McGill Residency Session Notes
Prepared by John Kildea
Contents
     Dosimetry, Treatment Planning and Setup
Penumbra and Collimation
Large Patients
Bolus, Compensator, Build-up Bolus, Beam Spoiler
Bolus
Compensator
Build-up Bolus
Beam Spoiler
Wedges
Orthovoltage Therapy
Interesting Facts re Orthovoltage
Orthovoltage Prescriptions (notes R3)
Orthovoltage X-ray Machines
Electron Therapy
Cerebrospinal Irradiation (CSI)
Field Matching - General Notes
Pelvis Targets
High and Low Energy Beams
Radiation Physics
HVL and TVL of Co-60 in Lead
Range of Electrons
Orthovoltage - Dose to Bone
General Physics and Math
The Inverse-Square Law
The Taylor Expansion & The Inverse-Square Law
Quick Dose Estimates
Units
Administrative
Ordering a New Machine or Something New for the Department
How much do things cost?
Quality Assurance
Ion Chambers and Beam Calibration Notes
TG-51 (Clinical Reference Dosimetry)
General Notes
Health Physics
Probability of Developing/Dying from Cancer
Effect of acute low LET uniform whole body radiation
All exposure categories – collective effective dose [%] - 2006
Effective radiation doses from image guidance in RT
General Radiation Protection
CNSC Dose Limits
Peripheral Dose
Physical Data
Linac Shielding Data
Treatment Site Details
Heterogeneity Corrections
Information Technology
Radiosurgery
Literature
Good-To-Know Reports:
Dosimetry, Treatment Planning and Setup
Penumbra and Collimation
- 4 types of penumbra in the beam profile:
- Transmission
- Geometrical
- Leakage
- Scatter
- Note: Transmission and leakage are different
- Transmission is through the edges of the jaws, leakage is through the head of the linac as a whole

- Geometrical penumbra is less for collimation near the patient's skin (actually on the patient's skin for orthovoltage) than it is for simple collimation using the main collimator
- Also, geometrical penumbra is greater for extended SSD treatments, as can be seen from the figure below.

- Orthovoltage collimation is done with lead right on the patient's skin for collimation
- this has the advantage of reduced penumbra
- and the added advantage of being able to use reduced margins
- reduced margins are possible because the patient motion is less (ie motion due to breathing and setup error is less as the field is right on the patient's skin and the collimation will move with the patient (William showed example using piece of paper with a hole in it and patient in high-E and ortho)
Top
Large Patients
- Large patients for which part of the patient is outside of the field-of-view of the planning CT image can have their tumor underdosed
- Basically, there is extra tissue that is not shown on the CT but which will cause attenuation of the beam.
Top
Bolus, Compensator, Build-up Bolus, Beam Spoiler
Bolus
- P 259 Kahn
- G-32 Dosimetry notes
- A layer of tissue-equivalent material placed directly on the skin surface to even out irregular contours of a patient in order to present a flat surface to the beam - ie this accounts for missing tissue and is a form of manual isodose line correction
- Bolus works well for orthovoltage radiation but for higher-energy beams it results in the loss of skin-sparing => use compensators for high-energy beams
- Example of bolus in clinic???


Compensator
- P 259 Kahn
- G-32 Dosimetry notes
- An irregularly shaped layer of material placed in the beam to account for missing tissue (the irregular shape accounts for the shape of the missing tissue)
- Design of the compensator must account for
- Beam divergence
- The difference in the attenuation coefficients between the compensator material and the soft tissue it is replacing - in order to replace exactly the right amount of tissue.
- The compensator basically reduces the amount of radiation into the body where there is missing tissue.
- The compensator is placed away from the skin in order to maintain skin-sparing.
- The compensator should be at least 20 cm away from the skin to avoid electron contamination.
Build-up Bolus
- P 259 Kahn
- A layer of tissue-equivalent material placed directly on the skin surface to provide adequate dose buildup over the skin surface.
- Ie build-up bolus is used to bring dmax closer to the surface - ie to maximize the dose on the surface
- This is similar to a beam-spoiler (I think), except that a beam-spoiler is held at an appropriate distance from the skin.
- For example, build-up bolus may be used to treat scar tissue with an MV beam by intentionally maximizing the dose on the skin surface (p280 Kahn)
Beam Spoiler
- P 281 Kahn
- A beam spoiler is used to bring the dmax for a high energy beam close to the surface, while maintaining the greater penetration advantage of the high-energy beam as desired
- Essentially, the secondary electrons produced in the beam spoiler provide electron contamination to increase the build-up dose under the skin of the patient
- For example for a large breast patient (in which the tumor extends deep into the breast from the surface of the breast), the beam spoiler can be used to bring the dmax close to the surface to ensure that the tumor close to the surface gets enough dose, while simultaneously ensuring that the deeper tumor gets adequate dose
- This is similar to build-up bolus (I think), except that build-up bolus is placed directly on the skin
Top
Wedges
- Advantages of dynamic wedges
- Physical wedges have a maximum field size limit, dynamic wedges do not
- Physics is easier without wedges - no need for calibration of physical wedges, etc
- Faster treatments - less MUs and no need to go into room to change wedge
- Clearance issues are simplified
- Disadvantages of dynamic wedges
- Jaws only do dynamic wedges in one direction, so planning more difficult
- Are wedges better for superficial or deep-seated tumors?
- Better for superficial tumors
- Dose at thin part of wedge will be higher than prescription
- Dose at thick part of wedge less than prescription

