Brachytherapy Notes
Prepared by John Kildea
Contents
Brachytherapy Notes
Contents
Advantages and Disadvantages of Brachytherapy?
Brachy Sources
Radial Dose Function for Ir-192
HDR vs LDR
Brachytherapy Dummy Wires
Radiation Safety in Brachytherapy
Brachytherapy Treatments
GYN Cases
Tandem and Colpostat Cases
Manual Calculation
Tandem and Colpostat Treatment - Manual Calculation
Tandem and Colpostat Treatment - Oncentra Post-planning Example
Butterfly Treatment Example
Vault Treatments
Tandem and Ring GYN Treatments
Prostate Cases
Physics Role at the CT Sim
Planning Notes
Putting Out an Oncentra Plan
Superficial Lesions
Example - Penis - Oncentra Planning
Example - Penis - Manual Calculation (ie sanity check)
Lung/Esophagus Cases
Dose Rate Formula
The Brachy Binder Example
Dr. X’s Actual Case - Lung Cancer Line Treatment
Brachytherapy Catheters
Brachytherapy Problems and Issues
Brachytherapy Best Practise
Advantages and Disadvantages of Brachytherapy?
- Physical advantage: improved localized delivery of dose to the target volume
- Shorter treatments for patients
- Can only treat smaller lesions
Brachy Sources
- Figure below taken from the book chapter “Isotopes and delivery systems for brachytherapy” by Anthony Flynn, additional tables from Podgorsak (IAEA) and Cember
- Ruthenium-106 is a beta emitter (hence it is packaged in a small plastic container inside a lead pig). It nominally has a half-life of 369 days and emits beta particles with maximum energy 3.53 MeV.
- In reality it has a half-life of 369 days but decays into Rhodium-106 by emission of a beta particle with maximum energy of just 39 keV. Rhodium-106, in turn, has a half life of just 30.4 seconds and decays into Paladium-106 by emission of 3.53 MeV beta particles. Given the brief half-life of Rh-106,, Ru-106 can be considered, overall, as emitting betas up to 3.53 MeV and having a half life of 369 days.



Radial Dose Function for Ir-192
- See spreadsheet and TG-43
HDR vs LDR
- LDR if for doses in the range 0.4 Gy/hr to 2 Gy/hr
- MDR is for doses in the range 2 G/hr to 12 Gy/hr
- HDR is for > 12 Gy/hr (ie 1 Gy per 5 mins)
- HDR at the MGH uses Ir-192 with a starting activity on the order of 10 Ci for a new source
- The Air KERMA rate constant for Ir-192 is about 108

