 37

In use in

Government Hospitals

  • ICU EEG Monitor : Cardiac ICU ราชวิถี, รพ.สรรพประสิทธิประสงค์ อุบลราชธานี
  • PSG : วชิรพยาบาล, รพ.โรคทรวงอก, รพ. ตำรวจ,
  • Software : PSG exam SST 2010,2011

Private Hospitals

  • EEG: Vibharam, PYT2
  • PSG: Chaophraya, ENT, PYT2, RAM, Rajthani, Ruma-Pathe, SMV, Synpathes, Sukumvit,Ubolrak Vibharam, Vibhawadee,
  • Lamdong Medical College, Vietnam

7 of 37

A Comparative Study, Efficacy of Phototherapy for Treatment of Neonatal Hyperbilirubinemia: �LED vs Fluorescence

Tawatchai Sarntipipat, MD

Tayard Desudchit, MD

Pimolrat Thaithumyanon, MD

Department of Pediatrics,

Faculty of Medicine, Chulalongkorn University

8 of 37

Phototherapy

9 of 37

การวัด Spectrometer [Wavelenght]�ที่คณะวิทยาศาสตร์ จุฬาลงกรณ์มหาวิทยาลัย�ดร. สนอง

10 of 37

​

LED

[Commercial]

Fluorescence

​

LED

[Custom built]

Intensity at 30 cm

[mW/cm2/nm]

30

20

20

Spectrum of wave length [nm]

450-470

400-500

450-470

Power and voltage

0.18 W,12 V

100 LEDs

20 W 220 V x 4

18 W 220 V x4

0.18 W,16 V

900 LEDs

Heat production

No

Yes

No

Life time [hours]

10,000

1,000

10,000

Price [Bahts]

50,000-100,000

5,000-8,000

8,000

11 of 37

Fluorescence

LED

12 of 37

Randomized Controlled Trials�LED vs Fluorescence Phototherapy

Patient outcomes

Simple randomization

​

70 cases

LED

​

70 cases

Fluorescence

Clinical practice guideline

KCMH

Treatment

Term infants with

jaundice

Exclusion criteria

ABO or Rh incompatibility

DIC

Hematoma

Respiratory distress

Emergency problems

*Study Protocol: IRB No. 332/53, January 1st-December 15th, 2011

​

1° : bilirubin decreasing rate

​

2°: duration of phototherapy

13 of 37

LED reduced bilirubin faster than traditional fluorescent phototherapy unit

13.31±1.30

13.30±1.32

p-value=0.96

10.80±1.36

9.78±1.31

Bilirubin level (mg/dL)

0.09±0.06mg/dL/hour

0.14±0.06 mg/dL/hour

*p-value<0.001

12.5

10

7.5

5

2.5

0

10

20

30

Time (hours)

*p-value<0.001

24.09±3.36 hours

26.27±7.37 hours

*p-value 0.027

LED

Fluorescent

14 of 37

Custom built LED�phototherapy

15 of 37

�YOUNG INVESTIGATOR AWARD 2012

THE ROYAL COLLEGE OF PEDIATRICIANS OF THAILAND

16 of 37

> 600 Units in Thailand

17 of 37

Accuracy of MDCU portable ECG �compare with standard 12 leads ECG�at King Chulalongkorn Memorial Hospital

Tawatchai Sarntipipat, MD

Tayard Desudchit, MD

Apichai Khongphatthanayothin, M.D.

​

18 of 37

What is This?

CPU + VGA +

Mainboard + RAM + etc.

19 of 37

Portable ECG with tablet or smartphone

20 of 37

ADS1298 + PSOC5-> USB

  • ADS1298 + PSOC5
  • USB power
  • Isolated USB (ADUM- by analog device)
  • Calibrated with Fluke ECG calibrator
  • Record as EDF
  • Display via EDF Viewer/ECG Viewer software
  • Clinically Validated

21 of 37

2.Comparative study Accuracy of ECG�Portable ECG Vs Standard ECG

ECG wave parameter

Standard 12 leads ECG

[Mortara]

Portable 12 leads ECG

[ADS1298FE]

Amplitude [mm.]

Amplitude [mm.]

Duration

[msec]

Duration

[msec]

Investigation

[Diagnositc test]

Patients at

Pediatric cardiology ward

Compare Mean between 2 groups with

Paired t-test

22 of 37

23 of 37

24 of 37

​

​

Standard ECG

Mortara

ELI250

Standard ECG

Mortara

ELI250

​

​

​

​

QRS Axis

Positive

QRS Axis

Negative

​

​

MDCU

Portable ECG

ADS 1298

QRS Axis

Positive

117

0

117

PPV

=117/117

= 100%

MDCU

Portable ECG

ADS 1298

​

QRS Axis

Negative

0

27

27

NPV

=117/117

= 100%

​

​

​

117

27

144

​

​

​

Sensitivity

=117/117

= 100%

Specificity

=27/27

= 100%

​

​

Agreement of Test

25 of 37

Conclusion

  • MDCU Portable ECG had a good accuracy as well as Standard ECG machine[Mortara instrument ELI250]

at King Chulalongkorn Memorial Hospital.

​

Mortara instrument

ELI 250

Price 3,400-4,000 USD

MDCU Portable ECG

Price 350-500 USD

26 of 37

Discussion: apply for Holter or Even recorder

  • Portable ECG device or machine which you can picked up everwhere.
  • Use as Holter monitoring[24 Hours-more than 30 days]
  • Use as Event recorder, Loop cycle event recorder for ECG.
  • Combine with other monitor as Vital sign monitor
  • Combine with Polysomnography test
  • Apply for Real time telemonitor ECG. Via internet, Wifi, 3G or 4G

27 of 37

28 of 37

Uroflow Meter

29 of 37

Validation

  • With Standard know flow compare to commercial uroflow meter
  • Then with patients in clinical validation

​

30 of 37

Biphasic Portable �External Defibrillator

Nipon Chattipakorn, Tayard Desudchit, Cherdkul Sopawanit, Apichai Khongpattanayothin

31 of 37

32 of 37

การทดสอบทางไฟฟ้า (Electrical Signal Test)

  • โครงการได้ทดสอบโปรแกรม Touch Screen Portable Vial Sign Monitor for Local/Remote ซึ่งเชื่อมต่อกับ Modules ต่าง ๆ ผ่านทาง Serial & USB Port เพื่อ นำ ECG wave, HR, SpO2. PR, และ Plethmogram wave ที่ได้มาแสดงผลบนจอ โดยใช้ Fluke Index II SpO2 Tested และ Fluke Impulse 7000 DP ECG/Defibrillator Tester กำเนิดสัญญาณ พบว่าสามารถรับและแสดงผลการวัดได้ถูกต้อง ตามคุณสมบัติของ Module แต่ละตัว

33 of 37

ECG Connectors

  • A) ภาพแสดง EKG/Chest impedance electrodes ที่ใช้ในการทดสอบกับคน B) การต่อเชื่อมกับเครื่อง Fluke EKG Simulator

34 of 37

  • Test with Fluke Impulse 7000 DP
  • Can detect HR 10-254 BPM
  • Amplitude .05-5 mV
  • Detect on the rising edge of R wave- 2mm after the peak of R wave

​

35 of 37

Waveforms

36 of 37

ผลการวิจัย(Results)