- Use when have two obliques - eg when treating superficial tumor at forehead
- Hinge angle - need to study
- (180 - angle between beams)/2
- Typical physical wedge factors
- If forget wedge for one fraction
- Investigate - see what the problem was exactly
- Possible solution is to use a steeper wedge for the remainder of the treatment so that the part of the field that got more dose than it should have gets less the next time round
MLCs
- If MLC doesn't work and patient treated with open beam
- Patient will get the more dose for the same number of MUs, based on the different between IMRT and open beams
Orthovoltage Therapy
Interesting Facts re Orthovoltage
- Used to treat surface lesions
- PTV is on the patient's skin, collimation is on the patient's skin
- PTV = CTV since margins tiny
- Tiny geometrical penumbra and collimation moves with the patient
- 50% dose is at what depth? About 7 cm

- In orthovoltage, electron contamination is an issue
- Especially from high-Z materials in the beam
- Applicators often made of plastic (low Z material) and may be closed to reduce electron contamination
- Orthovoltage is cheaper (x-ray tube versus linac)
- Geometrical penumbra is less
- Since collimation on the skin
- Shielding is more effective
- For example, if treating the eye, electrons would produce a higher dose behind the shield than orthovoltage photons would - why? because of bremsstrahlung? -John Kildea 2/2/10 1:29 PM
- Orthovoltage therapy is prescribed to the surface, electrons are prescribed to R90 or R80
Orthovoltage Prescriptions (notes R3)
- Prescribed to the surface in Roentgen (exposure) or Gray (dose)
- Can be prescribed in 3 ways and it is important to know exactly which way the physician is using:
- Exposure in air
- Exposure with full backscatter
- Dose in tissue (at surface)
Orthovoltage X-ray Machines
- Electrostatic x-ray production (as opposed to cyclic, which is a linac or a cyclotron)
Electron Therapy
- Good for concurrent boosts
- No need for additional infrastructure
- Difficult to achieve exact desired dose distribution
- Bremsstrahlung contamination
- Need for applicators on treatment machine - risk of collision
Cerebrospinal Irradiation (CSI)
- This is irradiation of the brain and the spinal cord
- About 12 cases per year at the MGH
- Of which maybe one is an adult
- It requires field matching
- In all cases the lateral brain fields need to be matched with the spinal cord field
- For adults, more than one spinal cord field will likely be required and must also be matched
- The patient can be either prone or supine
- Prone was the original position used in the 70s, since it was easy to feel the spinal cord
- Supine is the preferred position today, since it is easier to setup and is more comfortable for the patient
- Tumors treated with CSI include
- Meduloblastomas and apendenomas in which the cells metastasize by travelling in the cerebrospinal fluid (CSF)
- Today at the MGH, most of these cases are treated using Tomo
- Conventional treatment includes the following:
- Lateral fields for the brain, face blocked off with MLC
- Posterior field(s) for the spinal cord, with the patient in the prone position
- Don't use laterals for the cord, since the arms and the body get in the way
- Problem is the junctions - hot and cold spots (hot beyond junction, cold before junction due to diverging beams)
- Solve junction issue using:
- Half block technique (one jaw in at field center, other jaw out to the edge of the field) => gives sharp dose drop off at the center, that can be matched without hot and cold spots
- Extended SSD (means that a single beam can cover the whole spinal cord)
- Feather the junction(s) - move the junction around so that the affect of any mismatching are lessened. Typically, change the junction once per week
- For the cerebrospinal junction (CSJ),Â
- Rotate the collimator for the lateral fields to match the edge of the diverging posterior field
- kick (ie move slightly) the couch