- A 10 Ci source is

- Thus the Air Kerma rate at 1 cm would be

This is significantly higher than the definition of HDR, thus the MGH is definitely HDR when using Ir-192 wires
- HDR offers the advantage of not having permanent implants
- Permanent implants are not good for ALARA
- Patient should not to be near pregnant women for a long time
- Patient should wear condom during sex
- Probably HDR is less expensive as source can be reused multiple times
- HDR may have worse toxicity to OARs, since Ir-192 is more penetrating than I-125 or Pd-103
- LDR may be superior radiobiologically
- Longer protracted treatments may be better in terms of toxicity for OARs
- Fractionation is the solution
- The only difference between LDR and HDR is the dose rate, otherwise dose rate distributions are exactly the same for the same geometry and source type
Brachytherapy Dummy Wires
- Used to see distances in radiographs
- At the MGH, the following are used
- Colpostats: 1 cm apart dummies
- Tandem: 2 cm apart dummies
- Esophagus: 1 cm apart dummies
- Multiple catheter lung: various different bead distances used to distinguish between the different catheters
Radiation Safety in Brachytherapy
- Emergency stop button locations
- At and on console, inside door and on walls of suite
- Radiation survey meters and alerts
- Suture kit, wire cutters, pig
- In the event of an emergency
- Be sure it is an emergency
- Press interrupt button
- Press emergency stop buttons one at a time from outside room to inside
- Enter room with pocket dosimeter
- Go for gold - crank the gold hand crank on the afterloader
- If doesn’t come out, pull out catheter manually and place in pig
- Remove patient
- Survey the room
- Notify RSO
- Notify Nucleotron
- Close off room
- A NEW can be 1 m from the source for 1 hour before reaching 50 mSv annual limit
- Risk to patient is much higher since the source is inside them - remove it as fast as possible
Brachytherapy Treatments
Site | Prescription | Notes |
Cervix | 3 fractions of 8 Gy each with EBRT (typically 45 Gy with EBRT) | See paper by Souhami et al 2005 |
Prostate | 10 Gy in 1 fraction + 50 Gy in 20 fractions EBRT |
|
Lung |
|
|
Esophagus |
|
|
Eye (Choroidal melenoma, squamous cell carcinoma of the conjunctiva) | - Choroidal melenoma - 85 Gy - SCC of conjunctiva - 100 Gy | Eye plaque - prescription doses higher than for EBRT since better OAR sparing (eg 60 Gy in 10 fractions for SRS) |
GYN Cases
- See Guidelines by the ABS
- Ir-192 HDR intracavitary treatment using tandem and colpostats
- Prescription is to point A
- Point A is situated 2 cm laterally from the uterine canal and 2 cm above the lateral fornix, as shown in figure below (from Johns and Cunningham).
- From talking to William, point A is a point in reference to the applicator - 2 cm up from the cervical oss, which is represented by the flange, and 2 cm out laterally.
- There are technically two point As but they can be considered as just one since they should both receive the same dose. They may not receive the same dose if the setup for some reason is asymmetric
- OARs are on parametrial wall, represented by point B
- Point B is with reference to the anatomy, (in contrast to point A, which is in reference to the applicator)
- ICRU points are on the rectum and bladder walls
- Rectum point - 5 mm posterior to the vagina packing on a transverse slice
- Bladder point - at the most posterior part of the foley, which is inside the bladder
- If there are areas near the bladder and rectum points that are hotter than the points themselves, then move the points there - to error on the side of caution
- Typically the rectum and bladder get around 4 Gy per 8 Gy to point A
- In cases where colpostats are used without a tandem, the prescription may be to point X, which is a point 2 cm superior along an imaginary tandem from the point of intersection of the colpostats and an imaginary tandem placed centrally between them. See the figure below



From this figure it is clear that the bladder is anterior to the tandem and colpostats (which are inserted into the vagina) and the rectum is posterior to them.
- Tandem and colpostats along with packing (to move bladder and rectum out of the way) are inserted by physician and act as applicators for the HDR brachy source
- The tandem and colpostats are designed to bring the brachy source to the correct point for delivery of the pear-shaped radiation dose - see picture below
- After their insertion, the tandem and colpostats are clamped to treatment couch for immobilization.
- The ovoids of the colpostats provide the correct geometry for the pear-shaped dose distribution that is to be delivered.
- Patient gets AP and Lat kV radiographs and a CBCT
- The right side of the patient is marked with a CT marker
- Generally right and left are not an issue unless for some reason the applicator didn’t get inserted symmetrically - when the setup is assymetric computerised planning, rather than manual planning, should be done.
- Radiographs are used only for manual calculations
- CBCT is used for post-treatment (or pre-treatment) computerised planning with Oncentra
Manual (tables) versus Computerized Planning
- Manually planning uses tables that trace their origins to source trains of Co-60 pellets that were used when HDR was first implemented at the MUHC. Then, the source trains were prepared in advance and inserted into the patient as a single piece - thus all dwell positions had the same dwell time. The dose distributions were nice and so today manual planning uses similar distributions even though it involves a single source being stepped through each of the dwell positions.
- Manual planning is good in that it is fast and the patient can be treated quickly
- Manual planning should not be used when the setup is asymmetric or when the physician wants to deviate from the setup that the tables were designed for - see figure below for good and bad setups

Tandem and Colpostat Cases
- Picture of tandem and colpostats shown below
- The tandem goes through the flange and the colpostats go out to the level of the flange
- The flange sits against the cervical os, which is the circular opening from the vagina into the cervix.
- The flange helps steady the setup and it is also useful for visually determining the position of the cervical os (and hence points A and B) on the radiographs and CBCT

Photo of tandem and colpostats - tandem goes into the cervix, colpostats with their ovoids stay in the vagina. Notched positions at 1 cm apart are visible on the tandem. The dummy markers inserted into the tandem, however, are 2 cm apart.

Need to be careful when going from dummy positions (2 cm apart in tandem, 1 cm apart in colpostats) to source dwell positions, which are 2.5 mm apart. Note: the tandem used here is just less than 6 cm long, being the distance from the tip to the flange as evidenced by the dummies.

