��Medical Device From Chula 1997-2022
Tayard Desudchit, Md
Cherdkul Sopavanich
Chotsak Kulkitiyakorn
Pished Bunnin
Kanawat Senanan
Thaweesak Talthipwan
Pornlert Chatkraew,MD
Tawatchai Sarntipipat,MD
Suntchai Wirotsaengthong,MD
Ukrit Meukchit
Apichai Khongpatanayothin MD.
Medical Device From Chula 1997-2017
EEG 1997-2000
EEG-PSG 2000-2005
EEG-PSG Present
In use in
Government Hospitals
Private Hospitals
A Comparative Study, Efficacy of Phototherapy for Treatment of Neonatal Hyperbilirubinemia: �LED vs Fluorescence
Tawatchai Sarntipipat, MD
Tayard Desudchit, MD
Pimolrat Thaithumyanon, MD
Department of Pediatrics,
Faculty of Medicine, Chulalongkorn University
Phototherapy
การวัด Spectrometer [Wavelenght]�ที่คณะวิทยาศาสตร์ จุฬาลงกรณ์มหาวิทยาลัย�ดร. สนอง
| LED [Commercial] | Fluorescence | LED [Custom built] |
Intensity at 30 cm [mW/cm2/nm] | 30 | 20 | 20 |
Spectrum of wave length [nm] | 450-470 | 400-500 | 450-470 |
Power and voltage | 0.18 W,12 V 100 LEDs | 20 W 220 V x 4 18 W 220 V x4 | 0.18 W,16 V 900 LEDs |
Heat production | No | Yes | No |
Life time [hours] | 10,000 | 1,000 | 10,000 |
Price [Bahts] | 50,000-100,000 | 5,000-8,000 | 8,000 |
Fluorescence
LED
Randomized Controlled Trials�LED vs Fluorescence Phototherapy
Patient outcomes
Simple randomization
70 cases
LED
70 cases
Fluorescence
Clinical practice guideline
KCMH
Treatment
Term infants with
jaundice
Exclusion criteria
ABO or Rh incompatibility
DIC
Hematoma
Respiratory distress
Emergency problems
*Study Protocol: IRB No. 332/53, January 1st-December 15th, 2011
1° : bilirubin decreasing rate
2°: duration of phototherapy
LED reduced bilirubin faster than traditional fluorescent phototherapy unit
13.31±1.30
13.30±1.32
p-value=0.96
10.80±1.36
9.78±1.31
Bilirubin level (mg/dL)
0.09±0.06mg/dL/hour
0.14±0.06 mg/dL/hour
*p-value<0.001
12.5
10
7.5
5
2.5
0
10
20
30
Time (hours)
*p-value<0.001
24.09±3.36 hours
26.27±7.37 hours
*p-value 0.027
LED
Fluorescent
Custom built LED�phototherapy
�YOUNG INVESTIGATOR AWARD 2012
THE ROYAL COLLEGE OF PEDIATRICIANS OF THAILAND
> 600 Units in Thailand
Accuracy of MDCU portable ECG �compare with standard 12 leads ECG�at King Chulalongkorn Memorial Hospital
Tawatchai Sarntipipat, MD
Tayard Desudchit, MD
Apichai Khongphatthanayothin, M.D.
What is This?
CPU + VGA +
Mainboard + RAM + etc.
Portable ECG with tablet or smartphone
ADS1298 + PSOC5-> USB
2.Comparative study Accuracy of ECG�Portable ECG Vs Standard ECG
ECG wave parameter
Standard 12 leads ECG
[Mortara]
Portable 12 leads ECG
[ADS1298FE]
Amplitude [mm.]
Amplitude [mm.]
Duration
[msec]
Duration
[msec]
Investigation
[Diagnositc test]
Patients at
Pediatric cardiology ward
Compare Mean between 2 groups with
Paired t-test
| | Standard ECG Mortara ELI250 | Standard ECG Mortara ELI250 | | |
| | QRS Axis Positive | QRS Axis Negative | | |
MDCU Portable ECG ADS 1298 | QRS Axis Positive | 117 | 0 | 117 | PPV =117/117 = 100% |
MDCU Portable ECG ADS 1298 | QRS Axis Negative | 0 | 27 | 27 | NPV =117/117 = 100% |
| | 117 | 27 | 144 | |
| | Sensitivity =117/117 = 100% | Specificity =27/27 = 100% | | |
Agreement of Test
Conclusion
at King Chulalongkorn Memorial Hospital.
Mortara instrument
ELI 250
Price 3,400-4,000 USD
MDCU Portable ECG
Price 350-500 USD
Discussion: apply for Holter or Even recorder
Uroflow Meter
Validation
Biphasic Portable �External Defibrillator
Nipon Chattipakorn, Tayard Desudchit, Cherdkul Sopawanit, Apichai Khongpattanayothin
การทดสอบทางไฟฟ้า (Electrical Signal Test)
ECG Connectors
Waveforms
ผลการวิจัย(Results)