Field Matching - General Notes
- Field-matching at the skin is the easiest
- Can use the light fields
- However, the result will be overdosing at depth due to beam divergence
- Clinically, most field matching is done at depth
- Diverging edges must match at the required depth
- The 50% isodose levels should add up at the required depth
- Simple field matching can be done using a half-block technique and rotating the gantry
- Example is McGill breast technique
- A block is used to shield one side of the field from the isocenter to the edge
- This produces a non-diverging field edge
- The first field is irradiated
- The gantry is rotated and the first field is blocked while the second field is irradiated
- The second field has a non-diverging field edge that matches the non-diverging edge of the first field
- Can also kick the couch to obtain to account for beam divergence without having to use a half-block
- See examples from junction lab residency session
Image below shows how to calculate the surface gap between two fields based on them matching at depth z

Pelvis Targets
- Generally treated with a four-field box
- AP/PA and two laterals
- The laterals could be higher energy to account for the lateral width of the patient being greater than the anterior-posterior distance
- Could also treat with a 3-field arrangement with wedges if want to avoid a structure in the posterior or anterior beam - eg want to avoid the colon
High and Low Energy Beams
- When considering energy need to consider the following points
- Depth of tumor
- Depth of zmax/build-up region
- Location and size of tumor (eg a small tumor deep into the lung will be within the build-up region for the air-tissue interface within the lung)
- Leakage and neutron dose (high when put a lot of material in the beam, such as in IMRT)
- Low energy (6 MV) used for
- IMRT
- Shallow tumors
- Small tumors in the lung (since would be in electronic disequilibrium region for high-energy beams, particularly if beams are not parallel opposed where the exit dose might compensate)
- High energy (18 MV) used for
- Deep tumors
- Patients with large separation
Typical PDDs
- Simple numbers to remember
- Cobalt: 30%
- 6 MV: 15%
- 18 MV 10%
- Draw point for dmax location and PDD at 10 cm and interpolate between them
- 5 cm ~80% + E
- 10 cm ~60% + E
- 20 cm ~35% + E
- 30 cm ~15% + EÂ

Radiation Physics
HVL and TVL of Co-60 in Lead
- Co-60 => 1.25 MeV =>
 = 5.88 x 10-2 
- =>

- =>

- =>
 - would take a lot to make a lead jacket thick enough to wear in a Co-60 beam. - Of course TVL = 3.32HVL (since
 and
 =>Â
)
Range of Electrons
- Rules of thumb for photons->electrons->range in H20:
- Mean energy of an MV photon spectrum = 1/3rd the MV energy
- Examine mean energy position in curtains plot to see what interaction is dominantÂ
- Use aveage energy transfer fraction for the dominant process to determine the mean photon energy transferred to the initial kinetic energy of the secondary electrons (p254 Dr. Podgorsak's book for summmary of processes, p246 for curtins plot and p199 for the Compton graph 0.02 0.14 0.44 0.68 0.80)
                                                                        10-2 10-1 100 101 102
- Take 1/2 of the initial energy of the secondary electrons to be their mean kinetic energy (see "Slowing Down Electron Spectra" notes from Radiation Physics class - the square plot)
- Use the rule of thumb that the range of electrons in H2O in cm = 1/2 the kinetic energy of the electrons in MeV
- Example:
- 6 MV beam
- => 3 MeV mean photon energy
- => Compton dominant
- => Compton fraction about 0.5 or so
- => 0.5 x (3 MeV) = 1.5 MeV is mean initial electron kinetic energy
- => (1.5 MeV)/2 = 0.75 MeV mean kinetic energy of electron slowing down spectrum
- => 0.75/2 = 0.375 cm is range of electrons in water for 6 MV beam
- Actual value from NIST is more like 0.3 cm2/g = 0.3 cm => reasonably close!
- Density H2O = 1g/cm3
- Denisty air = 1.293 x 10-3Â g/cm3Â at STP
- So, air is 1000 times less dense than water
- So, electrons should go about 1000 times further in air than water
- So, 0.3 cm in H2O should give 0.3 x 1000 cm in air = 300 cm = 3 m (for the example of 0.75 MeV electrons above)
- Actual value from NIST is 3.136E-01 cm2/g => (3.136 x 10-1) / (1.293 x 10-3) cm = 242 cm = 2.42 m (bit less than estimate but not bad)
- Range of 6 MeV electrons
3 cm in H2O => 3 x 1000 cm in air = 3000 cm in air = 30 m by estimate
- By NIST it is 3.25 g/cm2 = (3.25)/ (1.293 x 10-3) cm = 2513 cm in air = 25.13 m (little less than estimate but not bad)
Top
Orthovoltage - Dose to BoneÂ
- The collision air Kerma in air is related to exposure in air by
- Dose to small mass of medium in air is such that :
 .  For X-ray below 350 kVp,Â
 = 1. SoÂ
  whereÂ
 .- The f-factor or Roentgen-to-tad factor converts Exposure to Dose for beam under 3 MeV.
  