This figure shows the loading as per the table. The dummy and actual source positions are shown. Dummies are 2 cm apart for the tandem and 1 cm apart for the colpostats. The source positions are 2.5 mm apart for both tandem and colpostats.
The loading should be done to the flange as shown here - this means that the length of tandem to choose in the table corresponds to the distance from the tip to the flange, in this case 6 cm. If unsure of the flange position in the lateral view, check it out on the AP view

AP radiograph of the tandem and colpostats inserted into the vagina and uterus. The packing is clearly visible in this photograph - packing is used to increase the separation of the source from the posterior bladder and anterior rectal walls.

Manual Calculation
- Use standard gynecological loadings to obtain a dose rate at the prescription points (A) as previously calculated by Horacio
- Need to confirm which tandem is being used (physician will say which he/she inserted but physicist should double-check by examining the radiographs).
- The distance to measure is from the tip to the flange - each dummy position is 2 cm and each source position is 2.5 mm (or 0.25 cm)
Distance (cm) | Dwell position | Dummy Number |
0 | 1 | 1 |
0.25 | 2 |
|
0.5 | 3 |
|
0.75 | 4 |
|
1 | 5 |
|
1.25 | 6 |
|
1.5 | 7 |
|
1.75 | 8 |
|
2 | 9 | 2 |
2.25 | 10 |
|
2.5 | 11 |
|
2.75 | 12 |
|
3 | 13 |
|
3.25 | 14 |
|
3.5 | 15 |
|
3.75 | 16 |
|
4 | 17 | 3 |
4.25 | 18 |
|
4.5 | 19 |
|
4.75 | 20 |
|
5 | 21 |
|
5.25 | 22 |
|
5.5 | 23 |
|
5.75 | 24 |
|
6 | 25 | 4 |
6.25 | 26 |
|
6.5 | 27 |
|
6.75 | 28 |
|
7 | 29 |
|
7.25 | 30 |
|
7.5 | 31 |
|
7.75 | 32 |
|
- Three colpostat dwell positions are used, corresponding to the three locations nearest the intersection of the colpostats and the tandem, as seen on the lateral kV radiograph
- The calculation boils down to taking the pre-determined dose rate, from the appropriate row in the tandem and colpostats table (in cGy/sec
Ci) and using the prescription dose (in cGy) and the source activity (in Ci) to determine the dwell times (in sec) for input to the treatment delivery computer
- The tandem and colpostats table lists tandem lengths along with the number of dwell positions and their locations and the expected resultant dose rate at the prescription point as a function of the source activity and equal dwell times at the dwell position
- Give 8 Gy = 800 cGy to point A, for large colpostats and tandem of length 6 cm
- From the table below, the dose rate at point A is 1.95 cGy/sec
Ci - Activity at the time of insertion is 8.74 Ci (13 July 2011)
- Need to determine time t
- Use definition of dose rate
- From the table, we see that for the 6 cm tandem, 7 dwell positions at locations 1, 7, 10, 13, 16, 19 and 22 should be loaded with dwell times of 46.94 seconds each
- The source (ie dwell position) spacing is 2.5 mm giving a dose rate at point A of 1.95 cGy/sec
Ci - Colpostats 3, 4 and 5 should also be loaded with dwell times of 46.94 seconds each (assuming they lie laterally across the tandem. If they do not, then the positions that do lie laterally across it should be loaded)

Notice that the tandem and colpostat source positions are 2.5 mm apart.
Tandem and Colpostat Treatment - Manual Calculation
- Following the manual calculation, the treatment is setup on the treatment console computer
- Essentially the dwell positions, as determined from the table for the tandem, and as appropriate for the colpostats, are set and given the determined dwell times
- Following delivery of the treatment, a post-treatment 3D planning is performed on the CBCT using Oncentra
- The purpose of the 3D post-planning is to determine the dose distribution that was actually delivered and to determine the dose that the organs at risk (rectum and bladder) actually received.
- This information can then be factored into the external beam planning for the external beam remainder of the treatment
- Manual treatment example (another example done already above)

