Piezomotor calibration
The scope of this procedure is to calibrate the Piezo motor. The Piezo takes a voltage to move the tip of a needle a certain amount. During this procedure you need to map spatial motion to voltage input sent to the Piezo. The program includes a semi-automated procedure for Piezo calibration, but you can also perform the calibration without it, which is pretty good and fast (I prefer manual - tibi).
Manual adjustment of the needle before the calibration
Using the Piezo Calibration.vi program for calibration
Going manual (a faster method)
You can use the Piezo Calibration.vi program for calibration located here C:\...\TRMFM
software\Libraries\Piezo. This program is also available from the Penguin_*.vi Instrument Driver, in the Displacement Calibration mode, by pressing the button Piezo Arm Calibration. Or from the Piezo_*.vi Instrument Driver, by pressing the calibrate button.
Using the Piezo Calibration program from the Penguin_*.vi Instrument Driver will transfer automatically the calibration factor directly to the Penguin, and the value will be used during the experiment. If you use the Piezo Calibration.vi program alone you need to manually input the calibration factor into the Penguin_*.vi Instrument Driver in the Experimental Mode, the Volt/microns control (see picture below). Once you input this value it is normally saved when you exit the program. If you want to make sure that the value is recorded, in case the program crashes, press the Configure control of the Penguin_*.vi Instrument Driver and close the configuration window by pressing the Back button. You don’t need to change anything in the configuration window, opening and closing it causes the program to save the controls of this driver in a log file
.
If you don’t want to use the Piezo Calibration.vi application to calibrate the piezo you can simply do the calibration using the Piezo driver with the Image_CCD Camera driver, within the Principle Program or individually. The Image_CCD Camera driver MUST be calibrated! If it isn't use the Calibration of Camera document to see how to calibrate it. The idea is to use the microscope at 60X, having the needle tip in the field of view, and to move the needle with the piezo program (using a simple jump) a given distance, say 10 microns. With the measuring tool available on the Image_CCD Camera Driver put a 10 microns bar on the image, and match on of its ends with the tip of the needle. When you move the needle for 10 microns with the piezo, if the piezo would be well calibrated, the tip of the needle should arrive on the other end of the measurement bar. If this is not the case, you need to adjust the calibration factor, which is the “Volt/microns” control.
For questions and discussions please use this FORUM