- Therefore for low energy beam like in Orthovoltage therapy treatment, absorbed dose to the bone can be up to about 4 times the absorbed dose to tissue.
General Physics and Math
The Inverse-Square Law
- for point sources only
- for primary beam only
- really applied to areas on a sphere, not on planes
- If going from shorter distance f1 to longer distance f2, expect dose will decrease from larger dose dp1 to lesser dose dp2
- => divide by larger number

- Say have orthovoltage beam with 50 mm SSD into curved eye socket for which outer part of socket is closer to beam by 5 mm than inner part - what is the % larger dose to the outer part of the socket and how can a better dose distribution be achieved? Say 5 Gy to central (lower) part of socket
- closer by 5 mm => more dose => divide by smaller number


- => % increase is

- How can a better dose distribution be achieved?
Top
The Taylor Expansion & The Inverse-Square Law
- Taylor expansion & inverse-square law used together to estimate dose differences at different SSD:
- Eg 1: Patient needs to be treated at SSD = 100 cm, but mistreated at SSD = 105 cm. Quick estimation of dose difference:
- using inverse-square law (further from source => dose reduced => divide by larger number)
 - So more or less 10% decrease in dose would be observed.
- Eg 2: Patient to be treated at SSD = 100 cm but accidentally at SSD = 98 cm
- Using inverse-square law (closer to source => dose increased => divide by the smaller number)
- Â

- This explains why in Orthovoltage the use of cones with larger SSD (50 cm) are preferred over cones with smaller SSD (20 cm). A small variation in treatment SSD for cone with small SSD has a higher impact on dose delivery as compared to the same variation in treatment SSD for cones with larger SSD.
Top
Quick Dose Estimates

Units
- Exposure: The charge of either sign collected for a given mass of air

- Unit: Roentgen
- Notes: Valid only for air and for photons with energy less than 3 MeV (measurement is impractical above 3 MeV)
- KERMA:Â
- Absorbed Dose:Â
- Equivalent Dose:
- Effective Dose:
- Collective Effective Dose
- Relationship between Dose and Exposure