Tandem and Colpostat Treatment - Oncentra Post-planning Example
- Import the patient’s CBCT into Oncentra
- Open Oncentra
- Choose the patient and click the + to show the series available
- Highlight the appropriate series and click Import
- It should detect the patient ID and name
- Click New Case
- Write in a Case Label and click OK
- Click OK when it asks to open the case
- Close the CM Import dialog
- Do not click Close on the Non-Active Case window
- Click on the red Brachy icon on the left hand side
- Brings up the CBCT images
- Click anywhere on an image to start a new plan
- Label the plan as appropriate
- Choose the machine, through a series of double clicks, and click finish
- The right side of the patient’s body is marked using a CT marker
- Note that the patient is imaged feet first supine
- Eclipse will show the image as such but Oncentra will flip it to appear head first supine
- Be careful of this by examining the CT marker location and by comparing Oncentra image with Eclipse image
- Reconstruct the catheters
- Click on Catheter Reconstruction
- Click Remove All
- Add three new catheters
- Don’t add them together, rather add one, contour it and then add the others - avoids mix-up
- 1 and 2 will be the colpostats, 3 will be the tandem
- 1 will be the colpostat on the right-hand-side
- Ie the side of the patient’s body containing the marker
- For each catheter add the following information
- Offset: -6.0 mm for the metalic (non-MRI) colpostats
- See memo by B. Reniers 03 Jan 2008
- Source Step: 2.5 mm
- Start At: Tip End
- Don’t need to draw on all slices
- However, should start at tip end and work backwards
- Keep the catheter points in order, to not go back a slice
- Be careful with the Activity gadget
- It allows the points to be drawn and it tends to be turned on by default - make sure it is off before clicking
- Adjust the window and level as appropriate
- Find the adjustment tools by clicking on the Toogle window icon
- Zoom in using CRTL-scroll
- Pan around using Shift-scroll
- Activate the dwell positions
- Click the Activation tab - if it is not lit up then check that all catheters are drawn, even catheters that will not be used (such as catheter 2 in tandem and ring)
- For the appropriate tandem points, as per the manual calculation, click on the points to activate them. Remember 1 and 2 are the colpostats and 3 is the tandem
- For the prescription and the OARs (bladder and rectum)
- Click on the green +
- Add bladder point (using the little green and red pencil highlighted)
- Go to the Foley and find the bottommost point on it, add the bladder point there
- If, after the dose is calculated, it looks like there may be another point on the bladder with a hotter dose, the bladder point should be moved to it, rather than simply leaving it at the lowest Foley point
- Add rectum point
- Go to the axial slice where the tandem and colpostats cross - ie the slice where all three are in a horizontal line
- Right click -> Measures -> Distance
- Go 5 mm below the packing and add the rectum point
- Again, as for bladder, err on the side of caution and make the rectum point hotter, if in doubt
- Rename the points by clicking on Points in the list at bottom left and double-clicking on P1 and P2 to change them
- Reset the coordinate system
- Using toggle window somehow get to a right-handside panel that gives a tab called ECS Views
- Click on the ECS Views tab
- Drag the axial, sagittal and coronal views into the screen
- In the sagittal view, drag the pink crosshairs to the flange
- The flange is the ring point of the applicator at the cervixal os
- Rotate the crosshairs until tangential to the tandem
- Can check using coronal view and fix the rotation if necessary
- Add the applicator (prescription) points
- Go to the tandem slice on the coronal view
- Turn on the cm grid from the view menu
- Add points A1 and A2 2 cm superior and 2 cm left and right of the origin
- Set the exact distance using the table at the bottom of the screen - ie 20 mm
- Can re-adjust the coordinate system if needed such that the points A are symmetric about the tandem and the two colpostats lie along “x” axis.
- Normalize the distribution
- Click the normalization tab
- Normalize to points -> applicator points
- Should provide a dose distribution
- Click the prescription tab
- Insert the prescribed dose in cGy (eg 800 cGy for 8 Gy)
- Examine the planned dwell times and compare with the actual dwell times
- In the explorer at the bottom of the screen, click on Applicator and choose Active Dwell Positions and look at the time shown in seconds
- Be sure that the apparent source activity corresponds to the actual source activity on the date of the actual treatment
- If it is different, then to compare the post-planned and actual treatments, the actual treatment time should be scaled by the ratio of the source activities
- Say, the activity during the actual treatment was 5.12 Ci and the manually calculated treatment time was 80.1 sec per dwell position
- Say, that the apparent activity in Oncentra, at the time of planning is 9.95 Ci, then, if the manual calculation had been done for the Oncentra activity it would have been given a duration
- 80.1 sec x (5.12 Ci/9.95 Ci) = 41.22 sec (remember, since the apparent source activity is greater the time of treatment will be less)
- This compares very well with the calculated time of 42.7 sec per dwell position
- Put the time that was actually delivered (or its equivalent scaled time) into the treatment planning system to see the actual dose value that was delivered to point A and to the OARs
- Click on the Optimization tab and chose Manual dwell weights/times
- Highlight all the dwell positions and change their times
- Go to the Patient and Applicator points in the explorer and see what the actual dose values delivered were. Tell the physician as appropriate.
Butterfly Treatment Example
- Draw both as a single applicator until they clearly separate
- Used for treating disease that is confined to the uterus - ie no cervical involvement
- No manual calculations done (yet) for this at the MUHC as tables not yet verified
- Start at the tip end
- Load every other dwell position
- Physician should say how far down to load - eg 5 cm from the top - see figure