- So, for 1 R:Â
 - ie 1 R is approximately equivalent to 1 cGy (or 1 RÂ
 1 cSv)
Administrative
Ordering a New Machine or Something New for the Department
Things to think about:
- Identify the needs for the machine - why is it interesting/useful/needed in the department?
- What (if any) other machines in the department can do the same job?
- Where will it go?
- What shielding/resources are required for it?
- Will a license be needed?
- How much will it be used?
- What sort of maintenance is needed?
- Is there enough staff?
- What sort of measuring/calibration equipment will be required?
- Will construction be required?
How much do things cost?
- An ion chamber: ~5 000 CAD
- Ion chamber calibration at standard's lab: ~2 000 CAD
- Full equipped Linac ~3M CAD
- Scanning Water Tank ~100K CAD
Quality Assurance
Ion Chambers and Beam Calibration Notes
- At low energy a graphite chamber with an aluminum central electrode are used. This ensures a flat energy response
- At low energies the photons get attenuated in the graphite wall but their loss is compensated for by increased photoelectric electrons from the aluminum central electrode
- In fact chambers can be "tuned" in all sorts of ways with all sorts of materials in order to ensure a flat energy response
- At high-energies the secondary electrons have a long range
- => they are produced in the phantom and penetrate into the cavity where they ionize the air
- So, at high energies the secondary electrons ionize the air in the ion chamber
- And it is assumed that the photons do not ionize the air
- At low energies the secondary electrons do not have a range long enough to penetrate the chamber wall
- So, at low energies the ionization of the air is as a result of the photons themselves
- Narrow beam geometry is preferred (even though real work is with broad beams) because all calibrations in the standard's lab and elsewhere are done using narrow beams
- It is easier to replicate narrow beams than it is to replicate broad beams => narrow beams are the standard
- Air Kerma is only defined for build up
- What does this statement mean?
- has to do with energy transferred, right?
- Parallel plate chambers versus cylindrical chambers
- PP use in the build-up region
- But need to worry about Compton current - check dosimetry notes (R2 and R6) -John Kildea 1/4/10 10:37 PM
Top
TG-51 (Clinical Reference Dosimetry)
General Notes
- Photons: 60Co to 50 MV
- Electrons: 4 MeV to 50 MeV
- TG-51 is used to determine absorbed dose to water at the point of measurement of the ion chamber placed under reference conditions
Â
- Where
 is the absorbed dose to water at the point of measurement, under reference conditions
 is the fully corrected ion chamber reading
 is the quality conversion factor to convert the calibration factor from
 to the beam quality Q- and
 is the calibration factor for a
 beam
- Essentially, when carrying out a TG-51 calibration, everything is provided, except for Q and
, which must be measured
 for photons- and
 for electrons
- M is the raw ion chamber reading corrected for ion recombination, temperature and pressure, electrometer (if electrometer and chamber calibrated separately as is the case in the US) and polarity
Health Physics
Probability of Developing/Dying from Cancer
- 40% of Canadian women and 45% of men will develop cancer during their lifetime.
- Approximately 1 out of every 4 Canadians will die from cancer.
- Three types of cancer account for the majority of new cases in each sex:
- prostate, lung and colorectal in males and
- breast, lung and colorectal in females.
- Lung cancer remains the leading cause of cancer death for both men and women.
- Colorectal cancer is the second leading cause of death from cancer.
- Risk of developing cancer & heritable effects as a function of dose (ICRP 103)
Table1 – Detriment-adjusted nominal risk coefficients (10-2 Sv-1) for stochastic effects after exposure to radiation at low dose rate.
Â
- The practical system of radiological protection recommended by the Commission is based upon the assumption that, at doses below about 100 mSv, a given increment in dose will produce a directly proportionate increment in the probability of incurring cancer or heritable effects attributable to radiation
- The Linear-Non-Threshold (LNT) model
- Conclusions on the in-utero risks of tissue injury and malformation at doses below about 100 mGy of low-LET radiation (ICRP Publication 90 - Biological effects after prenatal irradiation (embryo and fetus)):
- Embryonic susceptibility to the lethal effects of irradiation in the pre-implantation period of embryonic developments will be very infrequent at doses under 100 mGy.
- Risks of malformation after in-utero exposure to doses well below 100 mGy are not expected.
a Limits on effective dose are for the sum of the relevant effective doses from external exposure in the specified time period and the committed effective dose from intakes of radionuclides in the same period.
b Averaged over 1cm3 area of skin regardless of the area exposed.
c With the further provision that the effective dose should not exceed 50 mSv in any single year. Additional restrictions apply for the occupational exposure of pregnant women.
d In special circumstances, a higher value of effective dose could be allowed in a single year, provided that the average over 5 years does not exceed 1 mSv per year.
NCRP Report no. 160: Ionizing radiation exposure of the population of the US

All exposure categories – collective effective dose [%] - 2006

Effective radiation doses from medical examinations
o Radiography: chest 0.1 mSv ; upper GI tract 2 mSv ; lower GI tract 4 mSv
o Mammography: 0.7 mSv.
o CT: head 2 mSv ; chest 8 mSv ; abdomen 10 mSv
Effective radiation doses from image guidance in RT
o Portal vision: 4 – 10 cGy per image pair
o Cone beam CT: pelvis (125 kVp) à 3 – 5 cGy to the skin
head (100 kVp) à 0.5 – 4 cGy to the skin
General Radiation Protection
CNSC Dose Limits
- The CNSC dose limits are:
- NEW: 50 mSv/yr
- NEW: 100 mSv/(5 yr) - so effectively 20 mSv/yr over 5 years
- Pregnant NEW: 1 mSv for remainder of pregnancy
- Public: 1 mSv/yr
- An effective TADR of 20 mSv/hr is in place for instantaneous dose rate
- Not sure what the source of this is but it seems to be used in practise...
Peripheral Dose
- Could be dose to film left in the room during treatment or dose to a fetus
- Comprises
- Collimator scatter
- Patient scatter
- Neutron dose at high energies
- Dose from photo-activated radionuclides at high energies
- Important to know the leakage and collimator scatter components
- Since these can be reduced by placing shielding over the region of interest
- Dose due to scatter within the patient/phantom will still reach the region of interest by travelling within the patient/phantom
- TG 36Â deals with fetal dose in RT
- Experiment at MGH
- 3 TLDs left on the couch of the Novalis linac during irradiation  -John Kildea 4/19/10 12:35 PMÂ
- results will come with TLD readings from Health Canada
Physical Data
Linac Shielding Data
- Density of concrete: 2.2 g.cm-3
- Density of Lead: 11.35 g.cm-3Â
- Density of steel: 7.8 g.cm-3Â
- TVL of concrete for 6MV: 35 cm
- TVL of concrete for 18MV: 45 cm
- TVL of lead for Co-60: 46mm
- TVL of lead for 6MV: 57mm
- TVL of lead for 18MV: 57mm
- Z(Muscle) = 7.42
- Z(Water) = 7.42
- Z(Air) = 7.64
- Z(Bone) = 13.8
Treatment Site Details
- Two targets (and two plans)
- Some clinics use one, some use the other
- Use 18 MV as the lesion is deep
- so less dose outside the lung for more inside (see plot of opposing beams below)
- build-up in tumor not a concern since there are two beams, so although build-up deeper for 18 MV, the tumor will get exit dose from the opposing beam
- Use 6 MV since low density in lung
- so build-up in the tumor
- but more dose outside the lungs

- The 2 cm region in the air is known as the "flash"
- It's purpose is to account for patient motion/anatomical changes and penumbra
- Breast patients are setup lying on incline rather than flat so that breasts don't fall to side
- Having the breast to the sides would result in more dose to the lungs, which is avoided by having the patient on an incline
- lungs, and
- spinal cord (45 Gy)
- Spinal cord has dose constraint of 45 Gy
Heterogeneity Corrections
- There is a TG report that deals with hetrogeneity corrections
- Hetrogeneity corrections are important for lungs
- If turn on hetrogeneity corrections then patients are from then on treated differently than before
- A big issue on whether to turn on or not turn on hetrogeneity corrections
Information Technology
- Digital Imaging and COmmunications in Medicine
- Includes structures, beam, blocks, mlcs, etc
Radiosurgery
- Small-field dosimetry is difficult but not such a big deal in radiosurgery
- All the work is in localisation, less worried about dosimetry
- Want to blast the tumor with dose, if dosimetry off by 10% due to dosimetry not such a big deal
Literature
Good-To-Know Reports:
- ICRU 71 Prescribing, Recording, and Reporting for Electron Beams
- ICRU 50 & ICRU 62 Prescribing, Recording, and Reporting for Photon Beam
- ICRU 38 Intracavitary Therapy in Gynecology
- TG40 Comprehensive QA for Radiation Oncology
- TG45 AAPM Code of Practice for Radiotherapy Accelerators
- TG 65 Tissue Inhomogeneity Corrections for Megavoltage Photon Beams
- TG 66 Quality assurance for computed-tomography simulators and the computed-tomography-simulation process
- TG 106 Accelerator beam data commissioning equipment and procedures
- TG 53 Quality Assurance for Clinical Radiotherapy Treatment Planning
- TRS 430 Commissioning and Quality Assurance of Computerized Planning Systems for Radiation Treatment of Cancer
- TG 42 Stereotactic Radiosurgery
- TG 29 The Physical Aspects of Total and a Half Body Photon Irradiation
- TG 30 Total Skin Electron Therapy: Technique and Dosimetry
- TG 63 Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation.
- TG 65 Tissue Inhomogeneity Corrections for Megavoltage Photon Beams
- TG 76 The Management of Respiratory Motion in Radiation Oncology
- AAPM Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee.
- TG51 Protocol for Clinical Dosimetry of High-Energy Photon and Electron Beams
- TG61 AAPM protocol for 40–300 kV x-ray beam dosimetry in radiotherapy and radiobiology.
- TG43 Dosimetry of Interstitial Brachytherapy Sources
- TRS398 Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water
- TG36 Fetal Dose from Radiotherapy with Photon Beams
- ICRU 103 The 2007 Recommendations of the International Commission on Radiological Protection
- NCRP160 Ionizing Radiation Exposure of the Population of the United States
- NCRP49 Structural Shielding Design and Evaluation for Medical Use of X Rays and Gamma Rays of Energies Up to 10 MeV
- NCRP147 Structural Shielding Design for Medical X-Ray Imaging Facilities
- NCRP151 Structural Shielding Design and Evaluation for Megavoltage X- and Gamma-Ray Radiotherapy Facilities