- Physician should also say how far out from the applicator to prescribe - eg 2 cm (in which case, position prescription pt 2 cm out)


Vault Treatments
- Two colpostats - no tandem
- Used to treat disease which is confined to the cervix or situations in which the uterus has been removed.
- Tables are provided
- Can measure the separation between the colpostats on the radiograph to be sure that the correct tables have been used.
Tandem and Ring GYN Treatments
- Manual calculation is similar to tandem and colpostats
- Count distance along tandem to centre of ring to be sure of correct tandem to load
- Ask physician which ring was used - small, medium or large
- Just a few things to note for the post-planning
- When starting the post-planning, prepare ECS views and do the catheter reconstruction in the ECS views, aligned along the tandem and perpendicular to the ring centered at the centre of the ring
- Will be easier - just a few points for the tandem and a circle of points for the ring
- Use offset of -6 mm as per the tandem and colpostats
- The tandem goes into channel 3, the ring goes into channel 1
- Need to create a short two-point catheter for channel 2, otherwise Oncentra will complain
- It will not be loaded and can be drawn anywhere
- The ring is at the same point as the flange, ie at the cervical os, and so it should be used when determining point A

Ring contoured in Oncentra
Prostate Cases
Physics Role at the CT Sim
- Want to be sure of two things
- The CT scan does not cut off before the catheters end
- Both at the template end and at the tip end (should be able to see all the catheters as black air and then eventually no catheters, as move superiorly)
- The catheters do not end within the dyed-out bladder
- If they do it will be impossible to see the actual end of the catheters as the dye is too bright. The catheters should end before the bladder
- Other things to check include: (from Devlin, on ipad)
- Catheters deep enough to cover the PTV (ie prostate)
- Foley pulled down toward bladder neck
- End of all catheters included in scan
- All fiducial markers and volumes of interest in scan
- Identity of each catheter is clear - ie they don’t overlap or twist around one another
- Important that all the catheters are well covering the prostate. If the edge of the prostate is too far from the catheters then in order to cover the edge of the prostate with the prescribed dose, a symmetrical increase in radiation will need to be given, such that in the opposite direction a large (and possibly detrimental to the urethra) dose will be given - see figure below

In order to give the prescribed dose to the edge, an equal dose will be given to the urethra for the setup shown here. If the sources are closer to the urethra than to the edge,the urethra will be overdosed relative to the edge.
Planning Notes
- Export plan from Eclipse to Oncentra
- Open plan in Oncentra
- Turn off the body
- Change the contrast as desired
- Use 17 catheters
- Offset is 0 for the comfort catheters
- Use 5 mm spacing
- Start at the connector end just below where the contours of the prostate begin
- Note the start slice and use it for all
- Contour to the tip end of each individual catheter
- Contour just a few points on each catheter
- At the tip end, if there is any hint of the catheter then contour it in
- Don’t use the template (1, 2, 3,4 on left top to bottom, then 5, 6, 7, 8), 9 is by itself in the middle. (10, 11, 12, 13 to the right of 9, then 14, 15, 16, 17)

- When finished contouring catheters, check them in the 3D view
- Autoactivate (the A icon in the Activation tab)
- Autoactivate within 3-4 mm of the CTV
- Fill in any holes within the activated catheters
- Length is 1240 mm - double check with physician/techs
- Select all catheters and change the length
- Click on IPSA icon
- Load solution Postate_ID and fill in organs as appropriate
- If names don’t correspond then change in Target definition
- Under Plan - Define Isolines
- Choose autospectrum - get red (100%) covering red (CTV)
- Under Optimisation tab - Graphical, all the way to local
- Aim for complete coverage
- Don’t want the 100% line going inside the prostate. It should cover it from the outside
- Also want no hotspots within the CTV, so reduce the white where possible, without compromising coverage
- Generally make equal zero anything that is close to zero
- Want 100% of CTV volume to get 100% of the dose
- Don’t want the prostate to be too hot, so no more than about 30% getting 1500 cGy
- Want 1 cc rectum getting < 7 Gy
- Keep the urethra to less than 1250 cGy
- A small volume can go above this if needed
Putting Out an Oncentra Plan
- Print out treatment printout
- Under Plan menu
- To paper and pdf
- Browse the DICOM export folder on the desktop and move the RP1 file tot he desktop, then import it from the desktop into RadCalc
- Do radcalc - should be within a few percent
- Print radcalc report to pdf
- Import the following documents into ARIA
- Treatment printout
- Screenshot
- RadCalc report
Superficial Lesions
Example - Penis - Oncentra Planning
- Patient with penis cancer had three interstitial catheters inserted as per figure below
- Catheters to remain in during course of treatment

- Dose of 28 Gy in 8 fractions (3.5 Gy per fraction) prescribed
- Patient CT’ed after catheters inserted, in position shown in figure above
- Physician contoured, in Eclipse, the treatment volume (only) on CT slices
- CT image loaded into Oncentra from Eclipse
- CT projection planes in Oncentra re-orientated to be perpendicular to the direction of catheter travel, so as to better scroll and draw the contour the catheter
- Catheter was loaded with dwell positions
- Optimisation done in a box
- Physician graphically adjusted the isodose lines to provide best coverage
- Plan was approved, documents printed and plan sent to treatment planning system
- Plan was checked at console and patient was treated
Example - Penis - Manual Calculation (ie sanity check)
Using John’s and Cunninghan’s tables (for Radium) as a sanity check on the total treatment time for the penis patient
- Oncentra gave the following:
- Total treatment time of 47.9 seconds
- Source initial activity: 9.52 Ci
- Source decay factor on date of treatment: 0.89
- Source activity on date of treatment: 9.52 Ci x 0.89 = 8.47 Ci
- Volume receiving the prescribed dose (from the DVH): 5.85 cm3
- Treatment volume roughly rectangular, about 1.0 cm deep, or 0.5 cm either side of central plane - ie either side of catheters.
- Can thus be imagined as a rectangular source of 1 cm thickness and 6 cm length (thus 6 cm2 area along the long dimension)

- Looking at the table for planar interstitial implants in Johns and Cunningham gives 188 mg
hr of Radium for 10 Gy - treatment area of 6 cm2 and prescription distance of 0.5 cm

- But, we want to give 3.5 Gy (per fraction)
- So we have 188 mg.hr x
= 65.8 mg
hr of Radium for 3.5 Gy
- Converting into Ci we have approximately 65.8 mCi
hr of Radium for 3.5 Gy (remember 1 g Radium = 1 Ci, so 1 mg Radium = 1 mCi) - But we are actually using Ir-192, so we need to convert from Radium to Ir-192 - use table 13.2 of Johns and Cunningham to get ratio of exposure rate constants

- Ratio is 0.825/0.46 = 1.79 (using 0.46
for Ir-192 as it is the value that is accepted today according to Michael. Actually, according to the table by Cember, at the top of this document, it should be 0.48
)
- We know that the ratio must be larger than 1, since when we multiply by this factor we should increase the treatment time
- Iridium has a lower exposure rate constant than Radium, so it should require longer treatments, a little less than twice as long
- Converting from Radium to Iridium gives
- (65.8 mCi
hr)Ra x 1.79 = 118 mCi
hr for Iridium
- Our source is 9.52 Ci
- So our treatment time is
= 0.012 hr = 44.62 seconds - This compares very well with the 47.9 seconds given by Oncentra (about 7% different, only!)
Top
Lung/Esophagus Cases
- In an actual case, Dr. X wanted to treat a patient with 8 Gy in 1 fraction, using esophagus cylinder loading
- Dr. X’s patient was actually a lung cancer patient that he chose to treat with a line brachy treatment
- Here two examples are shown
- First the standard example given in the brachy binder
- Second, the actual case that Dr. X was treating
Dose Rate Formula

The Brachy Binder Example

The example from the brachy binder

The loading table from the brachy binder - note that the last source at each end is boosted by 8/3 times the treatment time calculated for the other dwell positions
Dr. X’s Actual Case - Lung Cancer Line Treatment
- Diagnosis: Lung cancer
- Prescription: 8 Gy in 1 fraction
- Application: Esophagus cylinder loading - ie line treatment
- Length of the treatment
- Determined by the physician and marks placed on radiograph by the physician to indicate the start and end points of the treatment - ie the start and end of the actual loading of the sources. It is understood in this clinic that the sources are loaded from the start until the end - confirm with the physician if there is any doubt
- Need to determine the dwell positions of the start and end points and determine from them the length to be treated
- Prescription distance: 1 cm from the central axis
- DosePerCi: Determine from the tables for the given length and prescription distance
- Source activity on the day in question: 5.473 Ci

In this figure the start is at the tip, thus at 995 mm
Counting dummies back from the tip we have: 995, 985, 975, 965, 955, 945, 935
start end
length = 995 mm - 935 mm = 60 mm
Quantity | Value | Unit | Note |
Date | 1 Sept 2011 |
|
|
Patient | XXXX |
|
|
Source Activity | 5.473 | Ci |
|
Prescription Dose | 800.0 | cGy |
|
Prescription Distance | 1.0 | cm |
|
True catheter length from afterloader to tip | 995 | mm |
|
Dummy separation | 10 | mm |
|
Source loading start point | 995 | mm | from radiograph - starting at the tip |
Source loading end point | 935 | mm | shown on radiograph - counting dummies back from tip |
Catheter length to use in TCS | 995 | mm | Use the position of the furthest dwell position as the catheter length in the TCS. As such, the tip dwell position is at number 1 in the TCS and loading is 1,2,3, etc as opposed to having to figure out where in the centre of the catheter they should fall |
Source loading length | 60 | mm | Distance |
Number of sources to load, 5 mm spacing (from table) | 13 |
|
|
Catheter length to use in TCS | 995 | mm | See note below |
Dose rate per Ci (from table) | 6.7 | 
|
|
Dose rate at time of insertion | 

| 
| Basically the dose rate accounting for the true activity |
Calculated dwell non-end positions | 

| sec |
|
Calculated dwell time end positions | 

= 58.2 | sec | End positions are boosted by 8/3 |
Note: For the catheter length in the TCS, use the value of the start loading point, this tricks the TCS into thinking that the tip of the catheter is at the start point and so source loading can start at the tip in the TCS - ie 1, 2, 3, etc rather than loading dwell positions further back in the catheter in the TCS.
Basically, the loading starts at the furthest point and works its way back to the least furthest point. Thus the length of the catheter to put into the TCS is the furthest point.
Brachytherapy Catheters
Catheters for Interstitial Brachytherapy
Comfort catheters
Inner catheter holds the source and goes inside outer catheter that stays in patient
Easy to remove in an emergency - just pull the inner catheter out

Head and neck catheters
Just one catheter
Tricky in an emergency as tip button may not be easy to remove and cannot be pushed back through the tongue
Prostate catheters (needles)
Can be pulled out directly
Brachytherapy Problems and Issues
- Dummy source gets stuck on its way out
- The TCS error message will say where the source got stuck, examine it and think about it
- Check that the correct catheter length was specified in the TCS
- If a length too short was used the source will jam into the end and immediately return to the safe with an abort
- This happened for an esophagus case where a length of 1500 mm was used when it should have been 995 mm. The dummy source jammed at length ~1000 mm and returned to the safe with an abort.
- If the catheter goes into an applicator check that the connection is correct. Some catheters (eg those used for esophagus) must be inside the appropriate applicator or the source will get stuck at the connection
- This was an issue when checking why the esophagus case was jamming. The catheter was laid out on the table and an attempt was made to run the source into it. It jammed at length about 450 mm, corresponding to the location of the connector.
- Most common error is that the lock ring is not turned to lock the catheters in place at the afterloader and the procedure won’t proceed.

Figure: In this situation where the catheter is supposed to go into an applicator, the connector is designed such that the source will jam if the catheter is not inside the applicator
Brachytherapy Best Practise
- Critical that the correct catheter length is used. No harm to double check.
- Always good to have radiographs to be sure that things are correct and to document.
- Radiographs are your friend!
- Take good photographs of the setup. Especially important when identifying catheters on the CT for interstitial cases (head and neck, breast, sarcoma, etc).